Ex-vivo product performance of Diskus and Turbuhaler inhalers using inhalation profiles from patients with severe chronic obstructive pulmonary disease.

Dosing performance of dry powder inhalers is dependent on patient's inspiratory effort. This study compares the inhalation profiles generated by patients with severe obstructive lung disease using Diskus and Turbuhaler inhalers. The patient profiles are subsequently used to determine the dosing performance of fluticasone propionate Diskus and budesonide Turbuhaler inhalers. Inhalation profiles were recorded in COPD patients (FEV1 < or = 30% predicted) as they inhaled with maximal effort through the inhalers. The profiles were used in an inhalation simulator to assess the dosing performance by measuring the total emitted dose and the fine particle mass for each inhaler type. Peak inspiratory flow was significantly higher through the Diskus (mean 82.31 min(-1)) compared with Turbuhaler (mean 53.51 min(-1), difference = 28.8 l min(-1); P < 0.0001). In addition, in direct comparison of the two devices. the Diskus was shown to deliver a more consistent dose irrespective of flow than the Turbuhaler in this patient population. These findings may be of importance in optimising selection of devices for patients with severe airway obstruction.

[1]  D. Sinclair,et al.  Inhalation devices for asthma. Choosing the right one could make all the difference. , 1993, Canadian family physician Medecin de famille canadien.

[2]  A Brindley,et al.  The characterisation of inhalation devices by an inhalation simulator: The Electronic Lung. , 1994, Journal of aerosol medicine : the official journal of the International Society for Aerosols in Medicine.

[3]  P. Paggiaro,et al.  Multicentre randomised placebo-controlled trial of inhaled fluticasone propionate in patients with chronic obstructive pulmonary disease , 1998, The Lancet.

[4]  S. Clarke,et al.  Terbutaline sulphate Turbuhaler: effect of inhaled flow rate on drug deposition and efficacy , 1991 .

[5]  K. Brown,et al.  Respiratory tract deposition of sodium cromoglycate is highly dependent upon technique of inhalation using the Spinhaler. , 1987, British journal of diseases of the chest.

[6]  B. Meakin,et al.  Drug delivery characteristics of Bricanyl TurbohalerTM dry powder inhalers , 1995 .

[7]  W. Abdullah Arch Dis Childhood , 1994 .

[8]  H. Bisgaard,et al.  Clinical effect of Diskus dry-powder inhaler at low and high inspiratory flow-rates in asthmatic children. , 1998, The European respiratory journal.

[9]  Morton B. Brown,et al.  Robust Tests for the Equality of Variances , 1974 .

[10]  M. Kenward,et al.  Design and Analysis of Cross-Over Trials , 1989 .

[11]  H. Levison,et al.  Spacing devices and metered-dose inhalers in childhood asthma. , 1985, The Journal of pediatrics.

[12]  L. S. Hill,et al.  A comparison of the performance of two modern multidose dry powder asthma inhalers. , 1998, Respiratory medicine.

[13]  S. Pedersen,et al.  Influence of inspiratory flow rate upon the effect of a Turbuhaler. , 1990, Archives of disease in childhood.

[14]  A. Hollingworth,et al.  The relationship between powder inhaler resistance and peak inspiratory conditions in healthy volunteers--implications for in vitro testing. , 1993, Journal of aerosol medicine : the official journal of the International Society for Aerosols in Medicine.

[15]  S. Pedersen,et al.  How to use a rotahaler. , 1986, Archives of disease in childhood.

[16]  H. Bisgaard,et al.  Fine particle mass from the Diskus inhaler and Turbuhaler inhaler in children with asthma. , 1998, The European respiratory journal.

[17]  P. Burnell,et al.  Design, validation and initial testing of the Electronic LungTM device☆ , 1998 .

[18]  S. Pedersen,et al.  Fenoterol powder inhaler technique in children: influence of inspiratory flow rate and breath-holding. , 1986, European journal of respiratory diseases.

[19]  S. Holgate,et al.  Inhalation rate of sodium cromoglycate determines plasma pharmacokinetics and protection against AMP-induced bronchoconstriction in asthma. , 1988, The European respiratory journal.

[20]  F. Kazim,et al.  Effect of administration technique on bronchodilator response to fenoterol in a metered-dose inhaler. , 1983, The Journal of pediatrics.