Intra- and Intersession Repeatability of an Optical Quality and Intraocular Scattering Measurement System in Children

Purpose To evaluate intra- and intersession repeatability of objective optical quality and intraocular scattering measurements with a double-pass system in children. Methods Forty-two eyes of 42 children were included in the study. An optical quality analysis system (OQAS) was used to measure optical quality parameters, including modulation transfer function cutoff frequency (MTFcutoff), Strehl ratio (SR), OQAS values (OV) at 3 different contrasts and objective scatter index (OSI). Three measurement sessions with 10-min intervals were operated by the same technician, and in each session four consecutive measurements were obtained. Results Mean values for MTFcutoff, SR and OSI were 46.85 ± 7.45cpd, 0.27 ± 0.06 and 0.34 ± 0.22 respectively. 1) The intraclass correlation coefficients were ranged from 0.89 to 0.97 and coefficients of variation from 0.06 to 0.16 for all the parameters in the first session; the relative repeatability were 11.1% (MTFcutoff), 22.5% (SR), 10.9% (OV100%), 16.6% (OV2%), 22.4% (OV9%) and 56.3% (OSI). Similar results were found in the second and third sessions. 2) Bland-Altman analysis showed that narrow 95% confidence intervals (compared between the first and second sessions) ranged from -5.42 to 5.28 (MTFcutoff), -0.05 to 0.07 (SR), -0.18 to 0.18 (OV100%), -0.26 to 0.29 (OV20%), -0.33 to 0.39 (OV9%) and -0.11 to 0.09 (OSI); the comparison between any two of the three sessions showed similar results. Conclusion Measurements of optical quality and intraocular scattering in children by the double-pass system showed good intra- and intersession repeatability. Retinal image quality is high and intraocular scattering is low in children.

[1]  K. Pesudovs,et al.  Anterior segment biometry with the Pentacam: Comprehensive assessment of repeatability of automated measurements , 2008, Journal of cataract and refractive surgery.

[2]  K. Shimizu,et al.  Effect of Aging on Optical Quality and Intraocular Scattering Using the Double-Pass Instrument , 2012, Current eye research.

[3]  K. Shimizu,et al.  Comparison of Optical Quality and Intraocular Scattering after Posterior Chamber Phakic Intraocular Lens with and without a Central Hole (Hole ICL and Conventional ICL) Implantation Using the Double-Pass Instrument , 2013, PloS one.

[4]  Jaume Pujol,et al.  Comparison of the retinal image quality with a Hartmann-Shack wavefront sensor and a double-pass instrument. , 2006, Investigative ophthalmology & visual science.

[5]  Richard Held,et al.  Wavefront aberrations in eyes of emmetropic and moderately myopic school children and young adults , 2002, Vision Research.

[6]  Yongfu Yu,et al.  Optical quality and intraocular scattering after femtosecond laser small incision lenticule extraction. , 2014, Journal of refractive surgery.

[7]  M. R. Hoffman,et al.  Relationship between ocular wavefront aberrations and refractive error in Chinese school children , 2012, Clinical & experimental optometry.

[8]  Raúl Martín,et al.  Optical quality and intraocular scattering assessed with a double-pass system in eyes with contact lens induced corneal swelling. , 2014, Contact lens & anterior eye : the journal of the British Contact Lens Association.

[9]  Jaume Pujol,et al.  An Objective Scatter Index Based on Double-Pass Retinal Images of a Point Source to Classify Cataracts , 2011, PloS one.

[10]  Hind Ibrahem Aldebasi,et al.  Ocular aberrations in amblyopic children. , 2013, Saudi journal of ophthalmology : official journal of the Saudi Ophthalmological Society.

[11]  Alain Saad,et al.  Repeatability of measurements with a double‐pass system , 2010, Journal of cataract and refractive surgery.

[12]  D. Altman,et al.  STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT , 1986, The Lancet.

[13]  Juan C. Ondategui,et al.  Optical quality after myopic photorefractive keratectomy and laser in situ keratomileusis: Comparison using a double‐pass system , 2012, Journal of cataract and refractive surgery.

[14]  K. Shimizu,et al.  Clinical evaluation of optical quality and intraocular scattering after posterior chamber phakic intraocular lens implantation. , 2012, Investigative ophthalmology & visual science.

[15]  Image quality in wearers of a centre distance concentric design bifocal contact lens , 2002, Ophthalmic & physiological optics : the journal of the British College of Ophthalmic Opticians.

[16]  J. Alió,et al.  Grading nuclear cataract opacity by densitometry and objective optical analysis , 2008, Journal of cataract and refractive surgery.

[17]  S. Pieh,et al.  Anterior Chamber Depth Measurement With A-Scan Ultrasonography, Orbscan II, and IOLMaster , 2005, Optometry and vision science : official publication of the American Academy of Optometry.

[18]  Jaume Pujol,et al.  Optical Quality Analysis System; Instrument for objective clinical evaluation of ocular optical quality. , 2004, Journal of cataract and refractive surgery.

[20]  A. Lam,et al.  Reliability and repeatability of the Pentacam on corneal curvatures , 2009, Clinical & experimental optometry.

[21]  Jaume Pujol,et al.  Intra- and Intersession Repeatability of a Double-Pass Instrument , 2010, Optometry and vision science : official publication of the American Academy of Optometry.

[22]  Aixa Alarcon,et al.  Optical quality and visual discrimination capacity after myopic LASIK with a standard and aspheric ablation profile. , 2011, Journal of refractive surgery.

[23]  A. Beckett,et al.  AKUFO AND IBARAPA. , 1965, Lancet.

[24]  Kimiya Shimizu,et al.  Longitudinal Assessment of Optical Quality and Intraocular Scattering Using the Double-Pass Instrument in Normal Eyes and Eyes with Short Tear Breakup Time , 2013, PloS one.

[25]  Jaume Pujol,et al.  Optical quality and intraocular scattering in a healthy young population , 2011, Clinical & experimental optometry.

[26]  J. Alió,et al.  Intra‐observer repeatability of optical quality measures provided by a double‐pass system , 2012, Clinical & experimental optometry.

[27]  P. Artal,et al.  Grading nuclear, cortical and posterior subcapsular cataracts using an objective scatter index measured with a double-pass system , 2012, British Journal of Ophthalmology.

[28]  Juan C. Ondategui,et al.  Effect of laser in situ keratomileusis on vision analyzed using preoperative optical quality , 2010, Journal of cataract and refractive surgery.

[29]  D. Altman,et al.  Statistics Notes: Measurement error and correlation coefficients , 1996, BMJ.

[30]  C. McAlinden,et al.  Objective assessment of crystalline lens opacity level by measuring ocular light scattering with a double-pass system. , 2013, American journal of ophthalmology.

[31]  D. Altman,et al.  Measurement error. , 1996, BMJ.

[32]  J Pujol,et al.  Double‐pass measurements of retinal image quality in monofocal contact lens wearers , 1997, Ophthalmic & physiological optics : the journal of the British College of Ophthalmic Opticians.