Human Imaging With Photon Counting–Based Computed Tomography at Clinical Dose Levels: Contrast-to-Noise Ratio and Cadaver Studies

ObjectivesThe purpose of this work was to measure and compare the iodine contrast-to-noise ratio (CNR) between a commercial energy-integrating detector (EID) computed tomography (CT) system and a photon-counting detector (PCD) CT scanner capable of human imaging at clinical dose rates, as well as to determine clinical feasibility using human cadavers. Materials and MethodsA research dual-source PCD-CT scanner was used, where the “A” tube/detector subsystem used an EID and the “B” tube/detector subsystem used a PCD. Iodine CNR was measured in 4 anthropomorphic phantoms, simulating 4 patient sizes, at 4 tube potential settings. After biospecimen committee approval, PCD scans were performed on a fresh-frozen human head and a whole-body cadaver using clinical dose rates. Scans were repeated using the EID and identical parameters, and qualitative side-by-side comparisons were performed. ResultsFor the same photon fluence, phantom measurements demonstrated a mean increase in CNR of 11%, 23%, 31%, 38% for the PCD system, relative to the EID system, at 80, 100, 120, and 140 kV, respectively. Photon-counting detector CT additionally provided energy-selective imaging, where low- and high-energy images reflected the energy dependence of the iodine signal. Photon-counting detector images of cadaveric anatomy demonstrated decreased beam hardening and calcium blooming in the high-energy bin images and increased contrast in the low-energy bins images relative to the EID images. Threshold-based PCD images were qualitatively deemed equivalent in other aspects. ConclusionsThe evaluated research PCD-CT system was capable of clinical levels of image quality at clinical dose rates. It further provided improved CNR relative to state-of-the-art EID-CT. The energy-selective bin images provide further opportunity for dual-energy and multienergy analyses.

[1]  E. Ritenour,et al.  Medical Imaging Physics , 1992 .

[2]  M Danielsson,et al.  Detective quantum efficiency dependence on x-ray energy weighting in mammography. , 1999, Medical physics.

[3]  Robert E. Melen,et al.  CdZnTe detector array for a scanning-beam digital x-ray system , 1999, Medical Imaging.

[4]  D. Maeding,et al.  Preliminary results obtained from a novel CdZnTe pad detector and readout ASIC developed for an automatic baggage inspection system , 2000, 2000 IEEE Nuclear Science Symposium. Conference Record (Cat. No.00CH37149).

[5]  Harold L. Kundel,et al.  Handbook of Medical Imaging, Volume 1. Physics and Psychophysics , 2000 .

[6]  Christin Wirth The Essential Physics of Medical Imaging , 2003, European Journal of Nuclear Medicine and Molecular Imaging.

[7]  K. Stierstorfer,et al.  Weighted FBP--a simple approximate 3D FBP algorithm for multislice spiral CT with good dose usage for arbitrary pitch. , 2004, Physics in medicine and biology.

[8]  G. Anton,et al.  The influence of energy weighting on X-ray imaging quality , 2004 .

[9]  E. Roessl,et al.  K-edge imaging in x-ray computed tomography using multi-bin photon counting detectors , 2007, Physics in medicine and biology.

[10]  P. Shikhaliev Computed tomography with energy-resolved detection: a feasibility study , 2008, Physics in medicine and biology.

[11]  D. Bale,et al.  CdZnTe Semiconductor Detectors for Spectroscopic X-ray Imaging , 2008, IEEE Transactions on Nuclear Science.

[12]  J. Schlomka,et al.  Multienergy photon-counting K-edge imaging: potential for improved luminal depiction in vascular imaging. , 2008, Radiology.

[13]  P.J. Bones,et al.  Medipix imaging - evaluation of datasets with PCA , 2008, 2008 23rd International Conference Image and Vision Computing New Zealand.

[14]  P. Shikhaliev Energy-resolved computed tomography: first experimental results , 2008, Physics in medicine and biology.

[15]  J. Schlomka,et al.  Experimental feasibility of multi-energy photon-counting K-edge imaging in pre-clinical computed tomography , 2008, Physics in medicine and biology.

[16]  J. Iwanczyk,et al.  Photon Counting Energy Dispersive Detector Arrays for X-ray Imaging , 2007, IEEE Transactions on Nuclear Science.

[17]  Taly Gilat Schmidt,et al.  Optimal "image-based" weighting for energy-resolved CT. , 2009, Medical physics.

[18]  K. Stierstorfer,et al.  Evaluation of z-axis resolution and image noise for nonconstant velocity spiral CT data reconstructed using a weighted 3D filtered backprojection (WFBP) reconstruction algorithm. , 2010, Medical physics.

[19]  J. Schlomka,et al.  Computed tomography in color: NanoK-enhanced spectral CT molecular imaging. , 2010, Angewandte Chemie.

[20]  Edgar Kraft,et al.  A research prototype system for quantum-counting clinical CT , 2010, Medical Imaging.

[21]  Marios E Myronakis,et al.  Monte Carlo investigation of charge-transport effects on energy resolution and detection efficiency of pixelated CZT detectors for SPECT/PET applications. , 2010, Medical physics.

[22]  Katsuyuki Taguchi,et al.  Modeling the performance of a photon counting x-ray detector for CT: energy response and pulse pileup effects. , 2011, Medical physics.

[23]  T. Flohr,et al.  Spectral Computed Tomography , 2012 .

[24]  Uwe Oelfke,et al.  Imaging properties of small-pixel spectroscopic x-ray detectors based on cadmium telluride sensors , 2012, Physics in medicine and biology.

[25]  Thorsten M. Buzug,et al.  Investigation of ultra low-dose scans in the context of quantum-counting clinical CT , 2012, Medical Imaging.

[26]  Katsuyuki Taguchi,et al.  Achieving routine submillisievert CT scanning: report from the summit on management of radiation dose in CT. , 2012, Radiology.

[27]  Thilo Hannemann,et al.  First results from a hybrid prototype CT scanner for exploring benefits of quantum-counting in clinical CT , 2012, Medical Imaging.

[28]  R. Alvarez Dimensionality and noise in energy selective x-ray imaging. , 2013, Medical physics.

[29]  K. Taguchi,et al.  Vision 20/20: Single photon counting x-ray detectors in medical imaging. , 2013, Medical physics.

[30]  Karl Stierstorfer,et al.  Multi-energy performance of a research prototype CT scanner with small-pixel counting detector , 2013, Medical Imaging.

[31]  Karl Stierstorfer,et al.  Photon counting CT at elevated X-ray tube currents: contrast stability, image noise and multi-energy performance , 2014, Medical Imaging.

[32]  A. Thran,et al.  Quantitative Spectral K-Edge Imaging in Preclinical Photon-Counting X-Ray Computed Tomography , 2015, Investigative radiology.

[33]  X Duan,et al.  Initial results from a prototype whole-body photon-counting computed tomography system , 2015, Medical Imaging.

[34]  C. McCollough,et al.  Dual- and Multi-Energy CT: Principles, Technical Approaches, and Clinical Applications. , 2015, Radiology.