Device thrombogenicity emulation: a novel methodology for optimizing the thromboresistance of cardiovascular devices.

Thrombotic complications with mechanical circulatory support (MCS) devices remain a critical limitation to their long-term use. Device-induced shear forces may enhance the thrombotic potential of MCS devices through chronic activation of platelets, with a known dose-time response of the platelets to the accumulated stress experienced while flowing through the device-mandating complex, lifelong anticoagulation therapy. To enhance the thromboresistance of these devices for facilitating their long-term use, a universal predictive methodology entitled device thrombogenicity emulation (DTE) was developed. DTE is aimed at optimizing the thromboresistance of any MCS device. It is designed to test device-mediated thrombogenicity, coupled with virtual design modifications, in an iterative approach. This disruptive technology combines in silico numerical simulations with in vitro measurements, by correlating device hemodynamics with platelet activity coagulation markers-before and after iterative design modifications aimed at achieving optimized thrombogenic performance. The design changes are first tested in the numerical domain, and the resultant device conditions are then emulated in a hemodynamic shearing device (HSD) in which platelet activity is measured under device emulated conditions. As such, DTE can be easily incorporated during the device research and development phase-achieving minimization of the device thrombogenicity before prototypes are built and tested thereby reducing the ultimate cost of preclinical and clinical trials. The robust capability of this predictive technology is demonstrated here in various MCS devices. The presented examples indicate the potential of DTE for reducing device thrombogenicity to a level that may obviate or significantly reduce the extent of anticoagulation currently mandated for patients implanted with MCS devices for safe long-term clinical use.

[1]  O.H. Frazier,et al.  The Total Artificial Heart: Where We Stand , 2004, Cardiology.

[2]  Danny Bluestein,et al.  Flow-induced platelet activation and damage accumulation in a mechanical heart valve: numerical studies. , 2007, Artificial organs.

[3]  Clement Kleinstreuer,et al.  Comparison of blood particle deposition models for non-parallel flow domains. , 2003, Journal of biomechanics.

[4]  L H Edmunds,et al.  Is prosthetic valve thrombogenicity related to design or material? , 1996, Texas Heart Institute Journal.

[5]  P. Perrotta,et al.  Platelet activation in a circulating flow loop: combined effects of shear stress and exposure time , 2003, Platelets.

[6]  Shmuel Einav,et al.  Device Thrombogenicity Emulator (DTE)--design optimization methodology for cardiovascular devices: a study in two bileaflet MHV designs. , 2010, Journal of biomechanics.

[7]  L. Harker,et al.  Studies of platelet and fibrinogen kinetics in patients with prosthetic heart valves. , 1970, The New England journal of medicine.

[8]  A. Gosman,et al.  Aspects of computer simulation of liquid-fuelled combustors , 1981 .

[9]  André Garon,et al.  Fast three-dimensional numerical hemolysis approximation. , 2004, Artificial organs.

[10]  D Morley,et al.  Clinical experience with nine patients supported by the continuous flow Debakey VAD. , 2001, The Journal of heart and lung transplantation : the official publication of the International Society for Heart Transplantation.

[11]  T. Feldman,et al.  Percutaneous Valve Repair and Replacement: Challenges Encountered, Challenges Met, Challenges Ahead , 2006, Circulation.

[12]  J R O'Brien,et al.  An independent haemostatic mechanism: shear induced platelet aggregation. , 1990, Advances in experimental medicine and biology.

[13]  J Jesty,et al.  Acetylated prothrombin as a substrate in the measurement of the procoagulant activity of platelets: elimination of the feedback activation of platelets by thrombin. , 1999, Analytical biochemistry.

[14]  A. Fraser,et al.  A New Scoring System to Determine Thromboembolic Risk After Heart Valve Replacement , 2003, Circulation.

[15]  J. D. Hellums,et al.  1993 Whitaker lecture: Biorheology in thrombosis research , 1994, Annals of Biomedical Engineering.

[16]  A. Fogelson,et al.  Surface-mediated control of blood coagulation: the role of binding site densities and platelet deposition. , 2001, Biophysical journal.

[17]  J Fisher,et al.  Three-dimensional study of the effect of two leaflet opening angles on the time-dependent flow through a bileaflet mechanical heart valve. , 1997, Medical engineering & physics.

[18]  M Cerrolaza,et al.  Analysis of 3D transient blood flow passing through an artificial aortic valve by Lattice-Boltzmann methods. , 1998, Journal of biomechanics.

[19]  S. Russell,et al.  Advanced heart failure treated with continuous-flow left ventricular assist device. , 2009, The New England journal of medicine.

[20]  Wei Yin,et al.  Flow induced platelet activation in mechanical heart valves - in vitro studies , 2002, Proceedings of the Second Joint 24th Annual Conference and the Annual Fall Meeting of the Biomedical Engineering Society] [Engineering in Medicine and Biology.

[21]  D J Wheatley,et al.  Detection of microemboli by transcranial Doppler ultrasound. , 1996, Texas Heart Institute journal.

[22]  L Schweiberer,et al.  Computational analysis of platelet adhesion and aggregation under stagnation point flow conditions , 1999, Comput. Biol. Medicine.

[23]  J F Antaki,et al.  Computational fluid dynamics as a development tool for rotary blood pumps. , 2001, Artificial organs.

[24]  Thierry G. Mesana Percutaneous valve technology: an historical opportunity for cardiac surgeons , 2006 .

[25]  A P Yoganathan,et al.  Turbulent shear stress measurements in the vicinity of aortic heart valve prostheses. , 1986, Journal of biomechanics.

[26]  C Bludszuweit,et al.  Model for a general mechanical blood damage prediction. , 1995, Artificial organs.

[27]  Anita Tierney,et al.  Neurological Events During Long-Term Mechanical Circulatory Support for Heart Failure: The Randomized Evaluation of Mechanical Assistance for the Treatment of Congestive Heart Failure (REMATCH) Experience , 2004, Circulation.

[28]  Christoph Benk,et al.  Non-surgical bleeding in patients with ventricular assist devices could be explained by acquired von Willebrand disease. , 2008, European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery.

[29]  S. Einav,et al.  A squeeze flow phenomenon at the closing of a bileaflet mechanical heart valve prosthesis. , 1994, Journal of biomechanics.

[30]  Louis E Samuels,et al.  Swan-Ganz monitoring of the AbioCor artificial heart. , 2005, The Annals of thoracic surgery.

[31]  G. Tenderich,et al.  Device and patient management in a bridge-to-transplant setting. , 2001, The Annals of thoracic surgery.

[32]  R L Kormos,et al.  Assessment of biventricular cardiac function in patients with a Novacor left ventricular assist device. , 1994, The Journal of heart and lung transplantation : the official publication of the International Society for Heart Transplantation.

[33]  H Reul,et al.  In Vitro Thrombogenicity Testing of Artificial Organs , 1998, The International journal of artificial organs.

[34]  D. Wilcox Simulation of Transition with a Two-Equation Turbulence Model , 1994 .

[35]  I. Krukenkamp,et al.  Free emboli formation in the wake of bi-leaflet mechanical heart valves and the effects of implantation techniques. , 2002, Journal of biomechanics.

[36]  Ajit P Yoganathan,et al.  Flow and thrombosis at orifices simulating mechanical heart valve leakage regions. , 2006, Journal of biomechanical engineering.

[37]  J. Tarbell,et al.  Numerical simulation of unsteady laminar flow through a tilting disk heart valve: prediction of vortex shedding. , 1994, Journal of biomechanics.

[38]  M. Gharib,et al.  Vortex shedding as a mechanism for free emboli formation in mechanical heart valves. , 2000, Journal of biomechanical engineering.

[39]  J M Anderson,et al.  Directions for improvement of substitute heart valves: National Heart, Lung, and Blood Institute's Working Group report on heart valves. , 1997, Journal of biomedical materials research.

[40]  K. Riemslagh,et al.  A Three-dimensional Analysis of Flow in the Pivot Regions of an ATS Bileaflet Valve , 1999, The International journal of artificial organs.

[41]  James F. Antaki,et al.  Computational Simulation of Platelet Deposition and Activation: II. Results for Poiseuille Flow over Collagen , 1999, Annals of Biomedical Engineering.

[42]  L V McIntire,et al.  Mathematical analysis of mural thrombogenesis. Concentration profiles of platelet-activating agents and effects of viscous shear flow. , 1989, Biophysical journal.

[43]  Danny Bluestein,et al.  Flow-Induced Platelet Activation in Bileaflet and Monoleaflet Mechanical Heart Valves , 2004, Annals of Biomedical Engineering.

[44]  A P Yoganathan,et al.  Fluid dynamic studies for the year 2000. , 1998, The Journal of heart valve disease.

[45]  James F. Antaki,et al.  Computational Simulation of Platelet Deposition and Activation: I. Model Development and Properties , 1999, Annals of Biomedical Engineering.

[46]  H Schmid-Schönbein,et al.  Towards a concept of thrombosis in accelerated flow: rheology, fluid dynamics, and biochemistry. , 1985, Biorheology.

[47]  L. Mockros,et al.  Hemolytic effects of energy dissipation in flowing blood , 2006, Medical and biological engineering.

[48]  M Grigioni,et al.  The influence of the leaflets' curvature on the flow field in two bileaflet prosthetic heart valves. , 2001, Journal of biomechanics.

[49]  Leslie W. Miller,et al.  Left Ventricular Assist Device as Destination for Patients Undergoing Intravenous Inotropic Therapy: A Subset Analysis From REMATCH (Randomized Evaluation of Mechanical Assistance in Treatment of Chronic Heart Failure) , 2004, Circulation.

[50]  O H Frazier,et al.  Surgery for Acquired Cardiovascular Disease Initial experience with the AbioCor Implantable Replacement Heart System , 2004 .

[51]  Markus J. Wilhelm,et al.  Treatment of Thrombus Formation Associated With the MicroMed DeBakey VAD Using Recombinant Tissue Plasminogen Activator , 2002, Circulation.

[52]  Klaus Affeld,et al.  Numerical estimation of blood damage in artificial organs. , 2004, Artificial organs.

[53]  N H Hwang,et al.  Estimation of the rotational undamped natural frequency of bileaflet cardiac valve prostheses. , 1990, Journal of biomechanical engineering.

[54]  Samer A. Kseibi,et al.  Impact of high intensity transient signals on the choice of mechanical aortic valve substitutes. , 2003, European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery.

[55]  Danny Bluestein,et al.  Biological effects of dynamic shear stress in cardiovascular pathologies and devices , 2008, Expert review of medical devices.

[56]  Danny Bluestein,et al.  Flow-induced platelet activation in a St. Jude mechanical heart valve, a trileaflet polymeric heart valve, and a St. Jude tissue valve. , 2005, Artificial organs.

[57]  Shmuel Einav,et al.  Design Optimization of a Mechanical Heart Valve for Reducing Valve Thrombogenicity—A Case Study with ATS Valve , 2010, ASAIO journal.

[58]  Edmunds Lh,et al.  Is prosthetic valve thrombogenicity related to design or material , 1996 .

[59]  W Jin,et al.  Experimental investigation of unsteady flow behaviour within a sac-type ventricular assist device (VAD). , 1993, Journal of biomechanics.

[60]  Larry V. McIntire,et al.  Real-time analysis of shear-dependent thrombus formation and its blockade by inhibitors of von Willebrand factor binding to platelets. , 1993 .

[61]  Danny Bluestein,et al.  Research approaches for studying flow-induced thromboembolic complications in blood recirculating devices , 2004, Expert review of medical devices.

[62]  J.,et al.  Whitaker Lecture : Biorheology in Thrombosis Research , 2022 .

[63]  Ajit P Yoganathan,et al.  An analysis of turbulent shear stresses in leakage flow through a bileaflet mechanical prostheses. , 2002, Journal of biomechanical engineering.

[64]  Danny Bluestein,et al.  Fluid mechanics of arterial stenosis: Relationship to the development of mural thrombus , 1997, Annals of Biomedical Engineering.

[65]  H Reul,et al.  Assessment of hemolysis related quantities in a microaxial blood pump by computational fluid dynamics. , 2001, Artificial organs.

[66]  Danny Bluestein,et al.  Thrombogenic Potential of Innovia Polymer Valves versus Carpentier-Edwards Perimount Magna Aortic Bioprosthetic Valves , 2011, ASAIO journal.

[67]  Stanley E. Rittgers,et al.  Particle Motion Within In Vitro Models of Stenosed Internal Carotid and Left Anterior Descending Coronary Arteries , 1998, Annals of Biomedical Engineering.

[68]  Todd D. Giorgio,et al.  Studies on the Mechanisms of Shear-Induced Platelet Activation , 1987 .

[69]  Shmuel Einav,et al.  Device Thrombogenicity Emulation: A Novel Method for Optimizing Mechanical Circulatory Support Device Thromboresistance , 2012, PloS one.

[70]  P M Portner,et al.  Improved outcomes with an implantable left ventricular assist system: a multicenter study. , 2001, The Annals of thoracic surgery.

[71]  G. Grunkemeier,et al.  Twenty years' experience with the Medtronic Hall valve. , 2001, The Journal of thoracic and cardiovascular surgery.

[72]  A. Gosman,et al.  Aspects of Computer Simulation of Liquid-Fueled Combustors , 1983 .

[73]  Hwa Liang Leo,et al.  Bileaflet Aortic Valve Prosthesis Pivot Geometry Influences Platelet Secretion and Anionic Phospholipid Exposure , 2001, Annals of Biomedical Engineering.

[74]  Jan Vierendeels,et al.  Comparison of the hemodynamic and thrombogenic performance of two bileaflet mechanical heart valves using a CFD/FSI model. , 2007, Journal of biomechanical engineering.