Development and characterisation of a new interface for coupling capillary LC with collision-cell ICP–MS and its application for phosphorylation profiling of tryptic protein digests

A comparison of different nebulisers for direct hyphenation of capillary and nano liquid chromatography (Cap-LC, Nano-LC) and quadrupole-based collision cell inductively coupled plasma mass spectrometry (CC-ICP–MS) for phosphorylation profiling of tryptic protein digests is described. Helium was used as cell gas and specially tuned instrumental conditions were used to achieve background minimisation at the mass of phosphorus, because of kinetic energy discrimination of the interfering polyatomic ions. The proposed set-up is based on a modified capillary electrophoresis interface and a home-made 4 mL spray chamber. It enables the use of gradient conditions with a highly concentrated organic mobile phase as often used in protein phosphorylation analysis, without the need to apply membrane desolvation for removal of the organic phase or further background minimisation. No significant signal suppression or other negative effects caused by the organic mobile phase occur, because of the low flow rates used in Cap-LC and the robust plasma conditions of the CC-ICP–MS instrument. A tryptic digest of beta-casein was investigated as model compound to demonstrate the applicability of the proposed set-up for phosphorylation profiling in protein analysis using quadrupole based collision-cell ICP–MS as phosphorus-specific detector. Detection limits for phosphorylated peptides down to the sub picomole level were obtained. As a complementary technique, electrospray ionisation tandem mass spectrometry (ESI–MS–MS) with data base searching was used for further characterisation of the phosphorylated peptides detected.

[1]  A. Stensballe,et al.  Characterization of phosphoproteins from electrophoretic gels by nanoscale Fe(III) affinity chromatography with off‐line mass spectrometry analysis , 2001, Proteomics.

[2]  K. Bennett,et al.  Phosphopeptide detection and sequencing by matrix-assisted laser desorption/ionization quadrupole time-of-flight tandem mass spectrometry. , 2002, Journal of mass spectrometry : JMS.

[3]  D. Pröfrock,et al.  Determination of phosphorus in phosphorylated deoxyribonucleotides using capillary electrophoresis and high performance liquid chromatography hyphenated to inductively coupled plasma mass spectrometry with an octopole reaction cell , 2003 .

[4]  K. Tomer,et al.  Detection and sequencing of phosphopeptides , 2000, Journal of the American Society for Mass Spectrometry.

[5]  Jack D. Henion,et al.  Packed-capillary liquid chromatography/ion-spray tandem mass spectrometry determination of biomolecules , 1991 .

[6]  J. Shabanowitz,et al.  Phosphoproteome analysis by mass spectrometry and its application to Saccharomyces cerevisiae , 2002, Nature Biotechnology.

[7]  N. Jakubowski,et al.  Quantitative determination of DNA adducts using liquid chromatography/electrospray ionization mass spectrometry and liquid chromatography/high-resolution inductively coupled plasma mass spectrometry. , 1999, Journal of mass spectrometry : JMS.

[8]  M. Zeller,et al.  The impact of chromatography and mass spectrometry on the analysis of protein phosphorylation sites , 2004, Analytical and bioanalytical chemistry.

[9]  H. Perreault,et al.  On-membrane digestion of ?-casein for determination of phosphorylation sites by matrix-assisted laser desorption/ionization quadrupole/time-of-flight mass spectrometry , 2001 .

[10]  A. Sickmann,et al.  Identification of post‐translationally modified proteins in proteome studies , 2001, Electrophoresis.

[11]  N. Packer,et al.  Protein phosphorylation: technologies for the identification of phosphoamino acids. , 1998, Journal of chromatography. A.

[12]  D. Schaumlöffel,et al.  Determination of element species at trace levels using capillary electrophoresis-inductively coupled plasma sector field mass spectrometry , 1999 .

[13]  W. Lehmann,et al.  Identification of phosphorylation sites in the polo‐like kinases Plx1 and Plk1 by a novel strategy based on element and electrospray high resolution mass spectrometry , 2002, Proteomics.

[14]  Andrew Emili,et al.  De novo peptide sequencing and quantitative profiling of complex protein mixtures using mass-coded abundance tagging , 2002, Nature Biotechnology.

[15]  J. Gebler,et al.  Identification of phosphopeptides by chemical modification with an isotopic tag and ion trap mass spectrometry. , 2002, Rapid communications in mass spectrometry : RCM.

[16]  Y. Sun,et al.  Sequencing of Gel-Isolated Proteins Using Microblotter Capillary Liquid Chromatography-Electrospray Mass Spectrometry , 1999, Journal of protein chemistry.

[17]  N. Yamada,et al.  Analysis of diluted sea-water at the ng L−1 level using an ICP-MS with an octopole reaction cell , 2002 .

[18]  W. Lehmann,et al.  Modified direct injection high efficiency nebulizer with minimized dead volume for the analysis of biological samples by micro- and nano-LC-ICP-MS , 2002 .

[19]  D. Schaumlöffel Capillary liquid separation techniques with ICP MS detection , 2004, Analytical and bioanalytical chemistry.

[20]  D. Muddiman,et al.  Selective, sensitive, and rapid phosphopeptide identification in enzymatic digests using ESI-FTICR-MS with infrared multiphoton dissociation. , 2001, Analytical chemistry.

[21]  N. Yamada,et al.  The effects of cell-gas impurities and kinetic energy discrimination in an octopole collision cell ICP-MS under non-thermalized conditions , 2002 .

[22]  W. Lehmann,et al.  Analysis of protein phosphorylation by capillary liquid chromatography coupled to element mass spectrometry with 31P detection and to electrospray mass spectrometry. , 2001, Analytical chemistry.

[23]  Joanna Szpunar,et al.  Hyphenated Techniques for Elemental Speciation in Biological Systems , 2003, Applied spectroscopy.

[24]  R. Aebersold,et al.  A systematic approach to the analysis of protein phosphorylation , 2001, Nature Biotechnology.

[25]  W. Lehmann,et al.  Stable isotope phospho-profiling of fibrinogen and fetuin subunits by element mass spectrometry coupled to capillary liquid chromatography. , 2003, Analytical biochemistry.

[26]  W. Lehmann,et al.  Protein phosphorylation degree: determination by capillary liquid chromatography and inductively coupled plasma mass spectrometry. , 2001, Analytical chemistry.

[27]  J. Banks,et al.  High-sensitivity peptide mapping using packed-capillary liquid chromatography and electrospray ionization mass spectrometry , 1996 .

[28]  William R. Pearson,et al.  A Strategy for the Rapid Identification of Phosphorylation Sites in the Phosphoproteome * , 2002, Molecular & Cellular Proteomics.

[29]  Richard D. Smith,et al.  Proteome analyses using accurate mass and elution time peptide tags with capillary LC time-of-flight mass spectrometry , 2003, Journal of the American Society for Mass Spectrometry.

[30]  D. Schaumlöffel,et al.  A new interface for combining capillary electrophoresis with inductively coupled plasma-mass spectrometry , 1999 .

[31]  R. Łobiński,et al.  Development of a sheathless interface between reversed-phase capillary HPLC and ICPMS via a microflow total consumption nebulizer for selenopeptide mapping. , 2003, Analytical chemistry.