Absolute quantification of myocardial blood flow with H(2)(15)O and 3-dimensional PET: an experimental validation.

UNLABELLED The purpose of this study was to assess a 3-dimensional (3D)-only PET scanner (ECAT EXACT3D) for its use in the absolute quantification of myocardial blood flow (MBF) using H(2)(15)O. METHODS Nine large white pigs were scanned with H(2)(15)O and C(15)O before and after partially occluding the circumflex (n = 4) or the left anterior descending (n = 5) coronary artery at rest and during hyperemia induced by intravenous dipyridamole. Radioactive microspheres labeled with either (57)Co or (46)Sc were injected during each of the H(2)(15)O scans, which allowed comparison between microsphere and PET measurements of regional MBF. PET analyses of 3D acquisition data were performed using filtered backprojection reconstruction and region-of-interest definition by factor and cluster analysis techniques and single-compartment model quantification. RESULTS The Hanning filter applied in image reconstruction resulted in a left atrial blood volume recovery factor of 0.84 +/- 0.06. Differences between repeated measurements of recovery were small (mean, -0.8%; range, -6.6% to 3.6%). In 256 paired measurements of MBF ranging from 0.05 to 4.4 mL. g(- 1). min(-1), microsphere and PET measurements were fairly well correlated. The mean difference between the 2 methods was - 0.11 mL. g(-1). min(-1) and the limits of agreement (+2 SD) were -0.82 and 0.60 mL. g(-1). min(-1) (Bland-Altman plot). CONCLUSION Dynamic measurements with H(2)(15)O using a 3D-only PET tomograph provide reliable and accurate measurements of absolute regional MBF over a wide flow range. The 3D acquisition technique can reduce the radiation dose to the subject while maintaining adequate counting statistics.

[1]  A. Lammertsma,et al.  A New Strategy for the Assessment of Viable Myocardium and Regional Myocardial Blood Flow Using 15O‐Water and Dynamic Positron Emission Tomography , 1992, Circulation.

[2]  Paul Kinahan,et al.  Analytic 3D image reconstruction using all detected events , 1989 .

[3]  F Shishido,et al.  Measurement of absolute myocardial blood flow with H215O and dynamic positron-emission tomography. Strategy for quantification in relation to the partial-volume effect. , 1988, Circulation.

[4]  T. Spinks,et al.  Physical characteristics of the ECAT EXACT3D positron tomograph. , 2000, Physics in medicine and biology.

[5]  P M Bloomfield,et al.  Histochemical correlates of (15)O-water-perfusable tissue fraction in experimental canine studies of old myocardial infarction. , 2000, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[6]  G Brix,et al.  Performance evaluation of a whole-body PET scanner using the NEMA protocol. National Electrical Manufacturers Association. , 1997, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[7]  M. Walsh,et al.  Noninvasive quantitation of myocardial blood flow in human subjects with oxygen-15-labeled water and positron emission tomography. , 1989, Journal of the American College of Cardiology.

[8]  William F. Jones,et al.  The architectural impact of single photon transmission measurements on full ring 3-D positron tomography , 1995, 1995 IEEE Nuclear Science Symposium and Medical Imaging Conference Record.

[9]  E. Hoffman,et al.  Quantitative measurement of myocardial blood flow with oxygen-15 water and positron computed tomography: an assessment of potential and problems. , 1985, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[10]  D L Bailey,et al.  Experience with fully 3D PET and implications for future high-resolution 3D tomographs. , 1998, Physics in medicine and biology.

[11]  M J Welch,et al.  Quantification of regional myocardial blood flow in vivo with H2150* , 2005 .

[12]  J. Yap,et al.  Assessment of the reproducibility of baseline and hyperemic myocardial blood flow measurements with 15O-labeled water and PET. , 1999, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[13]  A. Lammertsma,et al.  Myocardial blood flow: comparison of oxygen-15-water bolus injection, slow infusion and oxygen-15-carbon dioxide slow inhalation. , 1995, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[14]  J M Bland,et al.  Statistical methods for assessing agreement between two methods of clinical measurement , 1986 .

[15]  Adriaan A. Lammertsma,et al.  Measurement of myocardial blood flow with oxygen-15 labelled water: comparison of different administration protocols , 1998, European Journal of Nuclear Medicine.

[16]  D. Newport,et al.  A Single Scatter Simulation Technique for Scatter Correction in 3D PET , 1996 .

[17]  Georg Ertl,et al.  Fast High-Resolution Magnetic Resonance Imaging Demonstrates Fractality of Myocardial Perfusion in Microscopic Dimensions , 2001, Circulation research.

[18]  M. Gilardi,et al.  Physical performance of the latest generation of commercial positron scanner , 1988 .

[19]  J I Hoffman,et al.  Blood flow measurements with radionuclide-labeled particles. , 1977, Progress in cardiovascular diseases.

[20]  Michel Defrise,et al.  Exact and approximate rebinning algorithms for 3-D PET data , 1997, IEEE Transactions on Medical Imaging.

[21]  M. Phelps,et al.  Effect of short-term cardiovascular conditioning and low-fat diet on myocardial blood flow and flow reserve. , 1995, Circulation.

[22]  D. Newport,et al.  Evaluation of simulation-based scatter correction for 3-D PET cardiac imaging , 1995 .

[23]  Christopher J. Taylor,et al.  A cluster analysis approach for the characterization of dynamic PET data , 1996 .

[24]  M. Daube-Witherspoon,et al.  Treatment of axial data in three-dimensional PET. , 1987, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[25]  William Wijns,et al.  Relation Between Myocardial Blood Flow and the Severity of Coronary-Artery Stenosis , 1995 .

[26]  Adriaan A. Lammertsma,et al.  Noninvasive Quantification of Regional Myocardial Blood Flow in Coronary Artery Disease With Oxygen‐15–Labeled Carbon Dioxide Inhalation and Positron Emission Tomography , 1991, Circulation.

[27]  A. Luxen,et al.  Direct Comparison of [13N]Ammonia and [150]Water Estimates of Perfusion With Quantification of Regional Myocardial Blood Flow by Microspheres , 1993, Circulation.

[28]  J B Bassingthwaighte,et al.  Molecular and particulate depositions for regional myocardial flows in sheep. , 1990, Circulation research.

[29]  P. Camici,et al.  Role of positron emission tomography in the investigation of human coronary circulatory function. , 1994, Cardiovascular research.

[30]  M. Defrise,et al.  Three dimensional reconstruction of PET data from a multi-ring camera , 1989 .

[31]  T J Spinks,et al.  Generation of myocardial factor images directly from the dynamic oxygen-15-water scan without use of an oxygen-15-carbon monoxide blood-pool scan. , 1998, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[32]  Keishi Kitamura,et al.  Histochemical Correlates of 15O-Water–Perfusable Tissue Fraction in Experimental Canine Studies of Old Myocardial Infarction , 2000 .

[33]  P. Cutler,et al.  Local threshold for segmented attenuation correction of PET imaging of the thorax , 1993 .

[34]  A. Lammertsma,et al.  Use of the left ventricular time-activity curve as a noninvasive input function in dynamic oxygen-15-water positron emission tomography. , 1992, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[35]  R. Myers Quantification of brain function using PET , 1996 .