Precise CD4 T-cell counting using red diode laser excitation: for richer, for poorer.

BACKGROUND Measuring CD4 T-cell counts at low cost is relevant in dealing with the human immunodeficiency virus (HIV) epidemic throughout the developing world. The recently introduced novel concepts in gating strategies and sample stabilization facilitate affordable immunophenotyping by flow cytometry. However, the impact of these developments is still limited by the high cost of currently available flow cytometers. METHODS Diode lasers emitting 10-15 mW at 635 nm are one-tenth the size and cost and require one thousandth the power of an equivalent 488-nm argon ion laser. We used the available 635-nm diode-based flow cytometers, including PA-II, Luminex 100, SuperMot, and FACSCalibur, to investigate whether these instruments can generate reliable CD4 counts when used with allophycocyanin (APC) and cyanin-5 (Cy5)-labeled CD4 antibodies. RESULTS We document the feasibility of obtaining leucocyte differential counts using orthogonal side scatter (SSC) without the need for forward scatter (FSC). Accurate CD4% values among lymphocytes and leucocytes can be obtained by primary CD4 gating using a single CD4 monoclonal antibody conjugated to APC or Cy5. Double immunofluorescence (IF) staining with CD4-APC (FL1) and CD45-APC-Cy7 (FL2) introduces pan-leucogating for a convenient assessment of absolute CD4 counts on double platforms. We demonstrate that small flow cytometers with laser diodes are capable of delivering absolute CD4 T-cell counts with a precision similar to the performance of the current state-of-the-art single-platform instruments (e.g., the CytoronAbsolute; R(2) = 0.961). In this respect, they appear to be superior to the nonflow CD4 counting techniques. CONCLUSIONS Accurate CD4 counts can be generated at minimal cost on red diode laser-operated flow cytometers, retaining the potential for high throughput capacity without compromising precision. With further improvements in volumetric technology and clinical software, these cytometers may develop into a new generation of inexpensive battery-operated laboratory hardware that combines cellular phenotyping with bead-based multiplexing immunoassays for (HIV) serology.

[1]  D Barnett,et al.  Evaluation of a novel stable whole blood quality control material for lymphocyte subset analysis: results from the UK NEQAS immune monitoring scheme. , 1996, Cytometry.

[2]  Andrew N. Phillips,et al.  When to start highly active antiretroviral therapy in chronically HIV-infected patients: evidence from the ICONA study , 2001, AIDS.

[3]  M. O'gorman,et al.  Inter- and intrainstitutional evaluation of automated volumetric capillary cytometry for the quantitation of CD4- and CD8-positive T lymphocytes in the peripheral blood of persons infected with human immunodeficiency virus. Site Investigators and the NIAID New CD4 Technologies Focus Group , 1997, Clinical and diagnostic laboratory immunology.

[4]  C. Sabin,et al.  CD4 lymphocyte count as a predictor of the duration of highly active antiretroviral therapy-induced suppression of human immunodeficiency virus load. , 1999, The Journal of infectious diseases.

[5]  George Janossy,et al.  Multiplexed immunoassays by flow cytometry for diagnosis and surveillance of infectious diseases in resource-poor settings. , 2002, The Lancet. Infectious diseases.

[6]  M R Loken,et al.  Establishing optimal lymphocyte gates for immunophenotyping by flow cytometry. , 1990, Cytometry.

[7]  T. Strom,et al.  Immunofluorescence measurement in a flow cytometer using low-power helium-neon laser excitation. , 1983, Cytometry.

[8]  J P Nolan,et al.  Suspension array technology: new tools for gene and protein analysis. , 2001, Cellular and molecular biology.

[9]  George Janossy,et al.  Affordable CD4(+) T-cell counts on 'single-platform' flow cytometers I. Primary CD4 gating. , 2000 .

[10]  Visible diode lasers can be used for flow cytometric immunofluorescence and DNA analysis. , 1994, Cytometry.

[11]  G. Fleuren,et al.  Four-color multiparameter DNA flow cytometric method to study phenotypic intratumor heterogeneity in cervical cancer. , 2000, Cytometry.

[12]  Howard M. Shapiro,et al.  Practical Flow Cytometry , 1985 .

[13]  W. Guilford,et al.  Cell Cycle Redistribution of Cultured Cells After Treatment With Chemical Radiation Protectors , 1988, Cell and tissue kinetics.

[14]  A. Beckett,et al.  AKUFO AND IBARAPA. , 1965, Lancet.

[15]  M D Kazatchkine,et al.  Comparative assessment of five alternative methods for CD4+ T-lymphocyte enumeration for implementation in developing countries. , 2001, Journal of acquired immune deficiency syndromes.

[16]  M. Roederer,et al.  11-color, 13-parameter flow cytometry: Identification of human naive T cells by phenotype, function, and T-cell receptor diversity , 2001, Nature Medicine.

[17]  C. S. Scott,et al.  Microvolume fluorimetry for the determination of absolute CD4 and CD8 lymphocyte counts in patients with HIV: a comparative evaluation. , 1999, Clinical and laboratory haematology.

[18]  K. Reimann,et al.  Multisite Comparison of CD4 and CD8 T-Lymphocyte Counting by Single- versus Multiple-Platform Methodologies: Evaluation of Beckman Coulter Flow-Count Fluorospheres and the tetraONE System , 2000, Clinical Diagnostic Laboratory Immunology.

[19]  M. O'gorman,et al.  Evaluation of TruCount Absolute-Count Tubes for Determining CD4 and CD8 Cell Numbers in Human Immunodeficiency Virus-Positive Adults , 2000, Clinical Diagnostic Laboratory Immunology.

[20]  G. Janossy,et al.  Affordable CD4+ T cell counts by flow cytometry. II. The use of fixed whole blood in resource-poor settings. , 2001, Journal of immunological methods.

[21]  K. Pallangyo,et al.  Evaluation of the FACScount, TRAx CD4 and Dynabeads methods for CD4 lymphocyte determination. , 1996, Journal of immunological methods.

[22]  M. O'gorman,et al.  Adoption of Single-Platform Technologies for Enumeration of Absolute T-Lymphocyte Subsets in Peripheral Blood , 2000, Clinical Diagnostic Laboratory Immunology.

[23]  H M Shapiro,et al.  A flow cytometer designed for fluorescence calibration. , 1998, Cytometry.

[24]  M. Knight,et al.  Immunologic differentiation of absolute lymphocyte count with an integrated flow cytometric system: a new concept for absolute T cell subset determinations. , 1995, Cytometry.

[25]  D K Glencross,et al.  CD4+ T cell enumeration in HIV infection with limited resources. , 1999, Journal of immunological methods.

[26]  White blood cell differentiation using a solid state flow cytometer. , 1993, Cytometry.

[27]  Lesley E Scott,et al.  CD45-assisted PanLeucogating for accurate, cost-effective dual-platform CD4+ T-cell enumeration. , 2002, Cytometry.

[28]  B G de Grooth,et al.  Cell analysis system based on immunomagnetic cell selection and alignment followed by immunofluorescent analysis using compact disk technologies. , 2001, Cytometry.

[29]  T. Denny,et al.  Evaluation of a method for counting absolute numbers of cells with a flow cytometer , 1997, Clinical and diagnostic laboratory immunology.

[30]  A. Badley,et al.  Selection of lymphocyte gating protocol has an impact on the level of reliability of T-cell subsets in aging specimens. , 2002, Cytometry.

[31]  D. Altman,et al.  STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT , 1986, The Lancet.