Small field dosimetric characterization of a new 160-leaf MLC.

The goal of this work was to perform a 6 MV small field characterization of the new Agility 160-leaf multi-leaf collimator (MLC) from Elekta. This included profile measurement analysis and central axis relative output measurements using various diode detectors and an air-core fiber optic scintillation dosimeter (FOD). Data was acquired at a depth of 10.0 cm for field sizes of 1.0, 0.9, 0.8, 0.7, 0.6 and 0.5 cm. Three experimental data sets, comprised of five readings, were made for both the relative output and profile measurements. Average detector-specific output ratios (OR[overline](f(clin))(det))) were calculated with respect to a field size of 3.0 cm and small field replacement correction factors (k(f(clin),f(msr))(Q(clin),Q(msr))) derived for the diodes using the scintillation dosimeter readings as the baseline. The standard experimental uncertainty on OR[overline](f(clin))(det)) was calculated at a 90% confidence interval and the coefficient of variation (CV) used to characterize the detector-specific measurement precision. The positional accuracy of the collimation system was also investigated by analyzing the repeated profile measurements and field width constancy investigated as a function of collimator rotation. For comparison the output and profile measurements were repeated using the Elekta 80-leaf MLCi2 on a beam matched linac at 6 MV. The measured OR[overline](f(clin))(det)) varied as a function of detector and MLC design. At the smallest field size the standard experimental uncertainty on OR[overline](f(clin))(det)) was consistent across all detectors at approximately 0.5% and 1.0% for Agility and MLCi2 collimators respectively. The CV associated with the FOD measurements were greater than that of the diodes but did not translate into increased measurement uncertainty. At the smallest field size, the diode detector correction factors were approximately 2% greater for MLCi2 than that required for the Agility. Profile data revealed the Agility MLC to have a greater positional reproducibility than both the MLCi2 and the linac diaphragms (jaws), as also reflected in the experimental uncertainties on OR[overline](f(clin))(det)). The relative output, profile widths and associated uncertainties were all found to differ between the two MLC systems investigated, as were the field size specific diode detector replacement correction factors. The data also clearly showed that the Agility 160-leaf MLC performs to a tighter positional tolerance than the MLCi2.

[1]  W. Laub,et al.  The volume effect of detectors in the dosimetry of small fields used in IMRT. , 2003, Medical physics.

[2]  J. Seuntjens,et al.  A new formalism for reference dosimetry of small and nonstandard fields. , 2008, Medical physics.

[3]  S Cora,et al.  Calculation of k(Q(clin),Q(msr) ) (f(clin),f(msr) ) for several small detectors and for two linear accelerators using Monte Carlo simulations. , 2011, Medical physics.

[4]  L Souhami,et al.  Radiation Therapy Oncology Group: radiosurgery quality assurance guidelines. , 1993, International journal of radiation oncology, biology, physics.

[5]  W. Curran,et al.  Dosimetric performance of the new high‐definition multileaf collimator for intracranial stereotactic radiosurgery , 2010, Journal of applied clinical medical physics.

[6]  J. Perks,et al.  Comparison of a micro-multileaf collimator with a 5-mm-leaf-width collimator for intracranial stereotactic radiotherapy. , 2003, International journal of radiation oncology, biology, physics.

[7]  D. Shrieve,et al.  Is smaller better? Comparison of 3-mm and 5-mm leaf size for stereotactic radiosurgery: A dosimetric study , 2006 .

[8]  Natalka Suchowerska,et al.  Small field diode correction factors derived using an air core fibre optic scintillation dosimeter and EBT2 film , 2012, Physics in medicine and biology.

[9]  David I Thwaites,et al.  Notes on the construction of solid-state detectors. , 2006, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[10]  A Moutsatsos,et al.  On the output factor measurements of the CyberKnife iris collimator small fields: Experimental determination of the k(Q(clin),Q(msr) ) (f(clin),f(msr) ) correction factors for microchamber and diode detectors. , 2012, Medical physics.

[11]  John D Fenwick,et al.  Characterizing the influence of detector density on dosimeter response in non-equilibrium small photon fields , 2012, Physics in medicine and biology.

[12]  D. G. Rees,et al.  Introductory Medical Statistics. , 1976 .

[13]  David I Thwaites,et al.  Experimental small field 6 MV output ratio analysis for various diode detector and accelerator combinations. , 2011, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[14]  D I Thwaites,et al.  Monte Carlo modelling of diode detectors for small field MV photon dosimetry: detector model simplification and the sensitivity of correction factors to source parameterization , 2012, Physics in medicine and biology.

[15]  N Suchowerska,et al.  Plastic scintillation dosimetry: comparison of three solutions for the Cerenkov challenge. , 2011, Physics in medicine and biology.

[16]  Carlos De Wagter,et al.  Physical Characterization of a New Concept Design of an Elekta Radiation Head with Integrated 160-leaf Multi-leaf Collimator , 2009 .

[17]  Paul Z Y Liu,et al.  Real-time scintillation array dosimetry for radiotherapy: the advantages of photomultiplier detectors. , 2012, Medical physics.

[18]  A L Boyer,et al.  Clinical dosimetry for implementation of a multileaf collimator. , 1992, Medical physics.

[19]  D I Thwaites,et al.  Implementing a newly proposed Monte Carlo based small field dosimetry formalism for a comprehensive set of diode detectors. , 2011, Medical physics.

[20]  B. Salter,et al.  Conformity of LINAC-based stereotactic radiosurgery using dynamic conformal arcs and micro-multileaf collimator. , 2009, International journal of radiation oncology, biology, physics.

[21]  N Suchowerska,et al.  Cerenkov-free scintillation dosimetry in external beam radiotherapy with an air core light guide , 2008, Physics in medicine and biology.

[22]  S Cora,et al.  Monte Carlo simulated correction factors for machine specific reference field dose calibration and output factor measurement using fixed and iris collimators on the CyberKnife system , 2012, Physics in medicine and biology.

[23]  V. Budach,et al.  Commissioning of a micro multi-leaf collimator and planning system for stereotactic radiosurgery. , 1999, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[24]  Todd McNutt,et al.  Validation of physics improvements for IMRT with a commercial treatment‐planning system , 2005, Journal of applied clinical medical physics.

[25]  D. Harder,et al.  Dosimetric characteristics of a new unshielded silicon diode and its application in clinical photon and electron beams. , 2005, Medical physics.

[26]  Natalka Suchowerska,et al.  A prototype scintillation dosimeter customized for small and dynamic megavoltage radiation fields , 2010, Physics in medicine and biology.