A new pulsatile total artificial heart using a single centrifugal pump.

A new pulsatile total artificial heart (TAH) system, combining a single centrifugal pump (CFP) with two three-way valves, was developed. One port of each three-way valve was connected to the inlet and outlet of a CFP, respectively. The other two ports of each valve ware connected to the right and left atrium, pulmonary artery, and aorta. The CFP can perfuse the pulmonary and systemic circulation alternately with pulsatile flow. A prototype system composed of a Sarns' CFP and solenoid valves was connected to a mock circulatory system resulting in 1) a pulsatile TAH that could be produced with a single CFP, 2) 5 L/min of pulsatile output with a normal flow wave form that can be obtained alternately on the right and left side by switching the solenoid valves, and 3) flow balance between the left and the right that could be controlled easily by the length of switching duration. This new system could be miniaturized and is feasible for a totally implantable TAH.