The Effect of Targeting Scheme on Tidal Volume Delivery During Volume Control Mechanical Ventilation

BACKGROUND: Technological advances have increased ventilator mode complexity and risk of operator error. OBJECTIVE: To compare differences in volume control (VC) ventilation with set-point and dual targeting. Two hypotheses were tested: tidal volume (VT) delivery is different with VC using set-point versus dual targeting during active versus passive breathing; VC with dual targeting delivers VT similar to pressure support ventilation (PS) with active breathing. METHODS: The Ingmar Medical ASL 5000 lung model simulated pulmonary mechanics of an adult patient with ARDS during active and passive ventilation. Resistance was standardized at 10 cm H2O/L/s and compliance at 32 mL/cm H2O. Active breathing was simulated by setting the frequency (f) = 26 breaths/min, and adjusting the muscle pressure (Pmus) to produce a VT of 384 mL. VC was initiated with the Puritan Bennett 840 (set-point targeting) and the Servo-i (dual targeting) at VT = 430 mL, mandatory f = 15 breaths/min, and PEEP = 10 cm H2O. During PS, cycle threshold was set to 30% and peak inspiratory pressure adjusted to produce a VT similar to that delivered during VC. Expiratory VT was collected on 10 consecutive breaths during active and passive breathing with VC and PS. Mean VT differences (active vs passive model) were compared using analysis of variance. Statistical significance was established at P < .05. RESULTS: The mean ± SD VT difference varied with targeting schemes: VC set-point = 37.3 ± 3.5 mL, VC-dual = 77.1 ± 3.3 mL, and PS = 406.1 ± 1.5 mL (P < .001). Auto-triggering occurred during VC set-point with the active model. CONCLUSIONS: Dual targeting during VC allows increased VT, compared to set-point, but not as much as PS.

[1]  C. Sassoon Triggering of the Ventilator in Patient-Ventilator Interactions , 2011, Respiratory Care.

[2]  S. Epstein How Often Does Patient-Ventilator Asynchrony Occur and What Are the Consequences? , 2011, Respiratory Care.

[3]  Robert L Chatburn,et al.  Closed-Loop Control of Mechanical Ventilation: Description and Classification of Targeting Schemes , 2011, Respiratory Care.

[4]  R. Chatburn,et al.  Documentation issues for mechanical ventilation in pressure-control modes. , 2010, Respiratory care.

[5]  P. Pelosi,et al.  Pressure Support Ventilation and Biphasic Positive Airway Pressure Improve Oxygenation by Redistribution of Pulmonary Blood Flow , 2009, Anesthesia and analgesia.

[6]  S. Brusaferro,et al.  Errors and omissions in hospital prescriptions: a survey of prescription writing in a hospital , 2009, BMC clinical pharmacology.

[7]  L. Rose,et al.  Airway pressure release ventilation and biphasic positive airway pressure: a systematic review of definitional criteria , 2008, Intensive Care Medicine.

[8]  Kaija Saranto,et al.  Definition, structure, content, use and impacts of electronic health records: A review of the research literature , 2008, Int. J. Medical Informatics.

[9]  Robert L Chatburn,et al.  Classification of ventilator modes: update and proposal for implementation. , 2007, Respiratory care.

[10]  Martin J. Tobin,et al.  Principles and Practice of Mechanical Ventilation, 2nd Edition , 2006 .

[11]  R. Kallet,et al.  Effects of tidal volume on work of breathing during lung-protective ventilation in patients with acute lung injury and acute respiratory distress syndrome* , 2006, Critical care medicine.

[12]  L. Ho,et al.  The development and validation of a knowledge, attitude and behaviour questionnaire to assess undergraduate evidence‐based practice teaching and learning , 2003, Medical education.

[13]  S. P. Pilbeam,et al.  Mosby's Respiratory Care Equipment , 1999 .

[14]  Z. Karakurt,et al.  Pressure vs. volume control in COPD patients intubated due to ARF: a case-control study. , 2009, Tuberkuloz ve toraks.

[15]  Sanborn Wg Microprocessor-based mechanical ventilation , 1993 .

[16]  W. Sanborn Microprocessor-based mechanical ventilation. , 1993, Respiratory care.