Heating of metallic implants and instruments induced by gradient switching in a 1.5-Tesla whole-body unit.

PURPOSE To examine gradient switching-induced heating of metallic parts. MATERIALS AND METHODS Copper and titanium frames and sheets ( approximately 50 x 50 mm(2), 1.5 mm thick, frame width = 3 mm) surrounded by air were positioned in the scanner perpendicular to the static field horizontally 20 cm off-center. During the execution of a sequence (three-dimensional [3D] true fast imaging with steady precession [True-FISP], TR = 6.4 msec) exploiting the gradient capabilities (maximum gradient = 40 mT/m, maximum slew rate = 200 T/m/second), heating was measured with an infrared camera. Radio frequency (RF) amplitude was set to zero volts. Heating of a copper frame with a narrowing to 1 mm over 20 mm at one side was examined in air and in addition surrounded by several liters of gelled saline using fiber-optic thermography. Further heating studies were performed using an artificial hip made of titanium, and an aluminum replica of the hip prosthesis with the same geometry. RESULTS For the copper specimens, considerable heating (>10 degrees C) in air and in gelled saline (>1.2 degrees C) could be observed. Heating of the titanium specimens was markedly less ( approximately 1 degrees C in air). For the titanium artificial hip no heating could be detected, while the rise in temperature for the aluminum replica was approximately 2.2 degrees C. CONCLUSION Heating of more than 10 degrees C solely due to gradient switching without any RF irradiation was demonstrated in isolated copper wire frames. Under specific conditions (high gradient duty cycle, metallic loop of sufficient inductance and low resistance, power matching) gradient switching-induced heating of conductive specimens must be considered.

[1]  B. Schueler,et al.  MRI compatibility and visibility assessment of implantable medical devices , 1999, Journal of magnetic resonance imaging : JMRI.

[2]  F. Shellock,et al.  MRI of cervical fixation devices: Sensation of heating caused by vibration of metallic components , 1997, Journal of magnetic resonance imaging : JMRI.

[3]  F G Shellock,et al.  Implantable spinal fusion stimulator: Assessment of MR safety and artifacts , 2000, Journal of magnetic resonance imaging : JMRI.

[4]  F G Shellock,et al.  Metallic neurosurgical implants: Evaluation of magnetic field interactions, heating, and artifacts at 1.5‐Tesla , 2001, Journal of magnetic resonance imaging : JMRI.

[5]  Fritz Schick,et al.  Eddy‐current induction in extended metallic parts as a source of considerable torsional moment , 2006, Journal of magnetic resonance imaging : JMRI.

[6]  Jean A. Tkach,et al.  Evaluation of specific absorption rate as a dosimeter of MRI‐related implant heating , 2004, Journal of magnetic resonance imaging : JMRI.

[7]  Alexander V. Kildishev,et al.  Interactions of magnetic resonance imaging radio frequency magnetic fields with elongated medical implants , 2000 .

[8]  B. Condon,et al.  Investigation of the factors responsible for burns during MRI , 2001, Journal of magnetic resonance imaging : JMRI.

[9]  D N Buechler,et al.  Calculation of electric fields induced near metal implants by magnetic resonance imaging switched-gradient magnetic fields. , 1997, Magnetic resonance imaging.

[10]  F G Shellock,et al.  Metallic surgical instruments for interventional MRI procedures: Evaluation of MR safety , 2001, Journal of magnetic resonance imaging : JMRI.

[11]  W. Nitz,et al.  On the heating of linear conductive structures as guide wires and catheters in interventional MRI , 2001, Journal of magnetic resonance imaging : JMRI.

[12]  J. Reilly,et al.  Theoretical evaluation of peripheral nerve stimulation during MRI with an implanted spinal fusion stimulator. , 1997, Magnetic resonance imaging.

[13]  C J Bakker,et al.  Heating Around Intravascular Guidewires by Resonating RF Waves , 2000, Journal of magnetic resonance imaging : JMRI.