PURPOSE
Proof of principle study of the use of a CMOS active pixel sensor (APS) in producing proton radiographic images using the proton beam at the Massachusetts General Hospital (MGH).
METHODS
A CMOS APS, previously tested for use in s-ray radiation therapy applications, was used for proton beam radiographic imaging at the MGH. Two different setups were used as a proof of principle that CMOS can be used as proton imaging device: (i) a pen with two metal screws to assess spatial resolution of the CMOS and (ii) a phantom with lung tissue, bone tissue, and water to assess tissue contrast of the CMOS. The sensor was then traversed by a double scattered monoenergetic proton beam at 117 MeV, and the energy deposition inside the detector was recorded to assess its energy response. Conventional x-ray images with similar setup at voltages of 70 kVp and proton images using commercial Gafchromic EBT 2 and Kodak X-Omat V films were also taken for comparison purposes.
RESULTS
Images were successfully acquired and compared to x-ray kVp and proton EBT2/X-Omat film images. The spatial resolution of the CMOS detector image is subjectively comparable to the EBT2 and Kodak X-Omat V film images obtained at the same object-detector distance. X-rays have apparent higher spatial resolution than the CMOS. However, further studies with different commercial films using proton beam irradiation demonstrate that the distance of the detector to the object is important to the amount of proton scatter contributing to the proton image. Proton images obtained with films at different distances from the source indicate that proton scatter significantly affects the CMOS image quality.
CONCLUSION
Proton radiographic images were successfully acquired at MGH using a CMOS active pixel sensor detector. The CMOS demonstrated spatial resolution subjectively comparable to films at the same object-detector distance. Further work will be done in order to establish the spatial and energy resolution of the CMOS detector for protons. The development and use of CMOS in proton radiography could allow in vivo proton range checks, patient setup QA, and real-time tumor tracking.
[1]
H. Sadrozinski,et al.
Initial studies on proton computed tomography using a silicon strip detector telescope
,
2003
.
[2]
J. de Boer,et al.
A detector system for proton radiography on the gantry of the Paul-Scherrer-Institute ☆
,
1999
.
[3]
J. Molnár,et al.
Visualization of neutron and proton induced particle production in a CMOS image sensor
,
2003
.
[4]
Barbara Kaser-Hotz,et al.
First proton radiography of an animal patient.
,
2004,
Medical physics.
[5]
Cinzia Talamonti,et al.
Proton radiography for clinical applications
,
2010
.
[6]
V. Sipala,et al.
A proton imaging device: Design and status of realization
,
2010
.
[7]
E Pedroni,et al.
Proton radiography as a tool for quality control in proton therapy.
,
1995,
Medical physics.
[8]
J. Molnár,et al.
Proton therapy beam dosimetry with silicon CMOS image sensors
,
2004
.
[9]
F. Sauli.
GEM: A new concept for electron amplification in gas detectors
,
1997
.