On‐line flow quantification by low‐resolution phase‐contrast MR imaging and model‐based postprocessing

Over the past decade, magnetic resonance (MR) imaging has been developed toward a tool for guiding and evaluating diagnostic and therapeutic interventions. Within the field of vascular MR‐guided interventions, MR has potential for providing on‐line monitoring of the blood volume flow rate, which is relevant during procedures such as balloon angioplasty and stent placement. We recently reported a hardware and software environment for enabling flow quantification every 8 seconds using nontriggered phase‐contrast imaging. In the present study, the objective was to increase temporal resolution further to one evaluation per 4 seconds. We achieve this by lowering spatial resolution to 3 pixels per lumen diameter. The accuracy of the measurements is preserved by applying model‐based postprocessing for quantification of the volume flow rate. Phantom and volunteer studies are presented, demonstrating the accuracy of the model‐driven approach for the applied short acquisitions. The capabilities of the presented approach are illustrated by the results of several hypercapnia experiments and carotid compression tests performed on healthy volunteers. J. Magn. Reson. Imaging 2000;12:623–631. © 2000 Wiley‐Liss, Inc.

[1]  R van der Weide,et al.  Facilities for monitoring blood flow during MR‐guided diagnostic and therapeutic interventions , 1999, Journal of magnetic resonance imaging : JMRI.

[2]  R. Wendt,et al.  Nuclear magnetic resonance velocity spectra of steady flow , 1990, Magnetic resonance in medicine.

[3]  D L Parker,et al.  MR angiography by multiple thin slab 3D acquisition , 1991, Magnetic resonance in medicine.

[4]  R W Günther,et al.  Catheter visualization using locally induced, actively controlled field inhomogeneities , 1997, Magnetic resonance in medicine.

[5]  J Hennig,et al.  Quantitative flow measurement with the fast Fourier flow technique. , 1988, Radiology.

[6]  M. Viergever,et al.  MR phase‐contrast flow measurement with limited spatial resolution in small vessels: Value of model‐based image analysis , 1999, Magnetic resonance in medicine.

[7]  Max A. Viergever,et al.  Image guidance of endovascular interventions on a clinical MR scanner , 1998, IEEE Transactions on Medical Imaging.

[8]  F A Jolesz,et al.  Interventional use of magnetic resonance imaging. , 1994, Magnetic resonance quarterly.

[9]  S. Souza,et al.  Quantitative measurement of blood flow using cylindrically localized fourier velocity encoding , 1991, Magnetic resonance in medicine.

[10]  M Kouwenhoven,et al.  Nontriggered magnetic resonance velocity measurement of the time‐average of pulsatile velocity , 1993, Magnetic resonance in medicine.

[11]  阿部 俊昭,et al.  Interventional MRI を用いた定位的脳手術 , 2000 .

[12]  Mitsuaki Arakawa,et al.  Magnetic Resonance Imaging the Velocity Vector Components of Fluid Flow , 1985, Magnetic resonance in medicine.

[13]  S. Koskinen [Interventional magnetic resonance imaging]. , 1999, Duodecim; laaketieteellinen aikakauskirja.

[14]  D. Parker,et al.  Accuracy of phase‐contrast flow measurements in the presence of partial‐volume effects , 1993, Journal of magnetic resonance imaging : JMRI.

[15]  C L Dumoulin,et al.  Intravascular MR tracking catheter: preliminary experimental evaluation. , 1995, AJR. American journal of roentgenology.

[16]  R. Herfkens,et al.  Phase contrast cine magnetic resonance imaging. , 1991, Magnetic resonance quarterly.

[17]  R L Ehman,et al.  Analysis of systematic and random error in MR volumetric flow measurements , 1993, Magnetic resonance in medicine.

[18]  S. Souza,et al.  Real‐time position monitoring of invasive devices using magnetic resonance , 1993, Magnetic resonance in medicine.

[19]  M A Viergever,et al.  Visualization of dedicated catheters using fast scanning techniques with potential for MR‐guided vascular interventions , 1996, Magnetic resonance in medicine.

[20]  J. R. Singer,et al.  Blood Flow Rates by Nuclear Magnetic Resonance Measurements , 1959, Science.

[21]  W. Gedroyc Interventional magnetic resonance imaging. , 1997, BJU international.