Impeller-pump model derived from conservation laws applied to the simulation of the cardiovascular system coupled to heart-assist pumps

Previous numerical models of impeller pumps for ventricular assist devices utilize curve-fitted polynomials to simulate experimentally-obtained pressure difference versus flow rate characteristics of the pumps, with pump rotational speed as a parameter. In this paper the numerical model for the pump pressure difference versus flow rate characteristics is obtained by analytic derivation. The mass, energy and angular momentum conservation laws are applied to the working fluid passing through the impeller geometry and coupled with the turbomachine's velocity diagram. This results in the construction of a pressure difference versus flow rate characteristic for the specific pump geometry, with pump rotational speed as parameter. Overall this model allows modifications of the pump geometry, so that the pump avoids undesirable operating conditions, such as regurgitant flow. The HeartMate III centrifugal pump is used as an example to demonstrate the application of the technique. The parameterised numerical model for HeartMate III derived by this technique is coupled with a numerical model for the human cardiovascular system, and the combination is used to investigate the cardiovascular response under different conditions of impeller pump support. Conditions resulting in regurgitant pump flow, the pump resulting in aortic valve closure and taking over completely the pumping action from the diseased heart, and inner ventricular wall suction at pump inlet are predicted by the model. The simulation results suggest that for normal HeartMate III operation the pump speed should be maintained between 3,100 and 4,500 rpm to avoid regurgitant pump flow and ventricular suction. To obtain optimal overall cardiovascular system plus pump response, the pump operating speed should be 3,800 rpm.

[1]  F. Ghorbel,et al.  A human cardiopulmonary system model applied to the analysis of the Valsalva maneuver. , 2001, American journal of physiology. Heart and circulatory physiology.

[2]  F P T Baaijens,et al.  A three-dimensional computational analysis of fluid-structure interaction in the aortic valve. , 2003, Journal of biomechanics.

[3]  X. Mueller,et al.  Atrial, ventricular, or both cannulation sites to optimize left ventricular assistance? , 2000 .

[4]  M. Ursino,et al.  A mathematical model of the carotid baroregulation in pulsating conditions , 1999, IEEE Transactions on Biomedical Engineering.

[5]  N H Lovell,et al.  Sensorless flow and head estimation in the VentrAssist rotary blood pump. , 2000, Artificial organs.

[6]  Burkert Pieske,et al.  Reverse remodeling in heart failure – fact or fiction? , 2004 .

[7]  F. Casas,et al.  Minimal Sensor Count Approach to Fuzzy Logic Rotary Blood Pump Flow Control , 2007, ASAIO journal.

[8]  K. Litwak,et al.  HeartMate III: Pump Design for a Centrifugal LVAD with a Magnetically Levitated Rotor , 2001, ASAIO journal.

[9]  T. Korakianitis,et al.  Numerical simulation of cardiovascular dynamics with healthy and diseased heart valves. , 2006, Journal of biomechanics.

[10]  Mikhail Skliar,et al.  Modeling and Control of a Brushless DC Axial Flow Ventricular Assist Device , 2002, ASAIO journal.

[11]  A. Guyton,et al.  Textbook of Medical Physiology , 1961 .

[12]  Yong Zhao,et al.  Numerical simulation of opening process in a bileaflet mechanical heart valve under pulsatile flow condition. , 2003, The Journal of heart valve disease.

[13]  P. Havlík,et al.  Study of left ventricular bypass using Wankel type semipulsatile blood pump. , 1998, Artificial organs.

[14]  Pascal Verdonck,et al.  Unloading effect of a rotary blood pump assessed by mathematical modeling. , 2003, Artificial organs.

[15]  Patrick Segers,et al.  Effect of rotary blood pump failure on left ventricular energetics assessed by mathematical modeling. , 2002, Artificial organs.

[16]  Nigel H Lovell,et al.  Noninvasive average flow estimation for an implantable rotary blood pump: a new algorithm incorporating the role of blood viscosity. , 2007, Artificial organs.

[17]  R. Benkowski,et al.  Suction Detection for the MicroMed DeBakey Left Ventricular Assist Device , 2005, ASAIO journal.

[18]  Georg Wieselthaler,et al.  Advanced suction detection for an axial flow pump. , 2006, Artificial organs.

[19]  J. Bai,et al.  Simulation study of the Hemopump as a cardiac assist device , 2002, Medical and Biological Engineering and Computing.

[20]  Tomoyuki Yambe,et al.  Outflow control for avoiding atrial suction in a continuous flow total artificial heart. , 2003, Artificial organs.

[21]  F. J. Wiesner A Review of Slip Factors for Centrifugal Impellers , 1967 .

[22]  M Ursino,et al.  The role of pressure pulsatility in the carotid baroreflex control: A computer simulation study , 1996, Comput. Biol. Medicine.

[23]  G. Giridharan,et al.  Ascending aorta outflow graft location and pulsatile ventricular assist provide optimal hemodynamic support in an adult mock circulation. , 2005, Artificial organs.

[24]  H Reul,et al.  Collected Nondimensional Performance of Rotary Dynamic Blood Pumps , 2004, ASAIO journal.

[25]  M. A. Rezaienia,et al.  The Effect of Geometry on the Efficiency and Hemolysis of Centrifugal Implantable Blood Pumps , 2017, ASAIO journal.

[27]  David Gordon Wilson,et al.  The design of high-efficiency turbomachinery and gas turbines , 1984 .

[28]  G Bearnson,et al.  Motor feedback physiological control for a continuous flow ventricular assist device. , 1999, Artificial organs.

[29]  James F. Antaki,et al.  A sensorless approach to control of a turbodynamic left ventricular assist system , 2001, IEEE Trans. Control. Syst. Technol..

[30]  M Komeda,et al.  Physical model-based indirect measurements of blood pressure and flow using a centrifugal pump. , 2000, Artificial organs.

[31]  D. G. Shepherd,et al.  Principles of Turbomachinery , 1956 .

[32]  Theodosios Korakianitis,et al.  Numerical simulation of cardiovascular dynamics with left heart failure and in-series pulsatile ventricular assist device. , 2006, Artificial organs.

[33]  Akbar Rahideh,et al.  Optimization of Centrifugal Pump Characteristic Dimensions for Mechanical Circulatory Support Devices , 2016, ASAIO journal.