Phase Aberration Correction Using Echo Signals from Moving Targets I: Description and Theory

Inhomogeneous acoustic velocity in human tissue introduces phase aberration in ultrasonic imaging systems and degrades image quality. A novel technique that employs echo signals from moving diffuse targets, such as flowing blood, as an image quality factor to compensate for phase aberration is described. Such signals can be obtained by subtracting the images of two consecutive target interrogations. The fundamental statistics of the quality factor and other related parameters are developed to provide a theoretical basis for the technique.

[1]  Gregg E. Trahey,et al.  A statistical analysis of phase aberration correction using image quality factors in coherent imaging systems , 1992, IEEE Trans. Medical Imaging.

[2]  M. O’Donnell,et al.  Phase-aberration correction using signals from point reflectors and diffuse scatterers: basic principles , 1988, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[3]  D. Middleton An Introduction to Statistical Communication Theory , 1960 .

[4]  G. Trahey,et al.  Comparisons of image quality factors for phase aberration correction with diffuse and point targets: theory and experiments , 1991, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[5]  F. S. Foster,et al.  The Focussing of Ultrasound Beams Through Human Tissue , 1980 .

[6]  R. F. Wagner,et al.  Statistics of Speckle in Ultrasound B-Scans , 1983, IEEE Transactions on Sonics and Ultrasonics.

[7]  D Rachlin Direct estimation of aberrating delays in pulse-echo imaging systems. , 1990, The Journal of the Acoustical Society of America.

[8]  P. Cooperberg,et al.  Ultrasonic refraction by the rectus abdominis muscles: the double image artifact. , 1984, Journal of ultrasound in medicine : official journal of the American Institute of Ultrasound in Medicine.

[9]  D C Sullivan,et al.  In vivo measurements of ultrasonic beam distortion in the breast. , 1991, Ultrasonic imaging.

[10]  R C Waag,et al.  In vivo and in vitro ultrasound beam distortion measurements of a large aperture and a conventional aperture focussed transducer. , 1988, Ultrasound in medicine & biology.

[11]  R C Chivers,et al.  Ultrasonic wave fluctuations through tissue: an experimental pilot study. , 1988, Ultrasonics.

[12]  J. Goodman Statistical Properties of Laser Speckle Patterns , 1963 .

[13]  S.W. Smith,et al.  Experimental results with a real-time adaptive ultrasonic imaging system for viewing through distorting media , 1990, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.