Commissioning and initial stereotactic ablative radiotherapy experience with Vero
暂无分享,去创建一个
Chuxiong Ding | John Yordy | Timothy D. Solberg | Paul M. Medin | Ezequiel Ramirez | Ryan D. Foster | T. Solberg | C. Ding | P. Medin | E. Ramirez | R. Foster | J. Yordy
[1] M. Stasi,et al. [Dosimetric characterization of a multileaf collimator]. , 1999, La Radiologia medica.
[2] Robert D Timmerman,et al. An overview of hypofractionation and introduction to this issue of seminars in radiation oncology. , 2008, Seminars in radiation oncology.
[3] Fang-Fang Yin,et al. Task Group 142 report: quality assurance of medical accelerators. , 2009, Medical physics.
[4] Lei Xing,et al. An end-to-end examination of geometric accuracy of IGRT using a new digital accelerator equipped with onboard imaging system , 2012, Physics in medicine and biology.
[5] Andrea Bezjak,et al. Stereotactic body radiation therapy for inoperable early stage lung cancer. , 2010, JAMA.
[6] F. Lohr,et al. Stereotactic, single-dose irradiation of lung tumors: a comparison of absolute dose and dose distribution between pencil beam and Monte Carlo algorithms based on actual patient CT scans. , 2010, International journal of radiation oncology, biology, physics.
[7] Hiroshi Nakayama,et al. Initial validations for pursuing irradiation using a gimbals tracking system , 2017 .
[8] Dirk Verellen,et al. Quality assurance of a system for improved target localization and patient set-up that combines real-time infrared tracking and stereoscopic X-ray imaging. , 2003, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.
[9] M. Yamada,et al. Optimization of the x-ray monitoring angle for creating a correlation model between internal and external respiratory signals. , 2012, Medical physics.
[10] I. Chetty,et al. Dosimetric verification and clinical evaluation of a new commercially available Monte Carlo-based dose algorithm for application in stereotactic body radiation therapy (SBRT) treatment planning , 2010, Physics in medicine and biology.
[11] Hugo Palmans,et al. IPEM report 103: Small field MV photon dosimetry , 2010 .
[12] Y. Yamada,et al. Correlation of local failure with measures of dose insufficiency in the high-dose single-fraction treatment of bony metastases. , 2008, International journal of radiation oncology, biology, physics.
[13] F. Nüsslin,et al. A virtual photon energy fluence model for Monte Carlo dose calculation. , 2003, Medical physics.
[14] J J DeMarco,et al. Monte Carlo treatment planning for stereotactic radiosurgery. , 1998, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.
[15] Timothy D. Solberg,et al. Quality and safety considerations in stereotactic radiosurgery and stereotactic body radiation therapy: Executive summary , 2012, Practical radiation oncology.
[16] I. Kawrakow,et al. Investigation of variance reduction techniques for Monte Carlo photon dose calculation using XVMC , 2000, Physics in medicine and biology.
[17] T D Solberg,et al. Implications of tissue heterogeneity for radiosurgery in head and neck tumors. , 1995, International journal of radiation oncology, biology, physics.
[18] B. Bjärngard,et al. Doses on the central axes of narrow 6-MV x-ray beams. , 1990, Medical physics.
[19] Hui Yan,et al. Dosimetric characteristics of novalis Tx system with high definition multileaf collimator. , 2008, Medical physics.
[20] T. Solberg,et al. Quality assurance of immobilization and target localization systems for frameless stereotactic cranial and extracranial hypofractionated radiotherapy. , 2008, International journal of radiation oncology, biology, physics.
[21] Lech Papiez,et al. Stereotactic body radiation therapy for early-stage non-small-cell lung carcinoma: four-year results of a prospective phase II study. , 2009, International journal of radiation oncology, biology, physics.
[22] P. Gerszten,et al. Radiotherapy and Radiosurgery for Metastatic Spine Disease: What Are the Options, Indications, and Outcomes? , 2009, Spine.
[23] Fang-Fang Yin,et al. Use of the BrainLAB ExacTrac X-Ray 6D system in image-guided radiotherapy. , 2008, Medical dosimetry : official journal of the American Association of Medical Dosimetrists.
[24] Cristina Garibaldi,et al. Computer-aided analysis of star shot films for high-accuracy radiation therapy treatment units , 2012, Physics in medicine and biology.
[25] Masahiro Hiraoka,et al. Development of an ultrasmall C-band linear accelerator guide for a four-dimensional image-guided radiotherapy system with a gimbaled x-ray head. , 2007, Medical physics.
[26] Zheng Chang,et al. ExacTrac X-ray 6 degree-of-freedom image-guidance for intracranial non-invasive stereotactic radiotherapy: comparison with kilo-voltage cone-beam CT. , 2009, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.
[27] F. Yin,et al. Dosimetric characteristics of Novalis shaped beam surgery unit. , 2002, Medical physics.
[28] Hiroki Shirato,et al. Stereotactic body radiotherapy (SBRT) for operable stage I non-small-cell lung cancer: can SBRT be comparable to surgery? , 2011, International journal of radiation oncology, biology, physics.
[29] C. Cleeland,et al. Phase I/II study of stereotactic body radiotherapy for spinal metastasis and its pattern of failure. , 2007, Journal of neurosurgery. Spine.
[30] M. Schell,et al. A prospective pilot study of curative‐intent stereotactic body radiation therapy in patients with 5 or fewer oligometastatic lesions , 2008, Cancer.
[31] David S Followill,et al. Algorithms used in heterogeneous dose calculations show systematic differences as measured with the Radiological Physics Center's anthropomorphic thorax phantom used for RTOG credentialing. , 2012, International journal of radiation oncology, biology, physics.
[32] I. Chetty,et al. Image-guided localization accuracy of stereoscopic planar and volumetric imaging methods for stereotactic radiation surgery and stereotactic body radiation therapy: a phantom study. , 2011, International journal of radiation oncology, biology, physics.
[33] Jinkoo Kim,et al. Clinical commissioning and use of the Novalis Tx linear accelerator for SRS and SBRT , 2012, Journal of applied clinical medical physics.
[34] L. Gaspar,et al. Multi-institutional phase I/II trial of stereotactic body radiation therapy for lung metastases. , 2009, Journal of clinical oncology : official journal of the American Society of Clinical Oncology.
[35] P. Xia,et al. Dose calculation differences between Monte Carlo and pencil beam depend on the tumor locations and volumes for lung stereotactic body radiation therapy , 2013, Journal of applied clinical medical physics.
[36] Samuel Ryu,et al. American Society for Therapeutic Radiology and Oncology (ASTRO) and American College of Radiology (ACR) practice guideline for the performance of stereotactic body radiation therapy. , 2004, International journal of radiation oncology, biology, physics.
[37] Masahiro Hiraoka,et al. Development of a four-dimensional image-guided radiotherapy system with a gimbaled X-ray head. , 2006, International journal of radiation oncology, biology, physics.
[38] G. Ding,et al. Commissioning stereotactic radiosurgery beams using both experimental and theoretical methods , 2006, Physics in medicine and biology.
[39] Thierry Gevaert,et al. Geometric accuracy of a novel gimbals based radiation therapy tumor tracking system. , 2011, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.
[40] Kevin W Eliceiri,et al. NIH Image to ImageJ: 25 years of image analysis , 2012, Nature Methods.
[41] M. Schell,et al. Stereotactic body radiation therapy: the report of AAPM Task Group 101. , 2010, Medical physics.
[42] M. Fippel. Fast Monte Carlo dose calculation for photon beams based on the VMC electron algorithm. , 1999, Medical physics.
[43] T Solberg,et al. Commissioning of the Varian TrueBeam linear accelerator: a multi-institutional study. , 2013, Medical physics.
[44] T Sasaki,et al. Effect of inhomogeneity in a patient's body on the accuracy of the pencil beam algorithm in comparison to Monte Carlo , 2012, Physics in medicine and biology.
[45] S. Dieterich,et al. Experimental comparison of seven commercial dosimetry diodes for measurement of stereotactic radiosurgery cone factors. , 2011, Medical physics.
[46] Mitsuhiro Nakamura,et al. Dosimetric characterization of a multileaf collimator for a new four-dimensional image-guided radiotherapy system with a gimbaled x-ray head, MHI-TM2000a). , 2010, Medical physics.
[47] Jan Nyman,et al. Outcome in a prospective phase II trial of medically inoperable stage I non-small-cell lung cancer patients treated with stereotactic body radiotherapy. , 2009, Journal of clinical oncology : official journal of the American Society of Clinical Oncology.
[48] T. Zhu,et al. Accelerator beam data commissioning equipment and procedures: Report of the TG-106 of the Therapy Physics Committee of the AAPM. , 2008, Medical physics.
[49] Mitsuhiro Nakamura,et al. Positional accuracy of novel x-ray-image-based dynamic tumor-tracking irradiation using a gimbaled MV x-ray head of a Vero4DRT (MHI-TM2000). , 2012, Medical physics.
[50] W. Franklin,et al. Multi-institutional phase I/II trial of stereotactic body radiation therapy for liver metastases. , 2009, Journal of clinical oncology : official journal of the American Society of Clinical Oncology.
[51] G. Lockwood,et al. Phase I study of individualized stereotactic body radiotherapy of liver metastases. , 2009, Journal of clinical oncology : official journal of the American Society of Clinical Oncology.
[52] Thierry Gevaert,et al. Clinical evaluation of a robotic 6-degree of freedom treatment couch for frameless radiosurgery. , 2012, International journal of radiation oncology, biology, physics.