Clinical validation of a medical grade color monitor for chest radiology

Until recently, the specifications of medical grade monochrome LCD monitors outperformed those of color LCD monitors. New generations of color LCD monitors, however, show specifications that are in many respects similar to those of monochrome monitors typically used in diagnostic workstations. The aim of present study was to evaluate the impact of different medical grade monitors in terms of detection of simulated lung nodules in chest x-ray images. Specifically, we wanted to compare a new medical grade color monitor (Barco Coronis 6MP color) to a medical grade grayscale monitor (Barco Coronis 3MP monochrome) and a consumer color monitor (Philips 200VW 1.7MP color) by means of an observer performance experiment. Using the free-response acquisition data paradigm, seven radiologists were asked to detect and locate lung nodules (170 in total), simulated in half of the 200 chest X-ray images used in the experiment. The jackknife free-response receiver operating characteristic (JAFROC) analysis of the data showed a statistically significant difference between at least two monitors, F-value=3.77 and p-value =0.0481. The different Figure of Merit values were 0.727, 0.723 and 0.697 for the new color LCD monitor, the medical grade monitor and the consumer color monitor respectively. There was no difference between the needed reading times but there was a difference between the mean calculated Euclidian distances between the position marked by the observers and the center of the simulated nodule, indicating a better accuracy with both medical grade monitors. Present data suggests that the new generation of medical grade color monitors could be used as diagnostic workstations.

[1]  A. Burgess,et al.  Human observer detection experiments with mammograms and power-law noise. , 2001, Medical physics.

[2]  E Samei,et al.  Detection of subtle lung nodules: relative influence of quantum and anatomic noise on chest radiographs. , 1999, Radiology.

[3]  Ehsan Samei,et al.  Subtle lung nodules: influence of local anatomic variations on detection. , 2003, Radiology.

[4]  Maria Kallergi,et al.  Choosing a radiology workstation: technical and clinical considerations. , 2007, Radiology.

[5]  Ehsan Samei,et al.  A comparative contrast-detail study of five medical displays. , 2008, Medical physics.

[6]  K. Berbaum,et al.  Receiver operating characteristic rating analysis. Generalization to the population of readers and patients with the jackknife method. , 1992, Investigative radiology.

[7]  John F. Hamilton,et al.  A Free Response Approach To The Measurement And Characterization Of Radiographic Observer Performance , 1977, Other Conferences.

[8]  Hilde Bosmans,et al.  Preliminary validation of a new variable pattern for daily quality assurance of medical image display devices. , 2007, Medical physics.

[9]  Patrik Sund,et al.  Comparison of Color LCD and Medical-grade Monochrome LCD Displays in Diagnostic Radiology , 2007, Journal of Digital Imaging.

[10]  Dev P Chakraborty,et al.  Observer studies involving detection and localization: modeling, analysis, and validation. , 2004, Medical physics.

[11]  Dev P Chakraborty,et al.  Validation and statistical power comparison of methods for analyzing free-response observer performance studies. , 2008, Academic radiology.

[12]  E. Samei,et al.  Simulation of subtle lung nodules in projection chest radiography. , 1997, Radiology.

[13]  Jiahua Fan,et al.  Monochrome versus color softcopy displays for teleradiology: observer performance and visual search efficiency. , 2007, Telemedicine journal and e-health : the official journal of the American Telemedicine Association.

[14]  Dd Dorfman,et al.  ROC rating analysis : generalization to the population of readers and cases with the jackknife method , 1992 .

[15]  Tom Kimpe,et al.  15.3: Specificities of a Psycho‐Physical Test Room Dedicated for Medical Display Applications , 2007 .

[16]  Michael J. Flynn,et al.  Assessment of a novel, high-resolution, color, AMLCD for diagnostic medical image display: luminance performance and DICOM calibration , 2003, SPIE Medical Imaging.