A novel easy-to-use phantom for the determination of MTF in SPECT scanners.

PURPOSE To evaluate modulation transfer function (MTF) in single photon emission computed tomography (SPECT) systems using the line spread function (LSF) method and a novel flood source which can be easily fabricated with materials accessible in hospital facilities. METHODS A Tc-99m-based flood source (E(γ) = 140 keV) consisting of a radiopharmaceutical bound to the grains of a radiographic film was prepared in laboratory. Various films and radiopharmaceuticals were examined, in order to obtain a thin homogenous and reproducible flood source. The source showing best uniformity and reproducibility was placed between two PMMA blocks and images were obtained by using the brain tomographic acquisition protocol (brain) and the myocardial perfusion tomographic acquisition protocol (heart). MTF was evaluated by determining the LSF for various reconstruction methods and filters. MTF calculation was obtained by the utilization of a custom made software in which a method similar to the one proposed by Boone [Med. Phys. 28, 356-360 (2001)] was implemented. All imaging experiments were performed in a Siemens e-Cam γ-camera. Furthermore, MTF was assessed through the point spread function (PSF) following conventional methods. RESULTS The optimum homogeneity was obtained by immersing an Agfa MammoRay HDR Medical x-ray film in a solution of dithiothreitol (DTT, 10(-3) M)/Tc-99m(III)-DMSA (DMSA: trivalent technetium-99m-dimercapto-succinic acid, 40 mCi/40 ml) for 30 min in the dark. These films exhibited better uniformity (CV < 1.9%). Higher MTF values were obtained for the brain scan protocol with iterative 3D with eight iterations reconstruction method. MTF of the brain protocol was in all cases better than the heart protocol. MTFs derived from LSF were more precise compared with those obtained from PSF since their reproducibility was better in all cases, providing a mean standard deviation of 0.0065, in contrary to the PSF method which gave 0.0348. CONCLUSIONS The method presented here is novel and easy to implement, requiring materials commonly found in clinical practice. Furthermore, this technique which is based on the LSF method reduces measurement noise levels due to the larger amount of data averaging than in the conventional PSF method. Furthermore, MTF can be assessed easily, in three dimensions (3D), by placing the flood source either in sagittal or coronal direction.

[1]  R. Ning,et al.  Three-dimensional point spread function measurement of cone-beam computed tomography system by iterative edge-blurring algorithm. , 2004, Physics in medicine and biology.

[2]  Toru Matsumoto,et al.  An effective method to verify line and point spread functions measured in computed tomography. , 2006, Medical physics.

[3]  D Babot,et al.  A comparison of the ball, wire, edge, and bar/space pattern techniques for modulation transfer function measurements of linear x-ray detectors. , 1996, Journal of X-ray science and technology.

[4]  M B Williams,et al.  Analysis of the detective quantum efficiency of a developmental detector for digital mammography. , 1999, Medical Physics (Lancaster).

[5]  S. R. Amtey,et al.  Practical considerations in gamma camera line spread function measurement. , 1978, Medical physics.

[6]  Kunio Doi,et al.  A simple method for determining the modulation transfer function in digital radiography , 1992, IEEE Trans. Medical Imaging.

[7]  Willi A Kalender,et al.  Presampling, algorithm factors, and noise: considerations for CT in particular and for medical imaging in general. , 2005, Medical physics.

[8]  H. Genant,et al.  Accuracy of CT-based thickness measurement of thin structures: modeling of limited spatial resolution in all three dimensions. , 2002, Medical physics.

[9]  K. Stierstorfer,et al.  Image reconstruction and image quality evaluation for a 64-slice CT scanner with z-flying focal spot. , 2005, Medical physics.

[10]  Jeffrey A. Fessler,et al.  Optimizing the number of equivalent iterations of 3D OSEM in SPECT reconstruction of I-131 focal activities , 2006 .

[11]  Michael A. King,et al.  Investigation of the stationarity of the modulation transfer function and the scatter fraction in conjugate view SPECT restoration filtering , 1989 .

[12]  A R Benedetto,et al.  Simplified method for determining the modulation transfer function for the scintillation camera. , 1975, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[13]  Salem A. Sassi,et al.  A Comparison Between 3D OSEM and FBP Image Reconstruction Algorithms in SPECT , 2008, World Congress on Engineering.

[14]  J. Ehrhardt,et al.  Modulation transfer function of the EMI CT head scanner. , 1977, Medical physics.

[15]  J M Boone,et al.  Determination of the presampled MTF in computed tomography. , 2001, Medical physics.

[16]  Magnus Båth,et al.  The use of detective quantum efficiency (DQE) in evaluating the performance of gamma camera systems , 2005, Physics in medicine and biology.

[17]  E L Nickoloff,et al.  A simplified approach for modulation transfer function determinations in computed tomography. , 1985, Medical physics.

[18]  G. Dougherty,et al.  Measurement of thickness and density of thin structures by computed tomography: a simulation study. , 1999, Medical physics.

[19]  Timo Väyrynen,et al.  Methods for measuring the modulation transfer function of gamma camera systems , 2004, European Journal of Nuclear Medicine.

[20]  R L Morin,et al.  A practical method to measure the MTF of CT scanners. , 1982, Medical physics.

[21]  Toru Matsumoto,et al.  Image filtering as an alternative to the application of a different reconstruction kernel in CT imaging: feasibility study in lung cancer screening. , 2011, Medical physics.

[22]  W. Kalender,et al.  Accuracy limits for the determination of cortical width and density: the influence of object size and CT imaging parameters. , 1999, Physics in medicine and biology.

[24]  J T Dobbins,et al.  Effects of undersampling on the proper interpretation of modulation transfer function, noise power spectra, and noise equivalent quanta of digital imaging systems. , 1995, Medical physics.

[25]  Marco Brambilla,et al.  Characterization of ordered-subsets expectation maximization with 3d post-reconstruction gauss filtering and comparison with filtered backprojection in99mTc SPECT , 2005, Annals of nuclear medicine.

[26]  Lee-Tzuu Chang,et al.  A Method for Attenuation Correction in Radionuclide Computed Tomography , 1978, IEEE Transactions on Nuclear Science.