Embalmed and fresh frozen human bones in orthopedic cadaveric studies: which bone is authentic and feasible?

Background and purpose The most frequently used bones for mechanical testing of orthopedic and trauma devices are fresh frozen cadaveric bones, embalmed cadaveric bones, and artificial composite bones. Even today, the comparability of these different bone types has not been established. Methods We tested fresh frozen and embalmed cadaveric femora that were similar concerning age, sex, bone mineral density, and stiffness. Artificial composite femora were used as a reference group. Testing parameters were pullout forces of cortex and cancellous screws, maximum load until failure, and type of fracture generated. Results Stiffness and type of fracture generated (Pauwels III) were similar for all 3 bone types (fresh frozen: 969 N/mm, 95% confidence interval (CI): 897–1,039; embalmed: 999 N/mm, CI: 875–1,121; composite: 946 N/mm, CI: 852–1,040). Furthermore, no significant differences were found between fresh frozen and embalmed femora concerning pullout forces of cancellous screws (fresh frozen: 654 N, CI: 471–836; embalmed: 595 N, CI: 365–823) and cortex screws (fresh frozen: 1,152 N, CI: 894–1,408; embalmed: 1,461 N, CI: 880–2,042), and axial load until failure (fresh frozen: 3,427 N, CI: 2,564–4290; embalmed: 3,603 N, CI: 2,898–4,306). The reference group showed statistically significantly different results for pullout forces of cancellous screws (2,344 N, CI: 2,068–2,620) and cortex screws (5,536 N, CI: 5,203–5,867) and for the axial load until failure (> 7,952 N). Interpretation Embalmed femur bones and fresh frozen bones had similar characteristics by mechanical testing. Thus, we suggest that embalmed human cadaveric bone is a good and safe option for mechanical testing of orthopedic and trauma devices.

[1]  Theo H Smit,et al.  Effect of long-term preservation on the mechanical properties of cortical bone in goats , 2008, Acta orthopaedica.

[2]  G. Njus,et al.  Pullout strength and load to failure properties of self-tapping cortical screws in synthetic and cadaveric environments representative of healthy and osteoporotic bone. , 2008, The Journal of trauma.

[3]  E. Lange Ein mechanisches Problem , 1873 .

[4]  Sebastian Kuhn,et al.  Influence of formalin fixation on the biomechanical properties of human diaphyseal bone , 2010, Biomedizinische Technik. Biomedical engineering.

[5]  S. Roe,et al.  Biomechanical properties of canine cortical bone allografts: effects of preparation and storage. , 1988, American journal of veterinary research.

[6]  J C Goh,et al.  Effect of preservation medium on the mechanical properties of cat bones. , 1989, Acta orthopaedica Scandinavica.

[7]  P Zioupos,et al.  Effect of formaldehyde fixation on some mechanical properties of bovine bone. , 1995, Biomaterials.

[8]  R. Zdero,et al.  Cancellous bone screw purchase: A comparison of synthetic femurs, human femurs, and finite element analysis , 2008, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[9]  W. Bishai,et al.  Transmission of Mycobacterium tuberculosis from a cadaver to an embalmer. , 2000, The New England journal of medicine.

[10]  F. Linde,et al.  The effect of different storage methods on the mechanical properties of trabecular bone. , 1993, Journal of biomechanics.

[11]  K. Knobloch,et al.  Differences in the Mechanical Properties of Calcaneal Artificial Specimens, Fresh Frozen Specimens, and Embalmed Specimens in Experimental Testing , 2006, Foot & ankle international.

[12]  Marco Viceconti,et al.  The effects of embalming using a 4% formalin solution on the compressive mechanical properties of human cortical bone. , 2008, Clinical biomechanics.

[13]  E. F. Byars,et al.  EFFECT OF EMBALMING ON THE MECHANICAL PROPERTIES OF BEEF BONE. , 1964, Journal of applied physiology.

[14]  S. Ruchholtz,et al.  The benefit of wire cerclage stabilisation of the medial hinge in intramedullary nailing for the treatment of subtrochanteric femoral fractures: a biomechanical study , 2011, International Orthopaedics.

[15]  H. Benghuzzi,et al.  Mechanical strength repercussions of various fixative storage methods on bone. , 2006, Biomedical sciences instrumentation.

[16]  Y. S. Choi,et al.  Comparison of mechanical rigidity between plate augmentation leaving the nail in situ and interlocking nail using cadaveric fracture model of the femur , 2011, International Orthopaedics.

[17]  G. Njus,et al.  The effect of pilot hole size on the insertion torque and pullout strength of self-tapping cortical bone screws in osteoporotic bone. , 2008, The Journal of trauma.

[18]  L Cristofolini,et al.  Mechanical validation of whole bone composite femur models. , 1996, Journal of biomechanics.

[19]  S. Esterby American Society for Testing and Materials , 2006 .

[20]  M. Panjabi,et al.  Biomechanical time‐tolerance of fresh cadaveric human spine specimens , 1985, Journal of orthopaedic research : official publication of the Orthopaedic Research Society.

[21]  R. Pelker,et al.  Biomechanical properties of bone allografts. , 1983, Clinical orthopaedics and related research.

[22]  W. Mittelmeier,et al.  Biomechanical examinations of cancellous bone concerning the influence of duration and temperature of cryopreservation. , 2001, Journal of biomedical materials research.