Respiratory motion correction in positron emission tomography

In this thesis, we develop a motion correction method to overcome the degradation of image quality introduced by respiratory motion in positron emission tomography (PET), so that diagnostic performance for lung cancer can be improved. Lung cancer is currently the most common cause of cancer death both in the UK and in the world. PET/CT, which is a combination of PET and CT, providing clinicians with both functional and anatomical information, is routinely used as a non-invasive imaging technique to diagnose and stage lung cancer. However, since a PET scan normally takes 15-30 minutes, respiration is inevitable in data acquisition. As a result, thoracic PET images are substantially degraded by respiratory motion, not only by being blurred, but also by being inaccurately attenuation corrected due to the mismatch between PET and CT. If these challenges are not addressed, the diagnosis of lung cancer may be misled. The main contribution of this thesis is to propose a novel process for respiratory motion correction, in which non-attenuation corrected PET images (PET-NAC) are registered to a reference position for motion correction and then multiplied by a voxelwise attenuation correction factor (ACF) image for attenuation correction. The ACF image is derived from a CT image which matches the reference position, so that no attenuation correction artefacts would occur. In experiments, the motion corrected PET images show significant improvements over the uncorrected images, which represent the acquisitions typical of current clinical practice. The enhanced image quality means that our method has the potential to improve diagnostic performance for lung cancer. We also develop an automatic lesion detection method based on motion corrected images. A small lung lesion is only 2 or 3 voxels in diameter and of marginal contrast. It could easily be missed by human observers. Our method aims to provide radiologists with a map of potential lesions for decision so that diagnostic efficiency can be improved. It utilises both PET and CT images. The CT image provides a lung mask, to which lesion detection is confined, whereas the PET image provides distribution of glucose metabolism, according to which lung lesions are detected. Experimental results show that respiratory motion correction significantly increases the success of lesion detection, especially for small lesions, and most of the lung lesions can be detected by our method. The method can serve as a useful computer-aided image analysing tool to help radiologists read images and find malignant lung lesions. Finally, we explore the possibility of incorporating temporal information into respiratory motion correction. Conventionally, respiratory gated PET images are individually registered to the reference position. Temporal continuity across the respiratory period is not considered. We propose a spatio-temporal registration algorithm, which models temporally smooth deformation in order to improve the registration performance. However, we discover that the improvement introduced by temporal information is relatively small at the cost of a much longer computation time. Spatial registration with regularisation yields similar results but is superior in speed. Therefore, it is preferable for respiratory motion correction.

[1]  Balraj Naren,et al.  Medical Image Registration , 2022 .

[2]  Christophe Deroose,et al.  Performance of MAP Reconstruction for Hot Lesion Detection in Whole-Body PET/CT: An Evaluation With Human and Numerical Observers , 2009, IEEE Transactions on Medical Imaging.

[3]  Kris Thielemans,et al.  A framework for automated tumor detection in thoracic FDG pet images using texture-based features , 2009, 2009 IEEE International Symposium on Biomedical Imaging: From Nano to Macro.

[4]  O. Muzik,et al.  A novel computerized approach to enhancing lung tumor detection in whole-body PET images , 2004, The 26th Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[5]  Soo Chin Liew,et al.  Description of a simultaneous emission-transmission CT system , 1990, Medical Imaging.

[6]  Dean F. Wong,et al.  Accurate Event-Driven Motion Compensation in High-Resolution PET Incorporating Scattered and Random Events , 2008, IEEE Transactions on Medical Imaging.

[7]  Simon R. Cherry,et al.  Fast gradient-based methods for Bayesian reconstruction of transmission and emission PET images , 1994, IEEE Trans. Medical Imaging.

[8]  H. Malcolm Hudson,et al.  Accelerated image reconstruction using ordered subsets of projection data , 1994, IEEE Trans. Medical Imaging.

[9]  Richard M. Leahy,et al.  A theoretical study of the contrast recovery and variance of MAP reconstructions from PET data , 1999, IEEE Transactions on Medical Imaging.

[10]  C. Ling,et al.  Effect of respiratory gating on quantifying PET images of lung cancer. , 2002, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[11]  W A Kalender,et al.  Physical performance characteristics of spiral CT scanning. , 1991, Medical physics.

[12]  Adriaan A. Lammertsma,et al.  Noninvasive Quantification of Regional Myocardial Blood Flow in Coronary Artery Disease With Oxygen‐15–Labeled Carbon Dioxide Inhalation and Positron Emission Tomography , 1991, Circulation.

[13]  Ross Berbeco,et al.  Evaluation of the combined effects of target size, respiratory motion and background activity on 3D and 4D PET/CT images , 2008, Physics in medicine and biology.

[14]  Paul K Marsden,et al.  Simultaneous PET and NMR. , 2002, The British journal of radiology.

[15]  Guy Marchal,et al.  Multimodality image registration by maximization of mutual information , 1997, IEEE Transactions on Medical Imaging.

[16]  Michael Unser,et al.  Spatio-temporal nonrigid registration for ultrasound cardiac motion estimation , 2005, IEEE Transactions on Medical Imaging.

[17]  M. Lodge,et al.  Comparison of 2-dimensional and 3-dimensional acquisition for 18F-FDG PET oncology studies performed on an LSO-based scanner. , 2006, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[18]  Y. Erdi,et al.  FDG-PET standardized uptake values in normal anatomical structures using iterative reconstruction segmented attenuation correction and filtered back-projection , 2001, European Journal of Nuclear Medicine.

[19]  O. Muzik,et al.  Computerized Detection of Lung Tumors in PET/CT Images , 2006, 2006 International Conference of the IEEE Engineering in Medicine and Biology Society.

[20]  B. E. Peterson [Diagnosis of lung cancer]. , 1971, Voprosy onkologii.

[21]  Carole Lartizien,et al.  Automatic detection of active nodules in 3D PET oncology imaging using the Hotelling Observer and the Support Vector Machines: A comparison study , 2008, 2008 IEEE Nuclear Science Symposium Conference Record.

[22]  Daniel Rueckert,et al.  Nonrigid registration using free-form deformations: application to breast MR images , 1999, IEEE Transactions on Medical Imaging.

[23]  Randall K Ten Haken,et al.  Using fluorodeoxyglucose positron emission tomography to assess tumor volume during radiotherapy for non-small-cell lung cancer and its potential impact on adaptive dose escalation and normal tissue sparing. , 2009, International journal of radiation oncology, biology, physics.

[24]  N. Otsu A threshold selection method from gray level histograms , 1979 .

[25]  Benjamin M. W. Tsui,et al.  MCAT to XCAT: The Evolution of 4-D Computerized Phantoms for Imaging Research , 2009, Proceedings of the IEEE.

[26]  Loïc Fin,et al.  Motion correction based on an appropriate system matrix for statistical reconstruction of respiratory-correlated PET acquisitions , 2009, Comput. Methods Programs Biomed..

[27]  P. Piccini,et al.  Applications of positron emission tomography (PET) in neurology , 2004, Journal of Neurology, Neurosurgery & Psychiatry.

[28]  John W. Clark,et al.  A motion-incorporated reconstruction method for gated PET studies , 2006, Physics in medicine and biology.

[29]  Sébastien Ourselin,et al.  Fast free-form deformation using graphics processing units , 2010, Comput. Methods Programs Biomed..

[30]  Thomas Beyer,et al.  Dual-modality PET/CT imaging: the effect of respiratory motion on combined image quality in clinical oncology , 2003, European Journal of Nuclear Medicine and Molecular Imaging.

[31]  Bram van Ginneken,et al.  Toward automated segmentation of the pathological lung in CT , 2005, IEEE Transactions on Medical Imaging.

[32]  Georges El Fakhri,et al.  Impact of Acquisition Geometry, Image Processing, and Patient Size on Lesion Detection in Whole-Body 18F-FDG PET , 2007, Journal of Nuclear Medicine.

[33]  Yuji Nakamoto,et al.  Clinically significant inaccurate localization of lesions with PET/CT: frequency in 300 patients. , 2003, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[34]  Thomas Beyer,et al.  Non-small cell lung cancer: dual-modality PET/CT in preoperative staging. , 2003, Radiology.

[35]  D. Mankoff,et al.  Hypoxia imaging-directed radiation treatment planning , 2006, European Journal of Nuclear Medicine and Molecular Imaging.

[36]  S S Gambhir,et al.  ROC and localization ROC analyses of lesion detection in whole-body FDG PET: effects of acquisition mode, attenuation correction and reconstruction algorithm. , 1999, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[37]  Alan Perkins,et al.  Diagnostic imaging , 2012, The Lancet.

[38]  J Bogaert,et al.  Lymph node staging in non-small-cell lung cancer with FDG-PET scan: a prospective study on 690 lymph node stations from 68 patients. , 1998, Journal of clinical oncology : official journal of the American Society of Clinical Oncology.

[39]  Heinz-Peter Schlemmer,et al.  PET/MRI: Paving the Way for the Next Generation of Clinical Multimodality Imaging Applications , 2010, Journal of Nuclear Medicine.

[40]  C. Tsoumpas,et al.  STIR: software for tomographic image reconstruction release 2 , 2012, 2006 IEEE Nuclear Science Symposium Conference Record.

[41]  S. Nehmeh,et al.  An iterative technique to segment PET lesions using a Monte Carlo based mathematical model. , 2009, Medical physics.

[42]  R. Coleman,et al.  Focal pulmonary abnormalities: evaluation with F-18 fluorodeoxyglucose PET scanning. , 1993, Radiology.

[43]  W. Bradley,et al.  MRI: The Basics , 1997 .

[44]  Jean-Claude Liehn,et al.  Respiratory motion artefact in the liver dome on FDG PET/CT: comparison of attenuation correction with CT and a caesium external source , 2005, European Journal of Nuclear Medicine and Molecular Imaging.

[45]  Benjamin M. W. Tsui,et al.  Modeling respiratory mechanics in the MCAT and spline-based MCAT phantoms , 1999 .

[46]  Stephen M. Smith,et al.  SUSAN—A New Approach to Low Level Image Processing , 1997, International Journal of Computer Vision.

[47]  C A Giannone Monte Carlo Calculations in Nuclear Medicine: Applications in Diagnostic Imaging , 1999 .

[48]  C. Mountain,et al.  Revisions in the International System for Staging Lung Cancer. , 1997, Chest.

[49]  Ronald H. Huesman,et al.  List mode reconstruction for PET with motion compensation: a simulation study , 2002, Proceedings IEEE International Symposium on Biomedical Imaging.

[50]  D. Lalush,et al.  A realistic spline-based dynamic heart phantom , 1998, 1998 IEEE Nuclear Science Symposium Conference Record. 1998 IEEE Nuclear Science Symposium and Medical Imaging Conference (Cat. No.98CH36255).

[51]  A J Reader,et al.  List-mode-based reconstruction for respiratory motion correction in PET using non-rigid body transformations , 2007, Physics in medicine and biology.

[52]  M Dahlbom,et al.  A comparison of normalization effects on three whole-body cylindrical 3D PET systems. , 2000, Physics in medicine and biology.

[53]  Ursula Nestle,et al.  Nonrigid Versus Rigid Registration of Thoracic 18F-FDG PET and CT in Patients with Lung Cancer: An Intraindividual Comparison of Different Breathing Maneuvers , 2009, Journal of Nuclear Medicine.

[54]  Michael E. Phelps,et al.  PET: Molecular Imaging and Its Biological Applications , 2004 .

[55]  Paul Kinahan,et al.  A combined PET/CT scanner for clinical oncology. , 2000, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[56]  Donald W. Wilson,et al.  Noise properties of the EM algorithm. I. Theory , 1994 .

[57]  Edgar Arce Santana,et al.  Hidden Markov Measure Field Models for Image Segmentation , 2003, IEEE Trans. Pattern Anal. Mach. Intell..

[58]  K Thielemans,et al.  Enhanced 3D PET OSEM reconstruction using inter-update Metz filtering. , 2000, Physics in medicine and biology.

[59]  Michael F. McNitt-Gray,et al.  Method for segmenting chest CT image data using an anatomical model: preliminary results , 1997, IEEE Transactions on Medical Imaging.

[60]  Daniel Rueckert,et al.  Automatic tracking of the aorta in cardiovascular MR images using deformable models , 1997, IEEE Transactions on Medical Imaging.

[61]  F. Fazio,et al.  Scatter correction techniques in 3D PET: a Monte Carlo evaluation , 1998, 1998 IEEE Nuclear Science Symposium Conference Record. 1998 IEEE Nuclear Science Symposium and Medical Imaging Conference (Cat. No.98CH36255).

[62]  R. Castillo,et al.  Attenuation correction of PET images with respiration-averaged CT images in PET/CT. , 2005, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[63]  J. Fessler,et al.  Joint estimation of image and deformation parameters in motion-corrected PET , 2003, 2003 IEEE Nuclear Science Symposium. Conference Record (IEEE Cat. No.03CH37515).

[64]  Thomas Beyer,et al.  A combined PET/CT scanner: the path to true image fusion. , 2002, The British journal of radiology.

[65]  M. Mandelkern,et al.  Accuracy and clinical impact of mediastinal lymph node staging with FDG-PET imaging in potentially resectable lung cancer. , 2000, American journal of clinical oncology.

[66]  David J. Hawkes,et al.  Validation of nonrigid image registration using finite-element methods: application to breast MR images , 2003, IEEE Transactions on Medical Imaging.

[67]  David W Townsend,et al.  Dual-Modality Imaging: Combining Anatomy and Function* , 2008, Journal of Nuclear Medicine.

[68]  Michael Brady,et al.  A Biomathematical Modeling Approach to Central Nervous System Radioligand Discovery and Development , 2009, Journal of Nuclear Medicine.

[69]  R. Leahy,et al.  High-resolution 3D Bayesian image reconstruction using the microPET small-animal scanner. , 1998, Physics in medicine and biology.

[70]  David Gavaghan,et al.  Relationship between structural changes and hyperpolarized gas magnetic resonance imaging in chronic obstructive pulmonary disease using computational simulations with realistic alveolar geometry , 2009, Philosophical Transactions of the Royal Society A: Mathematical, Physical and Engineering Sciences.

[71]  Xiaoyi Jiang,et al.  Lung motion correction on respiratory gated 3-D PET/CT images , 2006, IEEE Transactions on Medical Imaging.

[72]  V. Roggli,et al.  Solitary pulmonary nodules: Part I. Morphologic evaluation for differentiation of benign and malignant lesions. , 2000, Radiographics : a review publication of the Radiological Society of North America, Inc.

[73]  A. Alavi,et al.  Use of a corrected standardized uptake value based on the lesion size on CT permits accurate characterization of lung nodules on FDG-PET , 2002, European Journal of Nuclear Medicine and Molecular Imaging.

[74]  John W. Clark,et al.  Implementation of an Automated Respiratory Amplitude Gating Technique for PET/CT: Clinical Evaluation , 2010, Journal of Nuclear Medicine.

[75]  Vincent Frouin,et al.  Correction of partial-volume effect for PET striatal imaging: fast implementation and study of robustness. , 2002, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[76]  D. Newport,et al.  Evaluation of simulation-based scatter correction for 3-D PET cardiac imaging , 1995 .

[77]  John F. Nunn,et al.  Respiratory Physiology—the essentials , 1975 .

[78]  Paul Kinahan,et al.  Attenuation correction for a combined 3D PET/CT scanner. , 1998, Medical physics.

[79]  Heinrich R Schelbert,et al.  Improvements in cancer staging with PET/CT: literature-based evidence as of September 2006. , 2007, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[80]  Lee-Tzuu Chang,et al.  A Method for Attenuation Correction in Radionuclide Computed Tomography , 1978, IEEE Transactions on Nuclear Science.

[81]  Yu Chi Hu,et al.  Radiotherapy treatment planning for patients with non-small cell lung cancer using positron emission tomography (PET). , 2002, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[82]  M O Leach,et al.  Implications of respiratory motion for the quantification of 2D MR spectroscopic imaging data in the abdomen. , 2000, Physics in medicine and biology.

[83]  Guobao Wang,et al.  Generalized Algorithms for Direct Reconstruction of Parametric Images From Dynamic PET Data , 2009, IEEE Transactions on Medical Imaging.

[84]  Vincent J. Cunningham,et al.  Parametric Imaging of Ligand-Receptor Binding in PET Using a Simplified Reference Region Model , 1997, NeuroImage.

[85]  Roger Fulton,et al.  The design and implementation of a motion correction scheme for neurological PET. , 2003, Physics in medicine and biology.

[86]  Michael Unser,et al.  Fast B-spline Transforms for Continuous Image Representation and Interpolation , 1991, IEEE Trans. Pattern Anal. Mach. Intell..

[87]  D. Perperidis,et al.  Spatio-temporal registration and modelling of the heart using cardiovascular MR imaging. , 2006 .

[88]  G. J. Klein,et al.  Real-time system for respiratory-cardiac gating in positron tomography , 1997 .

[89]  Christopher G. Harris,et al.  A Combined Corner and Edge Detector , 1988, Alvey Vision Conference.

[90]  I. Buvat,et al.  Partial-Volume Effect in PET Tumor Imaging* , 2007, Journal of Nuclear Medicine.

[91]  T. Turkington,et al.  Clinical applications of PET in oncology. , 2004, Radiology.

[92]  Michael Unser,et al.  Optimization of mutual information for multiresolution image registration , 2000, IEEE Trans. Image Process..

[93]  David S. Lalush,et al.  Effects of upward creep and respiratory motion in myocardial SPECT , 2000 .

[94]  H. K. Huang Handbook of Medical Imaging. Volumes 1-3 , 2007 .

[95]  Ronald H. Huesman,et al.  Theoretical study of lesion detectability of MAP reconstruction using computer observers , 2001, IEEE Transactions on Medical Imaging.

[96]  M. Wernick,et al.  Fast spatio-temporal image reconstruction for dynamic PET , 1999, IEEE Transactions on Medical Imaging.

[97]  Florian Büther,et al.  Optimal number of respiratory gates in positron emission tomography: a cardiac patient study. , 2009, Medical physics.

[98]  P. Danielsson Euclidean distance mapping , 1980 .

[99]  J. H. Scarffe,et al.  Cancer Medicine , 1982, British Journal of Cancer.

[100]  Jeffrey A. Fessler,et al.  Statistical image reconstruction methods for randoms-precorrected PET scans , 1998, Medical Image Anal..

[101]  P. Kaufmann,et al.  Myocardial blood flow measurement by PET: technical aspects and clinical applications. , 2005, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[102]  Cyrill Burger,et al.  Respiration-induced attenuation artifact at PET/CT: technical considerations. , 2003, Radiology.

[103]  E. Veklerov,et al.  Stopping Rule for the MLE Algorithm Based on Statistical Hypothesis Testing , 1987, IEEE Transactions on Medical Imaging.

[104]  Alejandro F. Frangi,et al.  Model-based quantitation of 3-D magnetic resonance angiographic images , 1999, IEEE Transactions on Medical Imaging.

[105]  Valerie Treyer,et al.  CT vs 68Ge attenuation correction in a combined PET/CT system: evaluation of the effect of lowering the CT tube current , 2002, European Journal of Nuclear Medicine and Molecular Imaging.

[106]  Xiaoyi Jiang,et al.  Respiratory Motion Correction in 3-D PET Data With Advanced Optical Flow Algorithms , 2008, IEEE Transactions on Medical Imaging.

[107]  Alan C. Evans,et al.  Creation and Application of a Simulated Database of Dynamic [$^18$F]MPPF PET Acquisitions Incorporating Inter-Individual Anatomical and Biological Variability , 2006, IEEE Transactions on Medical Imaging.

[108]  Olivier Barret,et al.  Monte Carlo simulation and scatter correction of the GE Advance PET scanner with SimSET and Geant4 , 2005, Physics in medicine and biology.

[109]  W. K. Luk,et al.  Adaptive segmented attenuation correction for whole-body PET imaging , 1994, Proceedings of 1994 IEEE Nuclear Science Symposium - NSS'94.

[110]  Gary E. Christensen,et al.  Consistent image registration , 2001, IEEE Transactions on Medical Imaging.

[111]  Tony Lindeberg,et al.  Detecting salient blob-like image structures and their scales with a scale-space primal sketch: A method for focus-of-attention , 1993, International Journal of Computer Vision.

[112]  Carsten Steger,et al.  An Unbiased Detector of Curvilinear Structures , 1998, IEEE Trans. Pattern Anal. Mach. Intell..

[113]  Lei Xing,et al.  Model-based image reconstruction for four-dimensional PET. , 2006, Medical physics.

[114]  C. Lartizien,et al.  Monte Carlo simulation-based design study of a LSO-LuAP small animal PET system , 2002, 2002 IEEE Nuclear Science Symposium Conference Record.

[115]  Wei Chen Clinical Applications of PET in Brain Tumors* , 2007, Journal of Nuclear Medicine.

[116]  Cordelia Schmid,et al.  Scale & Affine Invariant Interest Point Detectors , 2004, International Journal of Computer Vision.

[117]  Joel S Karp,et al.  Optimization of a fully 3D single scatter simulation algorithm for 3D PET. , 2004, Physics in medicine and biology.

[118]  J. Besag Spatial Interaction and the Statistical Analysis of Lattice Systems , 1974 .

[119]  Jens Krause,et al.  Spiral Interpolation Algorithms for Multi-Slice Spiral CT Part I: Theory , 2000, IEEE Trans. Medical Imaging.

[120]  T. Hebert,et al.  A generalized EM algorithm for 3-D Bayesian reconstruction from Poisson data using Gibbs priors. , 1989, IEEE transactions on medical imaging.

[121]  E. Kim PET and PET/CT: A Clinical Guide , 2010, Journal of Nuclear Medicine.

[122]  Jeffrey A. Fessler,et al.  Emission image reconstruction for randoms-precorrected PET allowing negative sinogram values , 2004, IEEE Transactions on Medical Imaging.

[123]  J. Mazziotta,et al.  MRI‐PET Registration with Automated Algorithm , 1993, Journal of computer assisted tomography.

[124]  Richard M. Leahy,et al.  Lesion Detection in Dynamic FDG-PET Using Matched Subspace Detection , 2009, IEEE Transactions on Medical Imaging.

[125]  Sung-Cheng Huang,et al.  Anatomy of SUV , 2000 .

[126]  Jeffrey A. Fessler Penalized weighted least-squares image reconstruction for positron emission tomography , 1994, IEEE Trans. Medical Imaging.

[127]  Philippe Lambin,et al.  Effects of radiotherapy planning with a dedicated combined PET-CT-simulator of patients with non-small cell lung cancer on dose limiting normal tissues and radiation dose-escalation: a planning study. , 2005, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[128]  T. Spinks,et al.  Correction for scatter in 3D brain PET using a dual energy window method. , 1996, Physics in medicine and biology.

[129]  A. Buck,et al.  PET attenuation coefficients from CT images: experimental evaluation of the transformation of CT into PET 511-keV attenuation coefficients , 2002, European Journal of Nuclear Medicine and Molecular Imaging.

[130]  Aggelos K. Katsaggelos,et al.  Bayesian and regularization methods for hyperparameter estimation in image restoration , 1999, IEEE Trans. Image Process..

[131]  W. J. Lorenz,et al.  Measured attenuation correction methods , 2004, European Journal of Nuclear Medicine.

[132]  Stefaan Vandenberghe,et al.  Fast reconstruction of 3D time-of-flight PET data by axial rebinning and transverse mashing , 2006, Physics in medicine and biology.

[133]  L. Shepp,et al.  Maximum Likelihood Reconstruction for Emission Tomography , 1983, IEEE Transactions on Medical Imaging.

[134]  Michael Unser,et al.  Fast parametric elastic image registration , 2003, IEEE Trans. Image Process..

[135]  Tomio Inoue,et al.  Use of PET and PET/CT for radiation therapy planning: IAEA expert report 2006-2007. , 2009, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[136]  Rachid Deriche,et al.  Vector-valued image regularization with PDEs: a common framework for different applications , 2005, IEEE Transactions on Pattern Analysis and Machine Intelligence.

[137]  Tony Lindeberg,et al.  Feature Detection with Automatic Scale Selection , 1998, International Journal of Computer Vision.

[138]  O. Schober,et al.  Respiratory gating in positron emission tomography: a quantitative comparison of different gating schemes. , 2007, Medical physics.

[139]  E. Hoffman,et al.  A Monte Carlo correction for the effect of Compton scattering in 3-D PET brain imaging , 1995 .

[140]  Haiying Liu,et al.  A Generic Framework for Non-rigid Registration Based on Non-uniform Multi-level Free-Form Deformations , 2001, MICCAI.

[141]  B. Sobel,et al.  Cardiac Positron Emission Tomography , 1996, Developments in Cardiovascular Medicine.

[142]  Richard M. Leahy,et al.  Statistical approaches in quantitative positron emission tomography , 2000, Stat. Comput..

[143]  Joel Karp,et al.  Consensus recommendations for the use of 18F-FDG PET as an indicator of therapeutic response in patients in National Cancer Institute Trials. , 2006, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[144]  Magnus Dahlbom,et al.  PET system calibrations and corrections for quantitative and spatially accurate images , 1989 .

[145]  G. Gindi,et al.  Noise analysis of MAP-EM algorithms for emission tomography. , 1997, Physics in medicine and biology.

[146]  Slobodan Devic,et al.  Impact of FDG-PET/CT on radiotherapy volume delineation in non-small-cell lung cancer and correlation of imaging stage with pathologic findings. , 2008, International journal of radiation oncology, biology, physics.

[147]  W. F. Jones Real-time event stream correction for patient motion in clinical 3-D PET , 2001, 2001 IEEE Nuclear Science Symposium Conference Record (Cat. No.01CH37310).

[148]  Carlos Ortiz-de-Solorzano,et al.  Consistent and Elastic Registration of Histological Sections Using Vector-Spline Regularization , 2006, CVAMIA.

[149]  V. Treyer,et al.  Objective and subjective comparison of standard 2-D and fully 3-D reconstructed data on a PET/CT system , 2007, Nuclear medicine communications.

[150]  C. C. Huang,et al.  Computer-aided lesion detection with statistical model-based features in PET images , 1997 .

[151]  R. Fulton,et al.  Accuracy of motion correction methods for PET brain imaging , 2004, IEEE Symposium Conference Record Nuclear Science 2004..

[152]  Tinsu Pan,et al.  Frequent Diagnostic Errors in Cardiac PET/CT Due to Misregistration of CT Attenuation and Emission PET Images: A Definitive Analysis of Causes, Consequences, and Corrections , 2007, Journal of Nuclear Medicine.

[153]  E. Beckmann,et al.  CT scanning the early days. , 2006, The British journal of radiology.

[154]  Isabelle Bloch,et al.  Explicit Incorporation of Prior Anatomical Information Into a Nonrigid Registration of Thoracic and Abdominal CT and 18-FDG Whole-Body Emission PET Images , 2007, IEEE Transactions on Medical Imaging.

[155]  S. Vandenberghe,et al.  MRI-Based Attenuation Correction for PET/MRI Using Ultrashort Echo Time Sequences , 2010, Journal of Nuclear Medicine.

[156]  M. Unser,et al.  Interpolation Revisited , 2000, IEEE Trans. Medical Imaging.

[157]  Michael Brady,et al.  Regularized B-spline deformable registration for respiratory motion correction in PET images , 2008, 2008 IEEE Nuclear Science Symposium Conference Record.

[158]  J R Perry,et al.  Correction of nonuniform attenuation in cardiac SPECT imaging. , 1989, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[159]  H. Sipilä,et al.  Dual gated PET/CT imaging of small targets of the heart: Method description and testing with a dynamic heart phantom , 2010, Journal of nuclear cardiology : official publication of the American Society of Nuclear Cardiology.

[160]  R. Huesman,et al.  Non-rigid summing of gated PET via optical flow , 1996 .

[161]  Eric A. Hoffman,et al.  Automatic lung segmentation for accurate quantitation of volumetric X-ray CT images , 2001, IEEE Transactions on Medical Imaging.

[162]  William H. Press,et al.  Numerical recipes in C , 2002 .

[163]  大隈 智尚 18F-FDG small-animal PET for monitoring the therapeutic effect of CT-guided radiofrequency ablation on implanted VX2 lung tumors in rabbits , 2006 .

[164]  Ran Klein,et al.  Quantification of myocardial blood flow with 82Rb dynamic PET imaging , 2007, European Journal of Nuclear Medicine and Molecular Imaging.

[165]  A. Rahmim,et al.  Motion compensation in histogram-mode and list-mode EM reconstructions: beyond the event-driven approach , 2004, IEEE Transactions on Nuclear Science.

[166]  M Slifstein,et al.  Models and methods for derivation of in vivo neuroreceptor parameters with PET and SPECT reversible radiotracers. , 2001, Nuclear medicine and biology.

[167]  Daniel Rueckert,et al.  Fast Spatio-temporal Free-Form Registration of Cardiac MR Image Sequences , 2004, FIMH.

[168]  Harold L. Kundel,et al.  Handbook of Medical Imaging, Volume 1. Physics and Psychophysics , 2000 .

[169]  Simon R. Cherry,et al.  Effects of scatter on model parameter estimates in 3D PET studies of the human brain , 1995 .

[170]  D. Rubin,et al.  Maximum likelihood from incomplete data via the EM - algorithm plus discussions on the paper , 1977 .

[171]  A. Rahmim,et al.  Data Processing Methods for a High Throughput Brain Imaging PET Research Center , 2006, 2006 IEEE Nuclear Science Symposium Conference Record.

[172]  David J. Hawkes,et al.  A New Validation Method for X-ray Mammogram Registration Algorithms Using a Projection Model of Breast X-ray Compression , 2007, IEEE Transactions on Medical Imaging.

[173]  A. Pevsner,et al.  The CT motion quantitation of lung lesions and its impact on PET-measured SUVs. , 2004, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[174]  Stefan Eberl,et al.  Lung segmentation and tumor detection from CT thorax volumes of FDG PET-CT scans by template registration and incorporation of functional information , 2008, 2008 IEEE Nuclear Science Symposium Conference Record.

[175]  F. Fazio,et al.  Implementation and evaluation of a 3D one-step late reconstruction algorithm for 3D positron emission tomography brain studies using median root prior , 2002, European Journal of Nuclear Medicine and Molecular Imaging.

[176]  Val J Lowe,et al.  NEMA NU 2-2001 performance measurements of an LYSO-based PET/CT system in 2D and 3D acquisition modes. , 2006, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[177]  S. Armato,et al.  Automated lung segmentation for thoracic CT impact on computer-aided diagnosis. , 2004, Academic radiology.

[178]  W. Segars,et al.  Study of the efficacy of respiratory gating in myocardial SPECT using the new 4-D NCAT phantom , 2001 .

[179]  T. Zeffiro,et al.  Head movement in normal subjects during simulated PET brain imaging with and without head restraint. , 1994, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[180]  Osman Ratib,et al.  Comparison between 18F-FDG PET, in-line PET/CT, and software fusion for restaging of recurrent colorectal cancer. , 2005, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[181]  D. Faul,et al.  A method for calibrating the CT-based attenuation correction of PET in human tissue , 2004, IEEE Transactions on Nuclear Science.

[182]  D M Hansell,et al.  PET/CT in non-small cell lung cancer staging-promises and problems. , 2007, Clinical radiology.

[183]  Piotr J Slomka,et al.  Software Approach to Merging Molecular with Anatomic Information , 2004 .

[184]  Jeffrey A. Fessler Mean and variance of implicitly defined biased estimators (such as penalized maximum likelihood): applications to tomography , 1996, IEEE Trans. Image Process..

[185]  Nassir Navab,et al.  Dual cardiac–respiratory gated PET: implementation and results from a feasibility study , 2007, European Journal of Nuclear Medicine and Molecular Imaging.

[186]  Bram van Ginneken,et al.  Computer analysis of computed tomography scans of the lung: a survey , 2006, IEEE Transactions on Medical Imaging.

[187]  Carole Lartizien,et al.  A lesion detection observer study comparing 2-dimensional versus fully 3-dimensional whole-body PET imaging protocols. , 2004, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[188]  S. Siegel,et al.  Implementation and evaluation of a calculated attenuation correction for PET , 1991, Conference Record of the 1991 IEEE Nuclear Science Symposium and Medical Imaging Conference.

[189]  Richard M. Leahy,et al.  Controlling Familywise Error Rate for Matched Subspace Detection in Dynamic FDG PET , 2009, IEEE Transactions on Medical Imaging.

[190]  A J Reader,et al.  Respiratory motion correction for PET oncology applications using affine transformation of list mode data , 2007, Physics in medicine and biology.

[191]  B. Tsui,et al.  Noise properties of the EM algorithm: II. Monte Carlo simulations. , 1994, Physics in medicine and biology.

[192]  J. Ollinger Model-based scatter correction for fully 3D PET. , 1996, Physics in medicine and biology.

[193]  Yasuyoshi Watanabe,et al.  18F-FDG small-animal PET for monitoring the therapeutic effect of CT-guided radiofrequency ablation on implanted VX2 lung tumors in rabbits. , 2006, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[194]  Bernhard Schölkopf,et al.  MRI-Based Attenuation Correction for PET/MRI: A Novel Approach Combining Pattern Recognition and Atlas Registration , 2008, Journal of Nuclear Medicine.

[195]  S. Cherry,et al.  A study of artefacts in simultaneous PET and MR imaging using a prototype MR compatible PET scanner. , 1999, Physics in medicine and biology.

[196]  Jorge Núñez de Murga,et al.  Statistical analysis of maximum likelihood estimator images of human brain FDG PET studies , 1993, IEEE Trans. Medical Imaging.

[197]  F. Guillemin,et al.  Management of SPN in France. Pathways for definitive diagnosis of solitary pulmonary nodule: a multicentre study in 18 French districts , 2008, BMC Cancer.

[198]  Alan C. Evans,et al.  PET-SORTEO: a Monte Carlo-based Simulator with high count rate capabilities , 2004, IEEE Transactions on Nuclear Science.

[199]  John F. Canny,et al.  A Computational Approach to Edge Detection , 1986, IEEE Transactions on Pattern Analysis and Machine Intelligence.

[200]  Maurizio Conti,et al.  Experimental comparison of lesion detectability for four fully-3D PET reconstruction schemes , 2007, 2007 IEEE Nuclear Science Symposium Conference Record.