A new microcalorimeter has been designed and realized for implementing the national power standard up to the frequency of 26.5 GHz at IEN Galileo Ferraris. It is a dry microcalorimeter, incorporating some new refinements that improve the global performances of the measurement system. This microcalorimeter has been originally designed for calibrating classic thermistor mounts fitted with PC 3.5-mm connector, but will be optimized for thermocouple power sensors, which can operate from direct current to 26.5 GHz.
[1] A. Fantom,et al. Radio Frequency and Microwave Power Measurement , 1990 .
[2] W. Peinelt,et al. Microcalorimetric measurement of the effective efficiency of microwave power sensors comprising thermocouples , 1994, Proceedings of Conference on Precision Electromagnetic Measurements Digest.
[3] F. R. Clague. A method to determine the calorimetric equivalence correction for a coaxial microwave microcalorimeter , 1994 .