High-resolution CMOS imaging detector

A 2.5 cm by 5 cm, 512 by 1024 pixel CMOS photodiode array detector designed specifically for digital radiography will be described. All necessary scanning and readout circuitry is integrated within the detector. The small pixel spacing of 48 micrometers allows the imager to easily achieve the 10 lp/mm resolution required for the targeted interventional mammography application. Direct coupling to the scintillator and a pixel fill factor of more than 80% leads to high DQE over a large range of exposure values. The detector exhibits very low dark current of about 30 pA/cm2 at room temperature, which allows for low-noise operation and long integration times. Read noise of less than 200 electrons rms and a saturation level of 2.8 X 106 electrons combine for a large dynamic range greater than 80 dB. The conversion gain of the detector is 0.5 (mu) V/electron. Combined with a Gd2O2S scintillator, the imager achieves up to 25% MTF at 10 lp/mm and a DC detective quantum efficiency of 50%. The detector design is optimized for x-ray energies between 10 kV and 50 kV, but can be retrofitted with different scintillator screens to cover a large range of imaging applications up to 160 kV.

[1]  Stephen E. Reichenbach,et al.  Characterizing digital image acquisition devices , 1991 .

[2]  Daniel R. Bednarek,et al.  Evaluation of a CMOS image detector for low-cost and power medical x-ray imaging applications , 1999, Medical Imaging.

[3]  John M. Boone Spectral modeling and compilation of quantum fluence in radiography and mammography , 1998, Medical Imaging.

[4]  H Liu,et al.  Charge-coupled device detector: performance considerations and potential for small-field mammographic imaging applications. , 1992, Medical physics.

[5]  L. Mortara,et al.  Evaluations Of Charge-Coupled Device (CCD) Performance For Astronomical Use , 1981, Other Conferences.

[6]  John M. Boone,et al.  High Resolution Digital Radiography Utilizing CCD Planar Array , 1988, Medical Imaging.

[7]  David Gur,et al.  X-Ray Imaging With Two-Dimensional Charge-Coupled Device (CCD) Arrays , 1984, Medical Imaging.

[8]  A Fenster,et al.  Scanned-projection digital mammography. , 1987, Medical physics.

[9]  Larry E. Antonuk,et al.  Theoretical performance of amorphous silicon imagers in diagnostic radiology , 1996, Medical Imaging.

[10]  Michael P. Andre,et al.  Integrated CMOS-selenium x-ray detector for digital mammography , 1998, Medical Imaging.

[11]  R. K. Swank Absorption and noise in x‐ray phosphors , 1973 .

[12]  Martin J. Yaffe,et al.  Imaging performance of a prototype scanned-slot digital mammography system , 1993, Medical Imaging.

[13]  Hong Liu,et al.  Operational characteristics and potential of scientific-grade charge-coupled devices in x-ray imaging applications , 1992, Electronic Imaging.

[14]  David P. Trauernicht,et al.  The Measurement Of Conversion Noise In X-Ray Intensifying Screens , 1988, Medical Imaging.

[15]  David Gur,et al.  X-ray imaging with 2048 x 2048 CCD array , 1990, Medical Imaging.

[16]  M J Yaffe,et al.  Analysis of the spatial-frequency-dependent DQE of optically coupled digital mammography detectors. , 1994, Medical physics.

[17]  William V. Schempp,et al.  Signal, noise, and detective quantum efficiency in CCD-based x-ray imaging systems for use in mammography , 1994, Medical Imaging.

[18]  P. Judy,et al.  The line spread function and modulation transfer function of a computed tomographic scanner. , 1976, Medical physics.

[19]  Daniel R. Bednarek,et al.  Region of interest micro-angiography: radiographic imaging optimized for vascular interventions , 1999, Medical Imaging.

[20]  H. Blume,et al.  DQE(f) of four generations of computed radiography acquisition devices. , 1995, Medical physics.

[21]  Daniel M. O'Mara,et al.  Prototype CCD-based detector for whole-breast digital mammography , 1997, Medical Imaging.

[22]  Hartmut Sklebitz,et al.  OPDIMA: large-area CCD-based x-ray image sensor for spot imaging and biopsy control in mammography , 1999, Medical Imaging.

[23]  J A Rowlands,et al.  X-ray imaging with amorphous selenium: detective quantum efficiency of photoconductive receptors for digital mammography. , 1995, Medical physics.

[24]  Hans Roehrig,et al.  Digital x-ray cameras for real-time stereotactic breast needle biopsy , 1993, Medical Imaging.

[25]  Kunio Doi,et al.  A simple method for determining the modulation transfer function in digital radiography , 1992, IEEE Trans. Medical Imaging.

[26]  J. C. Dainty,et al.  Image Science: Principles, Analysis and Evaluation of Photographic-Type Imaging Processes , 1974 .