Evaluation of Matrix9 silicon photomultiplier array for small-animal PET.

PURPOSE The MatrixSL-9-30035-OEM (Matrix9) from SensL is a large-area silicon photomultiplier (SiPM) photodetector module consisting of a 3 × 3 array of 4 × 4 element SiPM arrays (total of 144 SiPM pixels) and incorporates SensL's front-end electronics board and coincidence board. Each SiPM pixel measures 3.16 × 3.16 mm(2) and the total size of the detector head is 47.8 × 46.3 mm(2). Using 8 × 8 polished LSO/LYSO arrays (pitch 1.5 mm) the performance of this detector system (SiPM array and readout electronics) was evaluated with a view for its eventual use in small-animal positron emission tomography (PET). METHODS Measurements of noise, signal, signal-to-noise ratio, energy resolution, flood histogram quality, timing resolution, and array trigger error were obtained at different bias voltages (28.0-32.5 V in 0.5 V intervals) and at different temperatures (5 °C-25 °C in 5 °C degree steps) to find the optimal operating conditions. RESULTS The best measured signal-to-noise ratio and flood histogram quality for 511 keV gamma photons were obtained at a bias voltage of 30.0 V and a temperature of 5 °C. The energy resolution and timing resolution under these conditions were 14.2% ± 0.1% and 4.2 ± 0.1 ns, respectively. The flood histograms show that all the crystals in the 1.5 mm pitch LSO array can be clearly identified and that smaller crystal pitches can also be resolved. Flood histogram quality was also calculated using different center of gravity based positioning algorithms. Improved and more robust results were achieved using the local 9 pixels for positioning along with an energy offset calibration. To evaluate the front-end detector readout, and multiplexing efficiency, an array trigger error metric is introduced and measured at different lower energy thresholds. Using a lower energy threshold greater than 150 keV effectively eliminates any mispositioning between SiPM arrays. CONCLUSIONS In summary, the Matrix9 detector system can resolve high-resolution scintillator arrays common in small-animal PET with adequate energy resolution and timing resolution over a large detector area. The modular design of the Matrix9 detector allows it to be used as a building block for simple, low channel-count, yet high performance, small animal PET or PET/MRI systems.

[1]  Tadashi Watabe,et al.  Development of a Si-PM-based high-resolution PET system for small animals , 2010, Physics in medicine and biology.

[2]  C. Levin,et al.  Fast Timing Silicon Photomultipliers for Scintillation Detectors , 2013, IEEE Photonics Technology Letters.

[3]  Yongfeng Yang,et al.  Investigation of Depth of Interaction Encoding for a Pixelated LSO Array With a Single Multi-Channel PMT , 2009, IEEE Transactions on Nuclear Science.

[4]  K. Ote,et al.  Development of a new brain PET scanner based on single event data acquisition , 2012, 2012 IEEE Nuclear Science Symposium and Medical Imaging Conference Record (NSS/MIC).

[5]  S. Cherry,et al.  A Simple Capacitive Charge-Division Readout for Position-Sensitive Solid-State Photomultiplier Arrays , 2013, IEEE Transactions on Nuclear Science.

[6]  Andrew L. Goertzen,et al.  Multiplexing Approaches for a 12 x 4 Array of Silicon Photomultipliers , 2014, IEEE Transactions on Nuclear Science.

[7]  K. Shah,et al.  Performance Measurements of CMOS Position Sensitive Solid-State Photomultipliers , 2010, IEEE Transactions on Nuclear Science.

[8]  D. Renker,et al.  Development of a novel depth of interaction PET detector using highly multiplexed G-APD cross-strip encoding. , 2014, Medical physics.

[9]  H. Ikeda,et al.  Development of a large-area monolithic 4×4 MPPC array for a future PET scanner employing pixelized Ce:LYSO and Pr:LuAG crystals , 2011 .

[10]  H. Watabe,et al.  Performance comparison of Si-PM-based block detectors with different pixel sizes for an ultrahigh-resolution small-animal PET system , 2011, Physics in medicine and biology.

[11]  C. Thompson,et al.  Evaluation of a 16:3 Signal Multiplexor to Acquire Signals From a SPM Array With Dual and Single Layer LYSO Crystal Blocks , 2011, IEEE Transactions on Nuclear Science.

[12]  Nicola Zorzi,et al.  Performance of FBK high-density SiPM technology coupled to Ce:LYSO and Ce:GAGG for TOF-PET. , 2014, Physics in medicine and biology.

[13]  J. Karp,et al.  Continuous LYSO-SSPM array based PET detectors for clinical and small volume imaging studies , 2011, 2011 IEEE Nuclear Science Symposium Conference Record.

[14]  Volkmar Schulz,et al.  Sensitivity Encoded Silicon Photomultipliers (SeSPs): A novel detector design for uniform crystal identification , 2011, 2011 IEEE Nuclear Science Symposium Conference Record.

[15]  S. Siegel,et al.  A sub-millimeter resolution PET detector module using a multi-pixel photon counter array , 2008, 2008 IEEE Nuclear Science Symposium Conference Record.

[16]  Hao Peng,et al.  Investigation of analog charge multiplexing schemes for SiPM based PET block detectors , 2013, Physics in medicine and biology.

[17]  Y. Shao,et al.  Development of a prototype PET scanner with depth-of-interaction measurement using solid-state photomultiplier arrays and parallel readout electronics , 2014, Physics in medicine and biology.

[18]  R. R. Raylman,et al.  Construction and Evaluation of a Prototype High Resolution, Silicon Photomultiplier-Based, Tandem Positron Emission Tomography System , 2013, IEEE Transactions on Nuclear Science.

[19]  Craig S. Levin,et al.  Readout Electronics and Data Acquisition of a Positron Emission Tomography Time-of-Flight Detector Module With Waveform Digitizer , 2013, IEEE Transactions on Nuclear Science.

[20]  C. Piemonte,et al.  Characterization of the First FBK High-Density Cell Silicon Photomultiplier Technology , 2013, IEEE Transactions on Electron Devices.

[21]  Thomas K Lewellen,et al.  The challenge of detector designs for PET. , 2010, AJR. American journal of roentgenology.

[22]  A. Ganin,et al.  Time-of-Flight PET Detector Based on Multi-Pixel Photon Counter and Its Challenges , 2011, IEEE Transactions on Nuclear Science.

[23]  R Taschereau,et al.  NEMA NU-4 performance evaluation of PETbox4, a high sensitivity dedicated PET preclinical tomograph , 2013, Physics in medicine and biology.

[24]  C. Piemonte,et al.  New Developments of Near-UV SiPMs at FBK , 2013, IEEE Transactions on Nuclear Science.

[25]  Total Ionizing Dose Response of Triple-Well FET-Based Wideband, High-Isolation RF Switches in a 130 nm SiGe BiCMOS Technology , 2013, IEEE Transactions on Nuclear Science.

[26]  T. Frach,et al.  The digital silicon photomultiplier — Principle of operation and intrinsic detector performance , 2009, 2009 IEEE Nuclear Science Symposium Conference Record (NSS/MIC).

[27]  S. Cherry,et al.  A Time-Walk Correction Method for PET Detectors Based on Leading Edge Discriminators , 2017, IEEE Transactions on Radiation and Plasma Medical Sciences.

[28]  S. Cherry,et al.  Application of Silicon Photomultipliers to Positron Emission Tomography , 2011, Annals of Biomedical Engineering.

[29]  S. Cherry,et al.  Signal and noise properties of position-sensitive avalanche photodiodes , 2011, Physics in medicine and biology.

[30]  Yong Choi,et al.  MR insertable brain PET using tileable GAPD arrays , 2010, IEEE Nuclear Science Symposuim & Medical Imaging Conference.

[31]  H. Uchida,et al.  Development of a Multi-Pixel Photon Counter Module for Positron Emission Tomography , 2013, IEEE Transactions on Nuclear Science.

[32]  Massimo Mazzillo,et al.  Timing Performances of Large Area Silicon Photomultipliers Fabricated at STMicroelectronics , 2010, IEEE Transactions on Nuclear Science.

[33]  P. Fischer,et al.  Interpolating silicon photomultipliers , 2013 .

[34]  J. C. Jackson,et al.  Properties of silicon photon counting detectors and silicon photomultipliers , 2009 .

[35]  A. Pullia,et al.  Resistive or capacitive charge-division readout for position-sensitive detectors , 2002 .

[36]  Jae Sung Lee,et al.  Initial Results of Simultaneous PET/MRI Experiments with an MRI-Compatible Silicon Photomultiplier PET Scanner , 2012, The Journal of Nuclear Medicine.

[37]  S. Cherry,et al.  Comparison of large-area position-sensitive solid-state photomultipliers for small animal PET , 2012, Physics in medicine and biology.

[38]  P. Weilhammer,et al.  AX-PET: A novel PET concept with G-APD readout , 2012 .