Feasibility of markerless 3D position monitoring of the central airways using kilovoltage projection images: Managing the risks of central lung stereotactic radiotherapy.

BACKGROUND AND PURPOSE Central lung stereotactic body radiotherapy (SBRT) can cause proximal bronchial tree (PBT) toxicity. Information on PBT position relative to the high-dose could aid risk management. We investigated template matching + triangulation for high-frequency markerless 3D PBT position monitoring. MATERIALS AND METHODS Kilovoltage projections of a moving phantom (full-fan cone-beam CT [CBCT, 15 frames/second] without MV irradiation: 889 images/dataset + CBCT and 7 frames/second fluoroscopy with MV irradiation) and ten patients undergoing free-breathing stereotactic/hypofractionated lung irradiation (full-fan CBCT without MV irradiation, 470-500 images/dataset) were retrospectively analyzed. 2D PBT reference templates (1 filtered digitally reconstructed radiograph/°) were created from planning CT data. Using normalized cross-correlation, templates were matched to projection images for 2D position. Multiple registrations were triangulated for 3D position. RESULTS For the phantom, 2D right/left PBT position could be determined in 86.6/75.1% of the CBCT dataset without MV irradiation, and 3D position (excluding first 20° due to the minimum triangulation angle) in 84.7/72.7%. With MV irradiation, this was up to 2% less. For right/left PBT, root-mean-square errors of measured versus "known" position were 0.5/0.8, 0.4-0.5/0.7, and 0.4/0.5-0.6 mm for left-right, superior-inferior, and anterior-posterior directions, respectively. 2D PBT position was determined in, on average, 89.8% of each patient dataset (range: 79.4-99.2%), and 3D position (excluding first 20°) in 85.1% (range: 67.9-99.6%). Motion was mainly superior-inferior (range: 4.5-13.6 mm, average: 8.5 mm). CONCLUSIONS High-frequency 3D PBT position verification during free-breathing is technically feasible using markerless template matching + triangulation of kilovoltage projection images acquired during gantry rotation. Applications include organ-at-risk position monitoring during central lung SBRT.

[1]  S. Senan,et al.  Feasibility of using anatomical surrogates for predicting the position of lung tumours. , 2012, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[2]  Suresh Senan,et al.  Analysis of reproducibility of respiration-triggered gated radiotherapy for lung tumors. , 2008, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[3]  Jakub Cvek,et al.  Analysis of Lung Tumor Motion in a Large Sample: Patterns and Factors Influencing Precise Delineation of Internal Target Volume. , 2016, International journal of radiation oncology, biology, physics.

[4]  Hassan Mostafavi,et al.  Subsecond and Submillimeter Resolution Positional Verification for Stereotactic Irradiation of Spinal Lesions. , 2016, International journal of radiation oncology, biology, physics.

[5]  Maureen van Eijnatten,et al.  Using 3D printing techniques to create an anthropomorphic thorax phantom for medical imaging purposes , 2018, Medical physics.

[6]  Lech Papiez,et al.  Stereotactic body radiation therapy for early-stage non-small-cell lung carcinoma: four-year results of a prospective phase II study. , 2009, International journal of radiation oncology, biology, physics.

[7]  S. Senan,et al.  An analysis of planned versus delivered airway doses during stereotactic lung radiotherapy for central tumors , 2016, Acta oncologica.

[8]  Minsong Cao,et al.  Initial clinical observations of intra- and interfractional motion variation in MR-guided lung SBRT. , 2018, The British journal of radiology.

[9]  Koichi Yamazaki,et al.  Insertion and fixation of fiducial markers for setup and tracking of lung tumors in radiotherapy. , 2005, International journal of radiation oncology, biology, physics.

[10]  Max Dahele,et al.  An analysis of patient positioning during stereotactic lung radiotherapy performed without rigid external immobilization. , 2012, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[11]  Suresh Senan,et al.  Inferring positions of tumor and nodes in Stage III lung cancer from multiple anatomical surrogates using four-dimensional computed tomography. , 2010, International journal of radiation oncology, biology, physics.

[12]  T Ackerly,et al.  Variation in Lung Tumour Breathing Motion between Planning Four-dimensional Computed Tomography and Stereotactic Ablative Radiotherapy Delivery and its Dosimetric Implications: Any Role for Four-dimensional Set-up Verification? , 2016, Clinical oncology (Royal College of Radiologists (Great Britain)).

[13]  S. Senan,et al.  Outcomes of Hypofractionated High‐Dose Radiotherapy in Poor‐Risk Patients with “Ultracentral” Non–Small Cell Lung Cancer , 2016, Journal of thoracic oncology : official publication of the International Association for the Study of Lung Cancer.

[14]  Puneeth Iyengar,et al.  Virtual Bronchoscopy-Guided Treatment Planning to Map and Mitigate Radiation-Induced Airway Injury in Lung SAbR. , 2018, International journal of radiation oncology, biology, physics.

[15]  Dynamic simulation of motion effects in IMAT lung SBRT , 2014, Radiation oncology.

[16]  S. Senan,et al.  Use of megavoltage cine-images for studying intra-thoracic motion during radiotherapy for locally advanced lung cancer. , 2011, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[17]  S Senan,et al.  An analysis of anatomic landmark mobility and setup deviations in radiotherapy for lung cancer. , 1999, International journal of radiation oncology, biology, physics.

[18]  Hassan Mostafavi,et al.  Markerless tracking of small lung tumors for stereotactic radiotherapy. , 2015, Medical physics.

[19]  P. Parikh,et al.  Cone-Beam Computed Tomography Internal Motion Tracking Should Be Used to Validate 4-Dimensional Computed Tomography for Abdominal Radiation Therapy Patients. , 2016, International journal of radiation oncology, biology, physics.

[20]  Elisabeth Weiss,et al.  Respiratory motion variability of primary tumors and lymph nodes during radiotherapy of locally advanced non-small-cell lung cancers , 2015, Radiation oncology.

[21]  Paul J Keall,et al.  Tumor and normal tissue motion in the thorax during respiration: Analysis of volumetric and positional variations using 4D CT. , 2007, International journal of radiation oncology, biology, physics.

[22]  Jef Vandemeulebroucke,et al.  The long- and short-term variability of breathing induced tumor motion in lung and liver over the course of a radiotherapy treatment. , 2017, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[23]  Hassan Mostafavi,et al.  Markerless positional verification using template matching and triangulation of kV images acquired during irradiation for lung tumors treated in breath-hold , 2018, Physics in medicine and biology.

[24]  Lena Specht,et al.  Intra- and interfraction breathing variations during curative radiotherapy for lung cancer. , 2007, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[25]  Yoichi M Ito,et al.  Evaluation of the motion of lung tumors during stereotactic body radiation therapy (SBRT) with four-dimensional computed tomography (4DCT) using real-time tumor-tracking radiotherapy system (RTRT). , 2016, Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics.

[26]  S. Senan,et al.  Stereotactic ablative radiotherapy (SABR) for central lung tumors: Plan quality and long-term clinical outcomes. , 2015, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[27]  Mitsuhiro Nakamura,et al.  The accuracy of extracted target motion trajectories in four-dimensional cone-beam computed tomography for lung cancer patients. , 2016, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[28]  J. Sonke,et al.  3D surface imaging for monitoring intrafraction motion in frameless stereotactic body radiotherapy of lung cancer. , 2012, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[29]  Quynh-Thu Le,et al.  Quantification of motion of different thoracic locations using four-dimensional computed tomography: implications for radiotherapy planning. , 2007, International journal of radiation oncology, biology, physics.

[30]  S. Senan,et al.  Analysis of carina position as surrogate marker for delivering phase-gated radiotherapy. , 2008, International journal of radiation oncology, biology, physics.