Hazard analysis of fracture-reduction robot and its application to safety design of fracture-reduction assisting robotic system

In this paper, we discuss the issue of safety in robot-assisted fracture reduction. We define the hazards of robot-assisted fracture reduction and design safety control methods. Although a large reduction force is required to reduce femoral neck fractures, an unexpectedly large force produced by a robot may cause injury to the patient. We have designed two mechanical failsafe units and a software force limiter; this along with velocity control can guarantee a safe operation in reduction force. In addition, to reduce the movement of bone fragments as much as possible, we devised spatially constrained control methods for fracture-reduction robots. The fracture-reduction system was evaluated using simulated fracture reductions.

[1]  M. Parker,et al.  Internal fixation of intracapsular fractures. , 2002, Journal of the Royal College of Surgeons of Edinburgh.

[2]  N. Sugano,et al.  Development of a computer-integrated femoral head fracture reduction system , 2005, IEEE International Conference on Mechatronics, 2005. ICM '05..

[3]  Mamoru Mitsuishi,et al.  Robot-assisted femoral fracture reduction: Preliminary study in patients and healthy volunteers , 2008, Computer aided surgery : official journal of the International Society for Computer Aided Surgery.

[4]  Michael Nerlich,et al.  Reduction of femoral shaft fractures in vitro by a new developed reduction robot system 'RepoRobo'. , 2004, Injury.

[5]  Tsuyoshi Koyama,et al.  Fluoroscopic Bone Fragment Tracking for Surgical Navigation in Femur Fracture Reduction by Incorporating Optical Tracking of Hip Joint Rotation Center , 2007, IEEE Transactions on Biomedical Engineering.

[6]  Dror Paley,et al.  Principles of Deformity Correction , 2003 .

[7]  Christian Krettek,et al.  Robot-assisted Long Bone Fracture Reduction , 2009, Int. J. Robotics Res..

[8]  Tsuyoshi Koyama,et al.  A Robot Assisted Hip Fracture Reduction with a Navigation System , 2008, MICCAI.

[9]  Mamoru Mitsuishi,et al.  Development of a femur fracture reduction robot , 2004, IEEE International Conference on Robotics and Automation, 2004. Proceedings. ICRA '04. 2004.

[10]  P M Rommens,et al.  Radiation exposure to the hands and the thyroid of the surgeon during intramedullary nailing. , 1998, Injury.

[11]  C. Cooper,et al.  Hip fractures in the elderly: A world-wide projection , 1992, Osteoporosis International.