Across-vendor standardization of semi-LASER for single-voxel MRS at 3 Tesla

The semi‐adiabatic localization by adiabatic selective refocusing (sLASER) sequence provides single-shot full intensity signal with clean localization and minimal chemical shift displacement error and was recommended by the international MRS Consensus Group as the preferred localization sequence at high- and ultra-high fields. Across-vendor standardization of the sLASER sequence at 3 Tesla has been challenging due to the B 1 requirements of the adiabatic inversion pulses and maximum B 1 limitations on some platforms. The aims of this study were to design a short-echo sLASER sequence that can be executed within a B 1 limit of 15 µT by taking advantage of gradient-modulated RF pulses, to implement it on three major platforms and to evaluate the between-vendor reproducibility of its perfomance with phantoms and in vivo . In addition, voxel-based first and second order B 0 shimming and voxel-based B 1 adjustments of RF pulses were implemented on all platforms. Amongst the gradient-modulated pulses considered (GOIA, FOCI and BASSI), GOIA-WURST was identified as the optimal refocusing pulse that provides good voxel selection within a maximum B 1 of 15 µT based on localization efficiency, contamination error and ripple artifacts of the inversion profile. An sLASER sequence (30 ms echo time) that incorporates VAPOR water suppression and 3D outer volume suppression was implemented with identical parameters (RF pulse type and duration, spoiler gradients and inter-pulse delays) on GE, Philips and Siemens and generated identical spectra on the GE ‘Braino’ phantom between vendors. High-quality spectra were consistently obtained in multiple regions (cerebellar white matter, hippocampus, pons, posterior cingulate cortex and putamen) in the human brain across vendors (5 subjects scanned per vendor per region; mean signal-to-noise ratio > 33; mean water linewidth between 6.5 Hz to 11.4 Hz). The harmonized sLASER protocol is expected to produce high reproducibility of MRS across sites thereby allowing large multi-site studies with clinical cohorts.

[1]  C. Juchem,et al.  FAMASITO: FASTMAP Shim Tool towards user‐friendly single‐step B0 homogenization , 2021, NMR in biomedicine.

[2]  Anke Henning,et al.  Methodological consensus on clinical proton MRS of the brain: Review and recommendations , 2019, Magnetic resonance in medicine.

[3]  Christophe Lenglet,et al.  AutoVOI: real‐time automatic prescription of volume‐of‐interest for single voxel spectroscopy , 2018, Magnetic resonance in medicine.

[4]  E. Auerbach,et al.  Transverse relaxation time constants of the five major metabolites in human brain measured in vivo using LASER and PRESS at 3 T , 2018, Magnetic resonance in medicine.

[5]  Kejal Kantarci,et al.  Improved localization, spectral quality, and repeatability with advanced MRS methodology in the clinical setting , 2018, Magnetic resonance in medicine.

[6]  Lynn E Eberly,et al.  Test‐retest reproducibility of neurochemical profiles with short‐echo, single‐voxel MR spectroscopy at 3T and 7T , 2016, Magnetic resonance in medicine.

[7]  Lynn E Eberly,et al.  Feasibility and reproducibility of neurochemical profile quantification in the human hippocampus at 3 T , 2015, NMR in biomedicine.

[8]  Romain Valabregue,et al.  Two‐site reproducibility of cerebellar and brainstem neurochemical profiles with short‐echo, single‐voxel MRS at 3T , 2015, Magnetic resonance in medicine.

[9]  T. Scheenen,et al.  Multi‐center reproducibility of neurochemical profiles in the human brain at 7 T , 2015, NMR in biomedicine.

[10]  Małgorzata Marjańska,et al.  Effect of carr‐purcell refocusing pulse trains on transverse relaxation times of metabolites in rat brain at 9.4 Tesla , 2015, Magnetic resonance in medicine.

[11]  T. Scheenen,et al.  Multi-center Reproducibility of Short Echo Time Single Voxel H MRS of the Human Brain at 7T with Adiabatic Slice- Selective Refocusing Pulses , 2012 .

[12]  Gülin Öz,et al.  Short‐echo, single‐shot, full‐intensity proton magnetic resonance spectroscopy for neurochemical profiling at 4 T: Validation in the cerebellum and brainstem , 2011, Magnetic resonance in medicine.

[13]  Eva-Maria Ratai,et al.  Spectroscopic imaging with improved gradient modulated constant adiabaticity pulses on high-field clinical scanners. , 2010, Journal of magnetic resonance.

[14]  Dennis W J Klomp,et al.  Short echo time 1H‐MRSI of the human brain at 3T with minimal chemical shift displacement errors using adiabatic refocusing pulses , 2008, Magnetic resonance in medicine.

[15]  Jens Frahm,et al.  Use of phased array coils for a determination of absolute metabolite concentrations , 2005, Magnetic resonance in medicine.

[16]  Jan M Warnking,et al.  Bandwidth‐modulated adiabatic RF pulses for uniform selective saturation and inversion , 2004, Magnetic resonance in medicine.

[17]  Peter Andersen,et al.  Proton T2 relaxation study of water, N‐acetylaspartate, and creatine in human brain using Hahn and Carr‐Purcell spin echoes at 4T and 7T , 2002, Magnetic resonance in medicine.

[18]  L DelaBarre,et al.  The return of the frequency sweep: designing adiabatic pulses for contemporary NMR. , 2001, Journal of magnetic resonance.

[19]  R Gruetter,et al.  Field mapping without reference scan using asymmetric echo‐planar techniques , 2000, Magnetic resonance in medicine.

[20]  D P Auer,et al.  On the reliability of quantitative clinical magnetic resonance spectroscopy of the human brain , 1995, NMR in biomedicine.

[21]  R. Gruetter,et al.  In vivo 1H NMR spectroscopy of rat brain at 1 ms echo time , 1999, Magnetic resonance in medicine.

[22]  M. Garwood,et al.  Adiabatic pulses , 1997, NMR in biomedicine.

[23]  R. Ordidge,et al.  Frequency offset corrected inversion (FOCI) pulses for use in localized spectroscopy , 1996, Magnetic resonance in medicine.

[24]  S. Provencher Estimation of metabolite concentrations from localized in vivo proton NMR spectra , 1993, Magnetic resonance in medicine.

[25]  R Gruetter,et al.  Automatic, localized in Vivo adjustment of all first‐and second‐order shim coils , 1993, Magnetic resonance in medicine.

[26]  J. Carlson,et al.  Rapid radiofrequency calibration in MRI , 1990, Magnetic resonance in medicine.

[27]  P. Bottomley Spatial Localization in NMR Spectroscopy in Vivo , 1987, Annals of the New York Academy of Sciences.

[28]  E. Purcell,et al.  Effects of Diffusion on Free Precession in Nuclear Magnetic Resonance Experiments , 1954 .