Monte Carlo Optimization of Crystal Configuration for Pixelated Molecular SPECT Scanners

Abstract Resolution-sensitivity-PDA tradeoff is the most challenging problem in design and optimization of pixelated preclinical SPECT scanners. In this work, we addressed such a challenge from a crystal point-of-view by looking for an optimal pixelated scintillator using GATE Monte Carlo simulation. Various crystal configurations have been investigated and the influence of different pixel sizes, pixel gaps, and three scintillators on tomographic resolution, sensitivity, and PDA of the camera were evaluated. The crystal configuration was then optimized using two objective functions: the weighted-sum and the figure-of-merit methods. The CsI(Na) reveals the highest sensitivity of the order of 43.47 cps/MBq in comparison to the NaI(Tl) and the YAP(Ce), for a 1.5×1.5 mm 2 pixel size and 0.1 mm gap. The results show that the spatial resolution, in terms of FWHM, improves from 3.38 to 2.21 mm while the sensitivity simultaneously deteriorates from 42.39 cps/MBq to 27.81 cps/MBq when pixel size varies from 2×2 mm 2 to 0.5×0.5 mm 2 for a 0.2 mm gap, respectively. The PDA worsens from 0.91 to 0.42 when pixel size decreases from 0.5×0.5 mm 2 to 1×1 mm 2 for a 0.2 mm gap at 15° incident-angle. The two objective functions agree that the 1.5×1.5 mm 2 pixel size and 0.1 mm Epoxy gap CsI(Na) configuration provides the best compromise for small-animal imaging, using the HiReSPECT scanner. Our study highlights that crystal configuration can significantly affect the performance of the camera, and thereby Monte Carlo optimization of pixelated detectors is mandatory in order to achieve an optimal quality tomogram.

[1]  Keiichi Magota,et al.  Performance characterization of the Inveon preclinical small-animal PET/SPECT/CT system for multimodality imaging , 2011, European Journal of Nuclear Medicine and Molecular Imaging.

[2]  Joyoni Dey High performance SPECT camera design , 2009, 2009 IEEE Nuclear Science Symposium Conference Record (NSS/MIC).

[3]  C Lartizien,et al.  GATE: a simulation toolkit for PET and SPECT. , 2004, Physics in medicine and biology.

[4]  S. Meikle,et al.  Small animal SPECT and its place in the matrix of molecular imaging technologies , 2005, Physics in medicine and biology.

[5]  M. Ay,et al.  Quantitative assessment of crystal material and size on the performance of rotating dual head small animal PET scanners using Monte Carlo modeling. , 2012, Hellenic journal of nuclear medicine.

[6]  A. Rahmim,et al.  Resolution-recovery-embedded image reconstruction for a high-resolution animal SPECT system. , 2014, Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics.

[7]  M. Ay,et al.  Performance evaluation of a newly developed high‐resolution, dual‐head animal SPECT system based on the NEMA NU1–2007 standard , 2014, Journal of applied clinical medical physics.

[8]  Habib Zaidi,et al.  Recent developments and future trends in nuclear medicine instrumentation. , 2006, Zeitschrift fur medizinische Physik.

[9]  Angelika Bischof Delaloye EJNMMI Research: a new journal in nuclear medicine , 2010, European Journal of Nuclear Medicine and Molecular Imaging.

[10]  Jasbir S. Arora,et al.  Survey of multi-objective optimization methods for engineering , 2004 .

[11]  Michael V. Green,et al.  MONICA: a compact, portable dual gamma camera system for mouse whole-body imaging. , 2010, Nuclear Medicine and Biology.

[12]  Konstantina S. Nikita,et al.  Performance evaluation of a mouse-sized camera for dynamic studies in small animals , 2007 .

[13]  G. Loudos,et al.  GATE simulations for small animal SPECT/PET using voxelized phantoms and rotating-head detectors , 2006, 2006 IEEE Nuclear Science Symposium Conference Record.

[14]  E. Asma,et al.  Evaluation of the impact of resolution-sensitivity tradeoffs on detection performance for SPECT imaging , 2008, 2008 IEEE Nuclear Science Symposium Conference Record.

[15]  Arman Rahmim,et al.  Design and development of a high resolution animal SPECT scanner dedicated for rat and mouse imaging , 2014 .

[16]  Bin He,et al.  Low resolution brain electromagnetic tomography in a realistic geometry head model: a simulation study , 2005, Physics in medicine and biology.

[17]  M. Ay,et al.  Effects of Crystal Pixel Size and Collimator Geometry on the Performance of a Pixelated Crystal γ-Camera Using Monte Carlo Simulation , 2010, The Journal of Nuclear Medicine Technology.