Relationship between file size and stiffness of stainless steel instruments.

The purpose of this study was to evaluate the relationship existing between file size and stiffness for three endodontic files of different cross-sections. Three groups of instruments were tested: Flexofiles from the Maillefer company, K Files and K-Flex from the Kerr company. Flexofiles have a triangular cross-section, K Files a square cross-section, and K-Flex instruments a rhombus cross-section from size 15 to size 40. For each group, 10 files from size 15 to size 40 were compared according to ANSI/ADA specification N degrees 28 for bending moment evaluation. There was a statistical difference between the three groups: the square cross-section K files presented a larger bending moment than the rhombus cross-section K Flex, which presented a higher bending moment than the triangular cross-section Flexofiles. There was an exponential relationship between file size and bending moment.

[1]  S. Montgomery,et al.  A comparison of weights of debris extruded apically by conventional filing and Canal Master techniques. , 1991, Journal of endodontics.

[2]  R E Walton,et al.  The influence of the method of canal preparation on the quality of apical and coronal obturation. , 1979, Journal of endodontics.

[3]  S. Montgomery,et al.  Another look at root canal instrumentation. , 1992, Oral surgery, oral medicine, and oral pathology.

[4]  S. Sorin,et al.  Cutting efficiency of endodontic reamers. , 1973, Oral surgery, oral medicine, and oral pathology.

[5]  D. Southard,et al.  Instrumentation of curved molar root canals with the Roane technique. , 1987, Journal of endodontics.

[6]  R. Craig,et al.  Bending and torsion of endodontic files with rhombus cross sections. , 1982, Journal of endodontics.

[7]  H. Gerstein,et al.  An initial investigation of the bending and torsional properties of Nitinol root canal files. , 1988, Journal of endodontics.

[8]  G. Marshall,et al.  Brittle and ductile torsional failures of endodontic instruments. , 1977, Journal of endodontics.

[9]  A. L. Frank,et al.  The anticurvature filing method to prepare the curved root canal. , 1980, Journal of the American Dental Association.

[10]  T. Chow Mechanical effectiveness of root canal irrigation. , 1983, Journal of endodontics.

[11]  R G Craig,et al.  Bending and torsion properites of endodontic instruments. , 1968, Oral surgery, oral medicine, and oral pathology.

[12]  J B Roane,et al.  The "balanced force" concept for instrumentation of curved canals. , 1985, Journal of endodontics.

[13]  J. Moser,et al.  Flute design of endodontic instruments: its influence on cutting efficiency. , 1982, Journal of endodontics.

[14]  F S Weine,et al.  The effect of preparation procedures on original canal shape and on apical foramen shape. , 1975, Journal of Endodontics.

[15]  B. Seto,et al.  Torsional properties of twisted and machined endodontic files. , 1990, Journal of endodontics.