Automatic gradient preemphasis adjustment: A 15‐minute journey to improved diffusion‐weighted echo‐planar imaging

Image distortion caused by gradient eddy currents is a major problem in the use of diffusion tensor imaging (DTI), as using the uncorrected images for calculation of apparent diffusion coefficient (ADC) and diffusion anisotropy will result in areas of artificially increased anisotropy and ADC at the edge of the images, as well as decreased spatial resolution and accuracy in ADC computations overall. This distortion may be substantially reduced by careful adjustment of the gradient preemphasis unit. A completely automatic method of adjusting the preemphasis unit is proposed which finds the optimal settings for all three gradient directions in approximately 15 min by estimating the magnitudes of the eddy currents at various delay times after a test gradient. The pixel shifts in a 64 × 128 echo‐planar diffusion‐weighted image with a diffusion gradient strength of 30 mT/m were reduced to less than 0.2. Magn Reson Med 47:208–212, 2002. © 2002 Wiley‐Liss, Inc.

[1]  D. Parker,et al.  Elimination of eddy current artifacts in diffusion‐weighted echo‐planar images: The use of bipolar gradients , 1997, Magnetic resonance in medicine.

[2]  M E Bastin,et al.  Correction of eddy current-induced artefacts in diffusion tensor imaging using iterative cross-correlation. , 1999, Magnetic resonance imaging.

[3]  M A Horsfield,et al.  Mapping eddy current induced fields for the correction of diffusion-weighted echo planar images. , 1999, Magnetic resonance imaging.

[4]  R. E. Wysong,et al.  A simple method of measuring gradient induced eddy currents to set compensation networks , 1993, Magnetic resonance in medicine.

[5]  D. van Ormondt,et al.  Improved algorithm for noniterative time-domain model fitting to exponentially damped magnetic resonance signals , 1987 .

[6]  D G Gadian,et al.  Correction for eddy current induced Bo shifts in diffusion‐weighted echo‐planar imaging , 1999, Magnetic resonance in medicine.

[7]  C. Pierpaoli,et al.  Characterization of and correction for eddy current artifacts in echo planar diffusion imaging , 1998, Magnetic resonance in medicine.

[8]  D. G. Hughes,et al.  Analysis of the temporal and spatial dependence of the Eddy current fields in a 40‐cm bore magnet , 1992, Magnetic resonance in medicine.

[9]  D. G. Hughes,et al.  Quantitative characterization of the eddy current fields in a 40‐cm bore superconducting magnet , 1994, Magnetic resonance in medicine.

[10]  C. Ahn,et al.  A New Phase Correction Method in NMR Imaging Based on Autocorrelation and Histogram Analysis , 1987, IEEE Transactions on Medical Imaging.

[11]  N G Papadakis,et al.  Gradient preemphasis calibration in diffusion‐weighted echo‐planar imaging , 2000, Magnetic resonance in medicine.

[12]  M E Bastin,et al.  On the use of water phantom images to calibrate and correct eddy current induced artefacts in MR diffusion tensor imaging. , 2000, Magnetic resonance imaging.

[13]  James R. Moore,et al.  Correction for distortion of echo‐planar images used to calculate the apparent diffusion coefficient , 1996, Magnetic resonance in medicine.

[14]  M Cercignani,et al.  An optimized pulse sequence for isotropically weighted diffusion imaging. , 1999, Journal of magnetic resonance.

[15]  P A Narayana,et al.  Reduction of pulsed gradient settling time in the superconducting magnet of a magnetic resonance instrument. , 1987, Medical physics.

[16]  R Gruetter,et al.  Localized eddy current compensation using quantitative field mapping. , 1998, Journal of magnetic resonance.