Combined OCT distance and FBG force sensing cannulation needle for retinal vein cannulation: in vivo animal validation

PurposeRetinal vein cannulation is an experimental procedure during which a clot-dissolving drug is injected into an obstructed retinal vein. However, due to the fragility and minute size of retinal veins, such procedure is considered too risky to perform manually. With the aid of surgical robots, key limiting factors such as: unwanted eye rotations, hand tremor and instrument immobilization can be tackled. However, local instrument anatomy distance and force estimation remain unresolved issues. A reliable, real-time local interaction estimation between instrument tip and the retina could be a solution. This paper reports on the development of a combined force and distance sensing cannulation needle, and its experimental validation during in vivo animal trials.MethodsTwo prototypes are reported, relying on force and distance measurements based on FBG and OCT A-scan fibres, respectively. Both instruments provide an 80 $$\upmu \hbox {m}$$μm needle tip and have outer shaft diameters of 0.6 and 2.3 mm, respectively.ResultsBoth prototypes were characterized and experimentally validated ex vivo. Then, paired with a previously developed surgical robot, in vivo experimental validation was performed. The first prototype successfully demonstrated the feasibility of using a combined force and distance sensing instrument in an in vivo setting.ConclusionThe results demonstrate the feasibility of deploying a combined sensing instrument in an in vivo setting. The performed study provides a foundation for further work on real-time local modelling of the surgical scene. This paper provides initial insights; however, additional processing remains necessary.

[1]  Wessels If,et al.  Stereoacuity and depth perception decrease with increased instrument magnification: comparing a non-magnified system with lens loupes and a surgical microscope. , 2001 .

[2]  Shikha Ambastha,et al.  Spinal needle force monitoring during lumbar puncture using fiber Bragg grating force device , 2016, Journal of biomedical optics.

[3]  Mamoru Mitsuishi,et al.  Impact of Robotic Assistance on Precision of Vitreoretinal Surgical Procedures , 2013, PloS one.

[4]  Dominiek Reynaerts,et al.  Development and Experimental Validation of a Combined FBG Force and OCT Distance Sensing Needle for Robot-Assisted Retinal Vein Cannulation , 2018, 2018 IEEE International Conference on Robotics and Automation (ICRA).

[5]  Russell H. Taylor,et al.  Force-sensing microneedle for assisted retinal vein cannulation , 2014, IEEE SENSORS 2014 Proceedings.

[6]  F. Tempia,et al.  Early Enriched Environment Exposure Protects Spatial Memory and Accelerates Amyloid Plaque Formation in APPSwe/PS1L166P Mice , 2013, PloS one.

[7]  Tien Yin Wong,et al.  The prevalence of retinal vein occlusion: pooled data from population studies from the United States, Europe, Asia, and Australia. , 2010, Ophthalmology.

[8]  Eva Lankenau,et al.  Single Scan OCT-Based Retina Detection for Robot-Assisted Retinal Vein Cannulation , 2018, J. Medical Robotics Res..

[9]  Dominiek Reynaerts,et al.  Development and experimental validation of a force sensing needle for robotically assisted retinal vein cannulations , 2015, 2015 IEEE International Conference on Robotics and Automation (ICRA).

[10]  Iulian Iordachita,et al.  A Submillimetric 3-DOF Force Sensing Instrument With Integrated Fiber Bragg Grating for Retinal Microsurgery , 2014, IEEE Transactions on Biomedical Engineering.

[11]  Arianna Menciassi,et al.  Force feedback-based microinstrument for measuring tissue properties and pulse in microsurgery , 2001, Proceedings 2001 ICRA. IEEE International Conference on Robotics and Automation (Cat. No.01CH37164).

[12]  Russell H. Taylor,et al.  3-DOF Force-Sensing Motorized Micro-Forceps for Robot-Assisted Vitreoretinal Surgery , 2017, IEEE Sensors Journal.

[13]  H. Ahmadieh,et al.  Toward the Art of Robotic-assisted Vitreoretinal Surgery , 2017, Journal of ophthalmic & vision research.

[14]  Dominiek Reynaerts,et al.  Towards a clinically applicable robotic assistance system for retinal vein cannulation , 2016, 2016 6th IEEE International Conference on Biomedical Robotics and Biomechatronics (BioRob).

[15]  Dominiek Reynaerts,et al.  Combined Force and Distance Sensing for Robot-Assisted Vitreo-Retinal Surgery , 2017 .

[16]  Russell H. Taylor,et al.  Single Fiber Optical Coherence Tomography Microsurgical Instruments for Computer and Robot-Assisted Retinal Surgery , 2009, MICCAI.

[17]  Jan Peirs,et al.  A micro optical force sensor for force feedback during minimally invasive robotic surgery , 2003 .

[18]  Xiaojing Zhang,et al.  Silicon microsurgery‐force sensor based on diffractive optical MEMS encoders , 2004 .

[19]  B. Nelson,et al.  Characterization of puncture forces for retinal vein cannulation , 2011 .

[20]  M. Zimmerman,et al.  Incidence of various types of retinal vein occlusion and their recurrence and demographic characteristics. , 1994, American journal of ophthalmology.

[21]  Dennis P. Han,et al.  A study of surgical approaches to retinal vascular occlusions. , 2000, Archives of ophthalmology.

[22]  J D Auran,et al.  Stereoacuity and depth perception decrease with increased instrument magnification: comparing a non-magnified system with lens loupes and a surgical microscope. , 2001, Binocular vision & strabismus quarterly.

[23]  Jin U. Kang,et al.  Active tremor cancellation by a "smart" handheld vitreoretinal microsurgical tool using swept source optical coherence tomography. , 2012, Optics express.

[24]  Russell H. Taylor,et al.  Force-based puncture detection and active position holding for assisted retinal vein cannulation , 2015, 2015 IEEE International Conference on Multisensor Fusion and Integration for Intelligent Systems (MFI).

[25]  Norihiko Nishizawa,et al.  Development of a Fiber-Optic Optical Coherence Tomography Probe for Intraocular Use. , 2016, Investigative ophthalmology & visual science.

[26]  R P Murphy,et al.  Distribution of iatrogenic retinal breaks in macular hole surgery. , 1995, Ophthalmology.

[27]  J. Olver,et al.  Ophthalmology at a Glance , 2005 .

[28]  Andrew M. Rollins,et al.  Integrative Advances for OCT-Guided Ophthalmic Surgery and Intraoperative OCT: Microscope Integration, Surgical Instrumentation, and Heads-Up Display Surgeon Feedback , 2014, 2015 Conference on Lasers and Electro-Optics (CLEO).

[29]  Cameron N. Riviere,et al.  Toward active tremor canceling in handheld microsurgical instruments , 2003, IEEE Trans. Robotics Autom..

[30]  Berk Gonenc,et al.  FBG-based transverse and axial force-sensing micro-forceps for retinal microsurgery , 2016, 2016 IEEE SENSORS.

[31]  Russell H. Taylor,et al.  A sub-millimetric, 0.25 mN resolution fully integrated fiber-optic force-sensing tool for retinal microsurgery , 2009, International Journal of Computer Assisted Radiology and Surgery.

[32]  Herman Bruyninckx,et al.  The real-time motion control core of the Orocos project , 2003, 2003 IEEE International Conference on Robotics and Automation (Cat. No.03CH37422).