Servo-Controlled Pneumatic Pressure Oscillator for Respiratory Impedance Measurements and High-Frequency Ventilation

The ability to provide forced oscillatory excitation of the respiratory system can be useful in mechanical impedance measurements as well as high frequency ventilation (HFV). Experimental systems currently used for generating forced oscillations are limited in their ability to provide high amplitude flows or maintain the respiratory system at a constant mean pressure during excitation. This paper presents the design and implementation of a pneumatic pressure oscillator based on a proportional solenoid valve. The device is capable of providing forced oscillatory excitations to the respiratory system over a bandwidth suitable for mechanical impedance measurements and HVF. It delivers high amplitude flows (>1.4 l/s) and utilizes a servo-control mechanism to maintain a load at a fixed mean pressure during simultaneous oscillation. Under open-loop conditions, the device exhibited a static hysteresis of approximately 7%, while its dynamic magnitude and phase responses were flat out to 10 Hz. Broad-band measurement of total harmonic distortion was approximately 19%. Under closed-loop conditions, the oscillator was able to maintain a mechanical test load at both positive and negative mean pressures during oscillatory excitations from 0.1 to 10.0 Hz. Impedance of the test load agreed closely with theoretical predictions. We conclude that this servo-controlled oscillator can be a useful tool for respiratory impedance measurements as well as HFV.

[1]  Harmonic distortion from nonlinear systems with broadband inputs: Applications to lung mechanics , 1995, Annals of Biomedical Engineering.

[2]  Carlos Canudas de Wit,et al.  A survey of models, analysis tools and compensation methods for the control of machines with friction , 1994, Autom..

[3]  David W. Kaczka,et al.  Airway and lung tissue mechanics in asthma. Effects of albuterol. , 1999, American journal of respiratory and critical care medicine.

[4]  K. Lutchen,et al.  Pseudorandom signals to estimate apparent transfer and coherence functions of nonlinear systems: applications to respiratory mechanics , 1992, IEEE Transactions on Biomedical Engineering.

[5]  N. D. Vaughan,et al.  The Modeling and Simulation of a Proportional Solenoid Valve , 1996 .

[6]  E D Michaelson,et al.  Pulmonary mechanics by spectral analysis of forced random noise. , 1975, The Journal of clinical investigation.

[7]  Zoltán Hantos,et al.  Airway and tissue mechanics in anesthetized paralyzed children , 2003, Pediatric pulmonology.

[8]  D. Navajas,et al.  Respiratory input impedance in anesthetized paralyzed patients. , 1990, Journal of applied physiology.

[9]  D. Navajas,et al.  Recording pressure at the distal end of the endotracheal tube to measure respiratory impedance. , 1989, The European respiratory journal.

[10]  David W. Kaczka,et al.  Partitioning airway and lung tissue resistances in humans: effects of bronchoconstriction. , 1997, Journal of applied physiology.

[11]  John E. Dennis,et al.  An Adaptive Nonlinear Least-Squares Algorithm , 1977, TOMS.

[12]  R. Kacmarek,et al.  Optimal Mean Airway Pressure during High-frequency Oscillation: Predicted by the Pressure–Volume Curve , 2001, Anesthesiology.

[13]  C. Mackenzie,et al.  Effect of Lung Volume on Lung Resistance and Elastance in Awake Subjects Measured during Sinusoidal Forcing , 1993, Anesthesiology.

[14]  J. Mead,et al.  Contribution of compliance of airways to frequency-dependent behavior of lungs. , 1969, Journal of applied physiology.

[15]  B. Simon,et al.  Mean airway pressure and alveolar pressure during high-frequency ventilation. , 1984, Journal of applied physiology: respiratory, environmental and exercise physiology.

[16]  David W. Kaczka,et al.  Optimal ventilation waveforms for estimating low-frequency respiratory impedance. , 1993, Journal of applied physiology.

[17]  T.F. Schuessler,et al.  A computer-controlled research ventilator for small animals: design and evaluation , 1995, IEEE Transactions on Biomedical Engineering.

[18]  K. Lutchen,et al.  Relationship between heterogeneous changes in airway morphometry and lung resistance and elastance. , 1997, Journal of applied physiology.

[19]  L P Chua,et al.  Proportional assist ventilation system based on proportional solenoid valve control. , 2001, Medical engineering & physics.

[20]  Z. Hantos,et al.  Importance of low-frequency impedance data for reliably quantifying parallel inhomogeneities of respiratory mechanics , 1988, IEEE Transactions on Biomedical Engineering.

[21]  Kenneth R. Lutchen,et al.  Technique to Determine Inspiratory Impedance during Mechanical Ventilation: Implications for Flow Limited Patients , 1999, Annals of Biomedical Engineering.

[22]  B. Suki,et al.  Forced oscillatory impedance of the respiratory system at low frequencies. , 1986, Journal of applied physiology.

[23]  J. Krishnan,et al.  High-frequency ventilation for acute lung injury and ARDS. , 2000, Chest.

[24]  D. Navajas,et al.  A system to generate simultaneous forced oscillation and continuous positive airway pressure. , 1997, The European respiratory journal.

[25]  D. Navajas,et al.  Respiratory mechanics in ventilated COPD patients: forced oscillation versus occlusion techniques. , 1998, The European respiratory journal.

[26]  P. Lopes de Melo,et al.  Linear servo-controlled pressure generator for forced oscillation measurements , 2006, Medical and Biological Engineering and Computing.

[27]  David W. Kaczka,et al.  Assessment of time-domain analyses for estimation of low-frequency respiratory mechanical properties and impedance spectra , 1995, Annals of Biomedical Engineering.

[28]  David W. Kaczka,et al.  Inspiratory lung impedance in COPD: effects of PEEP and immediate impact of lung volume reduction surgery. , 2001, Journal of applied physiology.

[29]  D. Navajas,et al.  Servocontrolled generator to measure respiratory impedance from 0.25 to 26 Hz in ventilated patients at different PEEP levels. , 1995, The European respiratory journal.

[30]  B.A. Simon,et al.  Design and calibration of a high-frequency oscillatory ventilator , 1991, IEEE Transactions on Biomedical Engineering.