A review of improved fixation methods for dental implants. Part II: biomechanical integrity at bone-implant interface.

PURPOSE The purpose of this article is to review the mechanical requirements of the tissue-implant interface and analyze related theories. STUDY SELECTION The osseointegration capacity of titanium implants has been investigated over the past 50 years. We considered the ultimate goal of osseointegration to which form a desirable interfacial layer and a bone matrix with adequate biomechanical properties. RESULTS Occasionally, the interface comprises porous titanium and bone ingrowth that enables a functionally graded Young's modulus, thereby allowing reduction of stress shielding. However, the optimal biomechanical connection at the interface has not yet been fully clarified. There have been publications supporting several universal mechanical testing technologies in terms of bone-titanium bonding ability, although the separation of newly formed bone quality is unlikely. CONCLUSIONS The understanding of complex mechanical bone behavior and size-dependent properties ranging from a nano- to a macroscopic level are essential in the biomechanical optimization of implants. The requirements of regenerated tissue at the interface include high strength, fracture toughness related to ductility, and time-dependent energy dissipation and/or elastic-plastic stress distribution. Moreover, a strong relationship between strain signals and peri-implant tissue turnover could be expected, so that ideal implant biomechanics may enable longevity via adaptive bone remodeling.

[1]  M. Nugent,et al.  Lysyl Oxidase Propeptide Inhibits FGF-2-induced Signaling and Proliferation of Osteoblasts* , 2010, The Journal of Biological Chemistry.

[2]  C. Engh,et al.  Producing and avoiding stress shielding. Laboratory and clinical observations of noncemented total hip arthroplasty. , 1992, Clinical orthopaedics and related research.

[3]  E. Luo,et al.  Effect of combined local treatment with zoledronic acid and basic fibroblast growth factor on implant fixation in ovariectomized rats. , 2009, Bone.

[4]  Pasquale Franciosa,et al.  Effects of thread features in osseo-integrated titanium implants using a statistics-based finite element method. , 2012, Dental materials : official publication of the Academy of Dental Materials.

[5]  A. van der Plas,et al.  Sensitivity of osteocytes to biomechanical stress in vitro , 1995, FASEB journal : official publication of the Federation of American Societies for Experimental Biology.

[6]  Y Zilberman,et al.  Osseous adaptation to continuous loading of rigid endosseous implants. , 1984, American journal of orthodontics.

[7]  M. Ito,et al.  Contribution of trabecular and cortical components to the mechanical properties of bone and their regulating parameters. , 2002, Bone.

[8]  J. Klein-Nulend,et al.  MECHANOTRANSDUCTION IN BONE : ROLE OF THE LACUNOCANALICULAR NETWORK , 1999 .

[9]  V. Sikavitsas,et al.  Biomaterials and bone mechanotransduction. , 2001, Biomaterials.

[10]  Qing Li,et al.  Mandibular bone remodeling induced by dental implant. , 2010, Journal of biomechanics.

[11]  J. Galante,et al.  ESB Research Award 1992. The mechanism of bone remodeling and resorption around press-fitted THA stems. , 1993, Journal of biomechanics.

[12]  M. Dard,et al.  Predicting bone remodeling around tissue- and bone-level dental implants used in reduced bone width. , 2013, Journal of biomechanics.

[13]  U. Lekholm,et al.  Variations in bone regeneration adjacent to implants augmented with barrier membranes alone or with demineralized freeze-dried bone or autologous grafts: a study in dogs. , 1995, The International journal of oral & maxillofacial implants.

[14]  Karen A. Dehring,et al.  Correlating Changes in Collagen Secondary Structure with Aging and Defective Type II Collagen by Raman Spectroscopy , 2006, Applied spectroscopy.

[15]  Y. Shibata,et al.  Glow discharge plasma treatment of titanium plates enhances adhesion of osteoblast-like cells to the plates through the integrin-mediated mechanism. , 2002, The International journal of oral & maxillofacial implants.

[16]  Baohua Ji,et al.  Mechanical properties of nanostructure of biological materials , 2004 .

[17]  R M Pilliar,et al.  Porous-surfaced metallic implants for orthopedic applications. , 1987, Journal of biomedical materials research.

[18]  Michelle L. Oyen,et al.  Poroelastic nanoindentation responses of hydrated bone , 2008 .

[19]  T. Takano-Yamamoto,et al.  The Use of Small Titanium Screws for Orthodontic Anchorage , 2003, Journal of dental research.

[20]  H. Grootenboer,et al.  Adaptive bone-remodeling theory applied to prosthetic-design analysis. , 1987, Journal of biomechanics.

[21]  Yuan-Ting Chang,et al.  Effects of osteoporosis and nutrition supplements on structures and nanomechanical properties of bone tissue. , 2011, Journal of the mechanical behavior of biomedical materials.

[22]  Wurihan,et al.  Nanomechanical properties and molecular structures of in vitro mineralized tissues on anodically-oxidized titanium surfaces. , 2014, Nanomedicine : nanotechnology, biology, and medicine.

[23]  Alexei Bolshakov,et al.  Influences of stress on the measurement of mechanical properties using nanoindentation: Part II. Finite element simulations , 1996 .

[24]  J. Jurvelin,et al.  Precision of nanoindentation protocols for measurement of viscoelasticity in cortical and trabecular bone. , 2010, Journal of biomechanics.

[25]  George M. Pharr,et al.  On the generality of the relationship among contact stiffness, contact area, and elastic modulus during indentation , 1992 .

[26]  J. Galante,et al.  Metal Release in Patients Who Have Had a Primary Total Hip Arthroplasty. A Prospective, Controlled, Longitudinal Study* , 1998, The Journal of bone and joint surgery. American volume.

[27]  S. Cowin,et al.  A model for the excitation of osteocytes by mechanical loading-induced bone fluid shear stresses. , 1994, Journal of biomechanics.

[28]  Mitsuo Niinomi,et al.  Mechanical properties of biomedical titanium alloys , 1998 .

[29]  M. Swain,et al.  Micromechanical evaluation of mineralized multilayers. , 2008, Journal of biomechanics.

[30]  Nicole J. Crane,et al.  Identifying Chemical Changes in Subchondral Bone Taken from Murine Knee Joints Using Raman Spectroscopy , 2006, Applied spectroscopy.

[31]  T. Keaveny,et al.  Yield strain behavior of trabecular bone. , 1998, Journal of biomechanics.

[32]  S. Majumdar,et al.  Processing and Analysis of In Vivo High-Resolution MR Images of Trabecular Bone for Longitudinal Studies: Reproducibility of Structural Measures and Micro-Finite Element Analysis Derived Mechanical Properties , 2002, Osteoporosis International.

[33]  K. Karan,et al.  Molecular structural analysis of noncarious cervical sclerotic dentin using Raman spectroscopy , 2009 .

[34]  John B. Brunski,et al.  In Vivo Bone Response to Biomechanical Loading at the Bone/Dental-Implant Interface , 1999, Advances in dental research.

[35]  S. Weiner,et al.  On the relationship between the microstructure of bone and its mechanical stiffness. , 1992, Journal of biomechanics.

[36]  T. Hayakawa,et al.  Mode superposition transient dynamic analysis for dental implants with stress-absorbing elements: a finite element analysis. , 2006, Dental materials journal.

[37]  Jui-Ting Hsu,et al.  Biomechanical simulation of various surface roughnesses and geometric designs on an immediately loaded dental implant , 2010, Comput. Biol. Medicine.

[38]  G. Ozin,et al.  New nanocomposites: putting organic function 'inside' the channel walls of periodic mesoporous silica , 2000 .

[39]  Claudia Beleites,et al.  Chemical imaging of articular cartilage sections with Raman mapping, employing uni- and multi-variate methods for data analysis. , 2010, The Analyst.

[40]  Yang-Tse Cheng,et al.  Determining the instantaneous modulus of viscoelastic solids using instrumented indentation measurements , 2005 .

[41]  Engh Ca,et al.  Principles, techniques, results, and complications with a porous-coated sintered metal system. , 1986 .

[42]  A Scarano,et al.  Immediate implantation in fresh extraction sockets. A controlled clinical and histological study in man. , 2001, Journal of periodontology.

[43]  M V Swain,et al.  Micromechanical Property Recovery of Human Carious Dentin Achieved with Colloidal Nano-β-tricalcium Phosphate , 2008, Journal of dental research.

[44]  Yasuhiro Tanimoto,et al.  Influence of cortical bone quality on stress distribution in bone around dental implant. , 2005, Dental materials journal.

[45]  P. Bravetti,et al.  Interaction of bone-dental implant with new ultra low modulus alloy using a numerical approach. , 2014, Materials science & engineering. C, Materials for biological applications.

[46]  T. Tanaka,et al.  Light and electron microscopic studies of bone-titanium interface in the tibiae of young and mature rats. , 1996, Journal of biomedical materials research.

[47]  Wurihan,et al.  Antioxidant and osteogenic properties of anodically oxidized titanium. , 2012, Journal of the mechanical behavior of biomedical materials.

[48]  K. Berhane,et al.  Predictable crestal bone remodelling around two porous-coated titanium alloy dental implant designs. A radiographic study in dogs. , 1994, Clinical oral implants research.

[49]  Tadashi Kokubo,et al.  Pore throat size and connectivity determine bone and tissue ingrowth into porous implants: three-dimensional micro-CT based structural analyses of porous bioactive titanium implants. , 2006, Biomaterials.

[50]  Steve Weiner,et al.  THE MATERIAL BONE: Structure-Mechanical Function Relations , 1998 .

[51]  A. A. Griffith The Phenomena of Rupture and Flow in Solids , 1921 .

[52]  L Catherine Brinson,et al.  Mechanics considerations for microporous titanium as an orthopedic implant material. , 2004, Journal of biomedical materials research. Part A.

[53]  P. Niederer,et al.  Experimental elucidation of mechanical load-induced fluid flow and its potential role in bone metabolism and functional adaptation. , 1998, The American journal of the medical sciences.

[54]  Tan Wen-chang,et al.  A note on unsteady flows of a viscoelastic fluid with the fractional Maxwell model between two parallel plates , 2003 .

[55]  M. Mabuchi,et al.  Processing and mechanical properties of autogenous titanium implant materials , 2002, Journal of materials science. Materials in medicine.

[56]  L. Linder High-resolution microscopy of the implant-tissue interface. , 1985, Acta orthopaedica Scandinavica.

[57]  A. Virdi,et al.  Limitations of using micro‐computed tomography to predict bone–implant contact and mechanical fixation , 2012, Journal of microscopy.

[58]  F. Butz,et al.  Harder and Stiffer Bone Osseointegrated to Roughened Titanium , 2006, Journal of dental research.

[59]  Gang Feng,et al.  Effects of Creep and Thermal Drift on Modulus Measurement Using Depth-sensing Indentation , 2002 .

[60]  J. Szivek,et al.  Bone ingrowth and stress shielding with a porous surface coated fracture fixation plate. , 1979, Journal of biomedical materials research.

[61]  Bobyn Jd,et al.  The effect of stem stiffness on femoral bone resorption after canine porous-coated total hip arthroplasty. , 1990, Clinical orthopaedics and related research.

[62]  J. Jurvelin,et al.  Bone density at the proximal femur after total hip arthroplasty. , 1998, Clinical orthopaedics and related research.

[63]  Hongbing Lu,et al.  Measurements of Two Independent Viscoelastic Functions by Nanoindentation , 2007 .

[64]  Bernd Gludovatz,et al.  Fracture resistance of human cortical bone across multiple length-scales at physiological strain rates. , 2014, Biomaterials.

[65]  P. Thurner,et al.  Comparison of microfocus- and synchrotron X-ray tomography for the analysis of osteointegration around Ti6Al4V implants. , 2004, European cells & materials.

[66]  M. Yoshinari,et al.  An immunoelectron microscopic localization of noncollagenous bone proteins (osteocalcin and osteopontin) at the bone-titanium interface of rat tibiae. , 1998, Journal of biomedical materials research.

[67]  Michael Kjaer,et al.  Role of extracellular matrix in adaptation of tendon and skeletal muscle to mechanical loading. , 2004, Physiological reviews.

[68]  Lisa A. Pruitt,et al.  Nanoindentation of biological materials , 2006 .

[69]  Wurihan,et al.  Strain-rate stiffening of cortical bone: observations and implications from nanoindentation experiments. , 2014, Nanoscale.

[70]  M. Morris,et al.  Application of vibrational spectroscopy to the study of mineralized tissues (review). , 2000, Journal of biomedical optics.

[71]  S. Majumdar,et al.  In Vivo Assessment of Architecture and Micro-Finite Element Analysis Derived Indices of Mechanical Properties of Trabecular Bone in the Radius , 2002, Osteoporosis International.

[72]  S D Cook,et al.  A three-dimensional finite element analysis of a porous rooted Co-Cr-Mo alloy dental implant. , 1982, Journal of dental research.

[73]  D. Munz,et al.  Determination of elastic modulus of thin layers using nanoindentation , 1997 .

[74]  A. Boyde,et al.  Hydration effects on the micro-mechanical properties of bone , 2006 .

[75]  L. Lefebvre,et al.  Human alveolar bone cell proliferation, expression of osteoblastic phenotype, and matrix mineralization on porous titanium produced by powder metallurgy. , 2009, Clinical oral implants research.

[76]  R Mendelsohn,et al.  Spectroscopic Characterization of Collagen Cross‐Links in Bone , 2001, Journal of bone and mineral research : the official journal of the American Society for Bone and Mineral Research.

[77]  L E Lanyon,et al.  Functional strain in bone tissue as an objective, and controlling stimulus for adaptive bone remodelling. , 1987, Journal of biomechanics.

[78]  Abhay Pandit,et al.  Fabrication methods of porous metals for use in orthopaedic applications. , 2006, Biomaterials.

[79]  G. Pharr,et al.  An improved technique for determining hardness and elastic modulus using load and displacement sensing indentation experiments , 1992 .

[80]  Felix Beckmann,et al.  Osteoconductive modifications of Ti-implants in a goat defect model: characterization of bone growth with SR muCT and histology. , 2005, Biomaterials.

[81]  R. Huiskes,et al.  The relationship between stress shielding and bone resorption around total hip stems and the effects of flexible materials. , 1992, Clinical orthopaedics and related research.

[82]  Nanoindentation-derived elastic modulus of an amorphous polymer and its sensitivity to load-hold period and unloading strain rate , 2008 .

[83]  H. Grootenboer,et al.  The behavior of adaptive bone-remodeling simulation models. , 1992, Journal of biomechanics.

[84]  S. Bhaduri,et al.  Gradient surface porosity in titanium dental implants: relation between processing parameters and microstructure , 2004, Journal of materials science. Materials in medicine.

[85]  E. Rocha,et al.  Influence of microthreads and platform switching on stress distribution in bone using angled abutments. , 2012, Journal of prosthodontic research.

[86]  D. Puleo,et al.  Understanding and controlling the bone-implant interface. , 1999, Biomaterials.

[87]  C. Engh,et al.  Effect of Femoral Stiffness on Bone Remodeling After Uncemented Arthroplasty , 2001, Clinical orthopaedics and related research.

[88]  Giuseppe Vairo,et al.  The influence of implant diameter and length on stress distribution of osseointegrated implants related to crestal bone geometry: a three-dimensional finite element analysis. , 2008, The Journal of prosthetic dentistry.

[89]  M. Peñarrocha-Diago,et al.  Influence of implant neck design and implant-abutment connection type on peri-implant health. Radiological study. , 2012, Clinical oral implants research.

[90]  Xi Ding,et al.  Implant-bone interface stress distribution in immediately loaded implants of different diameters: a three-dimensional finite element analysis. , 2009, Journal of prosthodontics : official journal of the American College of Prosthodontists.

[91]  Mamoru Mabuchi,et al.  Processing of biocompatible porous Ti and Mg , 2001 .

[92]  V. Rangarajan,et al.  Influence of Implant Surface Topography and Loading Condition on Stress Distribution in Bone Around Implants: A Comparative 3D FEA , 2011, Journal of Indian Prosthodontic Society.

[93]  S. Arzanpour,et al.  The effects of implant angulation on the resonance frequency of a dental implant. , 2014, Medical engineering & physics.

[94]  L Cristofolini,et al.  Mechanical validation of whole bone composite femur models. , 1996, Journal of biomechanics.

[95]  S. Cowin,et al.  Bone remodeling I: theory of adaptive elasticity , 1976 .

[96]  A. Uchida,et al.  Porous-surfaced metallic implants for orthopedic applications. , 1987 .

[97]  Alexei Bolshakov,et al.  A critical examination of the fundamental relations used in the analysis of nanoindentation data , 1999 .

[98]  L. Lanyon,et al.  The influence of strain rate on adaptive bone remodelling. , 1982, Journal of biomechanics.

[99]  Bert Van Rietbergen,et al.  Finite Element Analysis Based on In Vivo HR‐pQCT Images of the Distal Radius Is Associated With Wrist Fracture in Postmenopausal Women , 2007, Journal of bone and mineral research : the official journal of the American Society for Bone and Mineral Research.

[100]  G W Marshall,et al.  Mechanical properties of mineralized collagen fibrils as influenced by demineralization. , 2008, Journal of structural biology.

[101]  P Zioupos,et al.  Mechanical properties and the hierarchical structure of bone. , 1998, Medical engineering & physics.

[102]  F. Isidor,et al.  Histological evaluation of peri-implant bone at implants subjected to occlusal overload or plaque accumulation. , 1997, Clinical oral implants research.

[103]  Wurihan,et al.  Lysyl oxidase like-2 reinforces unsatisfactory ossification induced by bone morphogenetic protein-2: relating nanomechanical properties and molecular changes. , 2013, Nanomedicine : nanotechnology, biology, and medicine.

[104]  R. Ritchie,et al.  Reduced size-independent mechanical properties of cortical bone in high-fat diet-induced obesity. , 2010, Bone.

[105]  Michael V. Swain,et al.  Surface morphology optimization for osseointegration of coated implants. , 2010, Biomaterials.

[106]  F. Ghalichi,et al.  Effect of surface treatment on stress distribution in immediately loaded dental implants--a 3D finite element analysis. , 2014, Dental materials : official publication of the Academy of Dental Materials.

[107]  C. R. Howlett,et al.  The effect of different titanium and hydroxyapatite-coated dental implant surfaces on phenotypic expression of human bone-derived cells. , 2004, Journal of biomedical materials research. Part A.

[108]  M. McKee,et al.  Morphological and immunocytochemical characterization of primary osteogenic cell cultures derived from fetal rat cranial tissue , 1998, The Anatomical record.

[109]  R M Pilliar,et al.  The optimum pore size for the fixation of porous-surfaced metal implants by the ingrowth of bone. , 1980, Clinical orthopaedics and related research.

[110]  Niklaus P Lang,et al.  De novo alveolar bone formation adjacent to endosseous implants. , 2003, Clinical oral implants research.

[111]  M. McKee,et al.  Osteopontin at mineralized tissue interfaces in bone, teeth, and osseointegrated implants: Ultrastructural distribution and implications for mineralized tissue formation, turnover, and repair , 1996, Microscopy research and technique.