Nonrigid Motion Modeling of the Liver From 3-D Undersampled Self-Gated Golden-Radial Phase Encoded MRI

Magnetic resonance imaging (MRI) has been commonly used for guiding and planning image guided interventions since it provides excellent soft tissue visualization of anatomy and allows motion modeling to predict the position of target tissues during the procedure. However, MRI-based motion modeling remains challenging due to the difficulty of acquiring multiple motion-free 3-D respiratory phases with adequate contrast and spatial resolution. Here, we propose a novel retrospective respiratory gating scheme from a 3-D undersampled high-resolution MRI acquisition combined with fast and robust image registrations to model the nonrigid deformation of the liver. The acquisition takes advantage of the recently introduced golden-radial phase encoding (G-RPE) trajectory. G-RPE is self-gated, i.e., the respiratory signal can be derived from the acquired data itself, and allows retrospective reconstructions of multiple respiratory phases at any arbitrary respiratory position. Nonrigid motion modeling is applied to predict the liver deformation of an average breathing cycle. The proposed approach was validated on 10 healthy volunteers. Motion model accuracy was assessed using similarity-, surface-, and landmark-based validation methods, demonstrating precise model predictions with an overall target registration error of TRE = 1.70 ± 0.94 mm which is within the range of the acquired resolution.

[1]  Christof Baltes,et al.  k‐t BLAST reconstruction from non‐Cartesian k‐t space sampling , 2006, Magnetic resonance in medicine.

[2]  K. Hynynen MRI-guided focused ultrasound treatments. , 2010, Ultrasonics.

[3]  Olaf Dössel,et al.  An Optimal Radial Profile Order Based on the Golden Ratio for Time-Resolved MRI , 2007, IEEE Transactions on Medical Imaging.

[4]  David J. Hawkes,et al.  Effects of Respiratory Liver Motion on Heating for Gated and Model-Based Motion-Compensated High-Intensity Focused Ultrasound Ablation , 2011, MICCAI.

[5]  J. Felmlee,et al.  Adaptive technique for high-definition MR imaging of moving structures. , 1989, Radiology.

[6]  P. Boesiger,et al.  Advances in sensitivity encoding with arbitrary k‐space trajectories , 2001, Magnetic resonance in medicine.

[7]  Jay B. West,et al.  Predicting error in point-based registration , 2001 .

[8]  Vira Chankong,et al.  The impact of respiratory motion and treatment technique on stereotactic body radiation therapy for liver cancer. , 2008, Medical physics.

[9]  K. Cleary,et al.  Image-guided interventions: technology review and clinical applications. , 2010, Annual review of biomedical engineering.

[10]  Debiao Li,et al.  Respiratory self‐gated four‐dimensional coronary MR angiography: A feasibility study , 2008, Magnetic resonance in medicine.

[11]  Tobias Schaeffter,et al.  Hierarchical adaptive local affine registration for fast and robust respiratory motion estimation , 2011, Medical Image Anal..

[12]  Armando Manduca,et al.  Self‐navigated motion correction using moments of spatial projections in radial MRI , 2004 .

[13]  C. Buerger,et al.  D Non-Rigid Motion Modeling of the Liver from Undersampled Golden-Radial Phase Encoding ( G-RPE ) Acquisitions , 2010 .

[14]  David Atkinson,et al.  GPU accelerated iterative SENSE reconstruction of radial phase encoded whole-heart MRI , 2010 .

[15]  Sung Yong Shin,et al.  Scattered Data Interpolation with Multilevel B-Splines , 1997, IEEE Trans. Vis. Comput. Graph..

[16]  D. Hill,et al.  Whole‐heart cine MRI using real‐time respiratory self‐gating , 2007, Magnetic resonance in medicine.

[17]  C. Buerger,et al.  Self-Gating Reconstruction of Multiple Respiratory Phases using Undersampled Golden-Radial Phase Encoding Trajectory , 2009 .

[18]  K. Scheffler,et al.  Principles and applications of balanced SSFP techniques , 2003, European Radiology.

[19]  S. Webb,et al.  Planning lung radiotherapy using 4D CT data and a motion model , 2008, Physics in medicine and biology.

[20]  J. McClelland,et al.  MRI-based measurements of respiratory motion variability and assessment of imaging strategies for radiotherapy planning , 2006, Physics in medicine and biology.

[21]  J. McClelland,et al.  A continuous 4D motion model from multiple respiratory cycles for use in lung radiotherapy. , 2006, Medical physics.

[22]  YingLiang Ma,et al.  A subject-specific technique for respiratory motion correction in image-guided cardiac catheterisation procedures , 2009, Medical Image Anal..

[23]  P. Börnert,et al.  Free-breathing cardiac MR imaging: study of implications of respiratory motion--initial results. , 2001, Radiology.

[24]  W. O'Dell,et al.  Modeling liver motion and deformation during the respiratory cycle using intensity-based nonrigid registration of gated MR images. , 2004, Medical physics.

[25]  David Atkinson,et al.  3D undersampled golden‐radial phase encoding for DCE‐MRA using inherently regularized iterative SENSE , 2010, Magnetic resonance in medicine.

[26]  P. Boesiger,et al.  SENSE: Sensitivity encoding for fast MRI , 1999, Magnetic resonance in medicine.

[27]  S. Riederer,et al.  Navigator-echo-based real-time respiratory gating and triggering for reduction of respiration effects in three-dimensional coronary MR angiography. , 1996, Radiology.

[28]  P. Börnert,et al.  Free‐breathing whole‐heart coronary MRA with 3D radial SSFP and self‐navigated image reconstruction , 2005, Magnetic resonance in medicine.