Validity of pulse oximetry during maximal exercise in normoxia, hypoxia, and hyperoxia.

During exercise, pulse oximetry is problematic due to motion artifact and altered digital perfusion. New pulse oximeter technology addresses these issues and may offer improved performance. We simultaneously compared Nellcor N-395 (Oxismart XLTM) pulse oximeters with an RS-10 forehead sensor (RS-10), a D-25 digit sensor (D-25), and the Ivy 2000 (Masimo SETTM)/LNOP-Adt digit sensor (Ivy) to arterial blood oxygen saturation (Sa(O(2))) by cooximetry. Nine normal subjects, six athletes, and four patients with chronic disease exercised to maximum oxygen consumption (VO(2 max)) under various conditions [normoxia, hypoxia inspired oxygen fraction (FI(O(2))) = 0.125; hyperoxia, FI(O(2)) = 1.0]. Regression analysis for normoxia and hypoxic data was performed (n = 161 observations, Sa(O(2)) = 73-99.9%), and bias (B) and precision (P) were calculated. RS10 offered greater validity than the other two devices tested (y = 1.009x - 0.52, R(2) = 0.90, B+/-P = 0.3 +/- 2.5). Finger sensors had low precision and a significant negative bias (D-25: y = 1.004x - 2.327, R(2) = 0.52, B+/-P = -2.0 +/- 7.3; Ivy: y = 1.237x - 24.2, R(2) = 0.78, B+/-P = -2.0 +/- 5.2). Eliminating measurements in which heart rate differed by >10 beats/min from the electrocardiogram value improved precision minimally and did not affect bias substantially (B+/-P = 0.5 +/- 2.0, -1.8 +/- 8.4, and -1.25+/-4.33 for RS-10, D-25, and Ivy, respectively). Signal detection algorithms and pulse oximeter were identical between RS-10 and D-25; thus the improved performance of the forehead sensor is likely because of sensor location. RS-10 should be considered for exercise testing in which pulse oximetry is desirable.

[1]  Inaccuracy of pulse oximetry in patients with severe tricuspid regurgitation , 1991, Anaesthesia.

[2]  S J Barker,et al.  Effects of Motion on the Performance of Pulse Oximeters in Volunteers , 1996, Anesthesiology.

[3]  R. G. Knowlton,et al.  Re-examination of the incidence of exercise-induced hypoxaemia in highly trained subjects. , 1993, British journal of sports medicine.

[4]  S. Barker,et al.  Effects of motion, ambient light, and hypoperfusion on pulse oximeter function , 1997 .

[5]  C. Bell,et al.  Effect of peripheral perfusion on accuracy of pulse oximetry in children. , 1999, Journal of clinical anesthesia.

[6]  S. Powers,et al.  Incidence of exercise induced hypoxemia in elite endurance athletes at sea level , 2004, European Journal of Applied Physiology and Occupational Physiology.

[7]  N. Pace,et al.  Delays in the detection of hypoxemia due to site of pulse oximetry probe placement. , 1999, Journal of clinical anesthesia.

[8]  S. Powers,et al.  Validity of pulse oximetry during exercise in elite endurance athletes. , 1992, Journal of applied physiology.

[9]  J A Dempsey,et al.  Exercise-induced arterial hypoxemia. , 1999, Journal of applied physiology.

[10]  Arthur S Slutsky,et al.  Ear oximetry during combined hypoxia and exercise. , 1986, Journal of applied physiology.

[11]  S. Powers,et al.  Accuracy of pulse oximetry to estimate HbO2 fraction of total Hb during exercise. , 1989, Journal of applied physiology.

[12]  R. Fronsko,et al.  Evaluation of the influence of movement on saturation readings from pulse oximeters , 1995, Anaesthesia.

[13]  R. K. Webb,et al.  Potential errors in pulse oximetry , 1991, Anaesthesia.

[14]  J. Severinghaus Simple, accurate equations for human blood O2 dissociation computations. , 1979, Journal of applied physiology: respiratory, environmental and exercise physiology.

[15]  D Martin,et al.  Effects of acute hypoxia on the VO2 max of trained and untrained subjects. , 1993, Journal of sports sciences.

[16]  Steven J. Barker,et al.  The effects of motion on the performance of pulse oximeters in volunteers (revised publication). , 1997 .