Lorentz force electrical impedance tomography

Abstract This article describes a method called Lorentz Force Electrical Impedance Tomography. The electrical conductivity of biological tissues can be measured through their sonication in a magnetic field: the vibration of the tissues inside the field induces an electrical current by Lorentz force. This current, detected by electrodes placed around the sample, is proportional to the ultrasonic pressure, to the strength of the magnetic field and to the electrical conductivity gradient along the acoustic axis. By focusing at different places inside the sample, a map of the electrical conductivity gradient can be established. In this study, experiments were conducted on a gelatin phantom and on a beef sample, successively placed in a 300 mT magnetic field and sonicated with an ultrasonic transducer focused at 21 cm emitting 500 kHz bursts. Although all interfaces are not visible, in this exploratory study a good correlation is observed between the electrical conductivity image and the ultrasonic image. This method offers an alternative to detecting pathologies invisible to standard ultrasonography.

[1]  C Gabriel,et al.  The dielectric properties of biological tissues: I. Literature survey. , 1996, Physics in medicine and biology.

[3]  Dieter Haemmerich,et al.  In vivo electrical conductivity of hepatic tumours. , 2003, Physiological measurement.

[4]  M. Unser,et al.  A bulk modulus dependent linear model for acoustical imaging. , 2009, Journal of the Acoustical Society of America.

[5]  J. Jossinet,et al.  Electric current generated by ultrasonically induced Lorentz force in biological media , 2006, Medical and Biological Engineering and Computing.

[6]  Bradley J Roth,et al.  The role of magnetic forces in biology and medicine , 2011, Experimental biology and medicine.

[7]  David Isaacson,et al.  Electrical Impedance Tomography , 1999, SIAM Rev..

[8]  Y. Xu,et al.  Magneto-acousto-electrical tomography: a potential method for imaging current density and electrical impedance. , 2008, Physiological measurement.

[9]  H. Ammari,et al.  Reconstruction of Small Inhomogeneities from Boundary Measurements , 2005 .

[10]  Peter Basser,et al.  A theoretical model for magneto-acoustic imaging of bioelectric currents , 1994, IEEE Transactions on Biomedical Engineering.

[11]  Jean-Louis Coatrieux,et al.  Image-Guided Therapy: Evolution and Breakthrough [A Look At] , 2010, IEEE Engineering in Medicine and Biology Magazine.

[12]  J. Mari,et al.  Electromagnetic hydrophone for high-intensity focused ultrasound ( HIFU ) measurement , 2013 .

[13]  Bin He,et al.  Magnetoacoustic tomography with magnetic induction (MAT-MI) , 2005, Physics in medicine and biology.

[14]  J. Mari,et al.  Electromagnetic hydrophone with tomographic system for absolute velocity field mapping , 2012, 1207.0324.

[15]  Bin He,et al.  Magnetoacoustic imaging of human liver tumor with magnetic induction. , 2011, Applied physics letters.

[16]  Eleanor Stride,et al.  An approximate nonlinear model for time gain compensation of amplitude modulated images of ultrasound contrast agent perfusion , 2010, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[17]  B. Roth,et al.  Comments on "Hall Effect Imaging" [with reply] , 1998 .

[18]  J Jossinet,et al.  Scanning Electric Conductivity Gradients with Ultrasonically-Induced Lorentz Force , 2001, Ultrasonic imaging.

[19]  Habib Ammari,et al.  Mathematical models and reconstruction methods in magneto-acoustic imaging , 2009, European Journal of Applied Mathematics.

[20]  J. Shah,et al.  Hall effect imaging , 1998, IEEE Transactions on Biomedical Engineering.