Design considerations to overcome cross talk in a photon counting silicon strip detector for computed tomography

This article presents a Monte Carlo simulation of the detector energy response in the presence of pileup in a segmented silicon microstrip detector designed for high flux spectral computed tomography with sub-millimeter pixel size. Currents induced on the collection electrode of a pixel segment are explicitly modeled and signals emanating from events in neighboring pixels are superimposed together with electronic noise before the entire pulse train is processed by a model of the readout electronics to obtain the detector energy response function. The article shows how the lower threshold and the time constant of the electronic filters need to be set in order to minimize the detrimental influence of cross talk from neighboring pixel segments, an issue that is aggravated by the sub-millimeter pixel size and the proposed segmented detector design.

[1]  K. Ogawa,et al.  Photon Counting X-ray CT System with a Semiconductor Detector , 2006, 2006 IEEE Nuclear Science Symposium Conference Record.

[2]  Polad M Shikhaliev,et al.  Projection x-ray imaging with photon energy weighting: experimental evaluation with a prototype detector , 2009, Physics in medicine and biology.

[3]  R. Nowotny Application of Si-microstrip-detectors in medicine and structural analysis , 1984 .

[4]  Robert E. Melen,et al.  CdZnTe detector array for a scanning-beam digital x-ray system , 1999, Medical Imaging.

[5]  Cheng Xu,et al.  Photon-counting CT with silicon detectors: feasibility for pediatric imaging , 2009, Medical Imaging.

[6]  S. Yoshida,et al.  Application of silicon strip detectors to X-ray computed tomography , 2005 .

[7]  J. H. Hubbell,et al.  XCOM: Photon Cross Section Database (version 1.2) , 1999 .

[8]  Katsuyuki Taguchi,et al.  Characterization of a novel photon counting detector for clinical CT: count rate, energy resolution, and noise performance , 2009, Medical Imaging.

[9]  S. Evans Catalogue of Diagnostic X-Ray Spectra and Other Data , 1998 .

[10]  H. Fitting,et al.  Electron penetration and energy transfer in solid targets , 1977 .

[11]  G. Knoll Radiation detection and measurement , 1979 .

[12]  B. Marangelli,et al.  A new Monte Carlo code for full simulation of silicon strip detectors , 2004 .

[13]  E. Roessl,et al.  K-edge imaging in x-ray computed tomography using multi-bin photon counting detectors , 2007, Physics in medicine and biology.

[14]  O. Klein,et al.  Über die Streuung von Strahlung durch freie Elektronen nach der neuen relativistischen Quantendynamik von Dirac , 1929 .

[15]  Willy Sansen,et al.  Limits of low noise performance of detector readout front ends in CMOS technology , 1990 .

[16]  M. Danielsson,et al.  Photon-counting spectral computed tomography using silicon strip detectors: a feasibility study , 2010, Physics in medicine and biology.

[17]  High-resolution X-ray imaging using the signal time dependence on a double-sided silicon detector , 1999 .

[18]  W. Shockley Currents to Conductors Induced by a Moving Point Charge , 1938 .