A 3D gantry single photon emission tomograph with hemispherical coverage for dedicated breast imaging

Abstract A novel tomographic gantry was designed, built and initially evaluated for single photon emission imaging of metabolically active lesions in the pendant breast and near chest wall. Initial emission imaging measurements with breast lesions of various uptake ratios are presented. Methods: A prototype tomograph was constructed utilizing a compact gamma camera having a field-of-view of Results: As iteration number increased for the tomographically measured data at all polar angles, contrasts increased while signal-to-noise ratios (SNRs) decreased in the expected way with OSEM reconstruction. The rollover between contrast improvement and SNR degradation of the lesion occurred at two to three iterations. The reconstructed tomographic data yielded SNRs with or without scatter correction that were >9 times better than the planar scans. There was up to a factor of ∼2.5 increase in total primary and scatter contamination in the photopeak window with increasing tilt angle from 15° to 45°, consistent with more direct line-of-sight of myocardial and liver activity with increased camera polar angle. Conclusion: This new, ultra-compact, dedicated tomographic imaging system has the potential of providing valuable, fully 3D functional information about small, otherwise indeterminate breast lesions as an adjunct to diagnostic mammography.