Sensitivity of coded aperture Raman spectroscopy to analytes beneath turbid biological tissue and tissue-simulating phantoms

Abstract. Traditional slit-based spectrometers have an inherent trade-off between spectral resolution and throughput that can limit their performance when measuring diffuse sources such as light returned from highly scattering biological tissue. Recently, multielement fiber bundles have been used to effectively measure diffuse sources, e.g., in the field of spatially offset Raman spectroscopy, by remapping the source (or some region of the source) into a slit shape for delivery to the spectrometer. Another approach is to change the nature of the instrument by using a coded entrance aperture, which can increase throughput without sacrificing spectral resolution. In this study, two spectrometers, one with a slit-based entrance aperture and the other with a coded aperture, were used to measure Raman spectra of an analyte as a function of the optical properties of an overlying scattering medium. Power-law fits reveal that the analyte signal is approximately proportional to the number of transport mean free paths of the scattering medium raised to a power of −0.47 (coded aperture instrument) or −1.09 (slit-based instrument). These results demonstrate that the attenuation in signal intensity is more pronounced for the slit-based instrument and highlight the scattering regimes where coded aperture instruments can provide an advantage over traditional slit-based spectrometers.

[1]  H. Bruining,et al.  In vivo confocal Raman microspectroscopy of the skin: noninvasive determination of molecular concentration profiles. , 2001, The Journal of investigative dermatology.

[2]  Jason R. Maher,et al.  Determination of Ideal Offset for Spatially Offset Raman Spectroscopy , 2010, Applied spectroscopy.

[3]  William F. Finney,et al.  Subsurface Probing in Diffusely Scattering Media Using Spatially Offset Raman Spectroscopy , 2005, Applied spectroscopy.

[4]  M. Golay Multi-slit spectrometry. , 1949, Journal of the Optical Society of America.

[5]  P. Matousek,et al.  Depth Profiling in Diffusely Scattering Media Using Raman Spectroscopy and Picosecond Kerr Gating , 2005, Applied spectroscopy.

[6]  N. Pitsianis,et al.  Static two-dimensional aperture coding for multimodal, multiplex spectroscopy. , 2006, Applied optics.

[7]  Pavel Matousek,et al.  Prospects for the diagnosis of breast cancer by noninvasive probing of calcifications using transmission Raman spectroscopy. , 2007, Journal of biomedical optics.

[8]  P. Matousek,et al.  Recent advances in the development of Raman spectroscopy for deep non‐invasive medical diagnosis , 2013, Journal of biophotonics.

[9]  H. J. van Staveren,et al.  Light scattering in Intralipid-10% in the wavelength range of 400-1100 nm. , 1991, Applied optics.

[10]  D. Katz,et al.  Quantitative Analysis of Microbicide Concentrations in Fluids, Gels and Tissues Using Confocal Raman Spectroscopy , 2013, PloS one.

[11]  H. Barr,et al.  Raman spectroscopy for identification of epithelial cancers. , 2004, Faraday discussions.

[12]  Akira Ishimaru,et al.  Wave propagation and scattering in random media , 1997 .

[13]  William F. Finney,et al.  Subsurface Raman Spectroscopy and Mapping Using a Globally Illuminated Non-Confocal Fiber-Optic Array Probe in the Presence of Raman Photon Migration , 2006, Applied spectroscopy.

[14]  Wei-Chuan Shih,et al.  Determination of uncertainty in parameters extracted from single spectroscopic measurements. , 2007, Journal of biomedical optics.

[15]  Ines Latka,et al.  Fiber optic probes for linear and nonlinear Raman applications – Current trends and future development , 2013 .

[16]  Pavel Matousek,et al.  Inverse Spatially Offset Raman Spectroscopy for Deep Noninvasive Probing of Turbid Media , 2006, Applied spectroscopy.

[17]  R. Dasari,et al.  Prospects for in vivo Raman spectroscopy , 2000 .

[18]  C. Krafft,et al.  Biomedical applications of Raman and infrared spectroscopy to diagnose tissues , 2006 .

[19]  A. Welch,et al.  A review of the optical properties of biological tissues , 1990 .

[20]  Ingo Gersonde,et al.  Influence of tissue absorption and scattering on the depth dependent sensitivity of Raman fiber probes investigated by Monte Carlo simulations , 2011, Biomedical optics express.

[21]  Richard P Van Duyne,et al.  Transcutaneous glucose sensing by surface-enhanced spatially offset Raman spectroscopy in a rat model. , 2010, Analytical chemistry.

[22]  David F. Katz,et al.  Multicompartmental Pharmacokinetic Model of Tenofovir Delivery by a Vaginal Gel , 2013, PloS one.

[23]  Vanni Nardino,et al.  Radiometric and signal-to-noise ratio properties of multiplex dispersive spectrometry. , 2010, Applied optics.

[24]  Shmuel Sternklar,et al.  Transition from the ballistic to the diffusive regime in a turbid medium. , 2011, Optics letters.

[25]  R. Richards-Kortum,et al.  Raman spectroscopy for the detection of cancers and precancers. , 1996, Journal of biomedical optics.

[26]  Jun Q. Lu,et al.  Determination of refractive indices of porcine skin tissues and intralipid at eight wavelengths between 325 and 1557 nm. , 2005, Journal of the Optical Society of America. A, Optics, image science, and vision.

[27]  Michael S Feld,et al.  Optical fiber probe for biomedical Raman spectroscopy. , 2004, Applied optics.

[28]  B. Pogue,et al.  Review of tissue simulating phantoms for optical spectroscopy, imaging and dosimetry. , 2006, Journal of biomedical optics.

[29]  S L Jacques,et al.  Optical properties of intralipid: A phantom medium for light propagation studies , 1992, Lasers in surgery and medicine.

[30]  D. Brady,et al.  High-throughput, multiplexed pushbroom hyperspectral microscopy. , 2008, Optics express.

[31]  M E Gehm,et al.  Coded Aperture Raman Spectroscopy for Quantitative Measurements of Ethanol in a Tissue Phantom , 2006, Applied spectroscopy.

[32]  A. Goodship,et al.  Numerical Simulations of Subsurface Probing in Diffusely Scattering Media Using Spatially Offset Raman Spectroscopy , 2005, Applied spectroscopy.

[33]  Walter J. Riker A Review of J , 2010 .

[34]  Ashwin A. Wagadarikar,et al.  Performance comparison of aperture codes for multimodal, multiplex spectroscopy. , 2007, Applied optics.

[35]  R. Dasari,et al.  Investigation of the specificity of Raman spectroscopy in non-invasive blood glucose measurements , 2011, Analytical and bioanalytical chemistry.

[36]  Anita Mahadevan-Jansen,et al.  Spatially offset Raman spectroscopy of layered soft tissues. , 2009, Optics letters.

[37]  Valery V. Tuchin,et al.  OPTICAL PROPERTIES OF SKIN, SUBCUTANEOUS, AND MUSCLE TISSUES: A REVIEW , 2011 .

[38]  Jason R. Maher,et al.  Overconstrained library-based fitting method reveals age- and disease-related differences in transcutaneous Raman spectra of murine bones , 2013, Journal of biomedical optics.

[39]  A. Goodship,et al.  Prediction of sublayer depth in turbid media using spatially offset Raman spectroscopy. , 2008, Analytical chemistry.