High speed acquisition of multiparameter data using a Macintosh IIcx
暂无分享,去创建一个
Abstract Accelerator mass spectrometry systems based on > 3 MV tandem accelerators often use multianode ionization detectors and/or time-of-flight detectors to identify individual isotopes through multiparameter analysis. A Macintosh IIcx has been programmed to collect AMS data from a CAMAC-implemented analyzer and to display the histogrammed individual parameters and a doubleparameter array. The computer-CAMAC connection is through a NuBus to CAMAC dataway interface which allows direct addressing to all functions and registers in the crate. Asynchronous data from the rare isotope are sorted into a CAMAC memory module by a list sequence controller. Isotope switching is controlled by a one-cycle timing generator. A rate-dependent amount of time is used to transfer the data from the memory module at the end of each timing cycle. The present configuration uses 10–75 ms for rates of 500–10000 cps. Parameter analysis occurs during the rest of the 520 ms data collection cycle. Completed measurements of the isotope concentrations of each sample are written to files which are compatible with standard Macintosh databases or other processing programs. The system is inexpensive and operates at speeds comparable to those obtainable using larger computers.