Study of 403.5 MHz Path Loss Models for Indoor Wireless Communications with Implanted Medical Devices on the Human Body

This paper contains simulated and measured data for 402-405 MHz radio propagation path loss in the consultation room for the allocated Medical Implant Communication System (MICS) band. The propagation models have been developed based on the number of partitions, concrete walls and objects between the transmitter and receiver. Unfurnished and furnished rooms were studied for indoor path loss and room penetration loss in a narrow band measurement. The received signals were measured, and effects from the indoor environment were evaluated to determine accurate impacts on the communication system. The fading in path loss for unfurnished and furnished indoor models with different polarizations was also considered. The path loss from the proposed models was illustrated and compared with the free space model. In this paper, the indoor wave propagation at the 403.5 MHz band was studied with both simulations and measurements to provide information that may aid the development of futuristic indoor communication for biotelemetry systems.

[1]  Paul Lukowicz,et al.  AMON: a wearable multiparameter medical monitoring and alert system , 2004, IEEE Transactions on Information Technology in Biomedicine.

[2]  H.T. Friis,et al.  A Note on a Simple Transmission Formula , 1946, Proceedings of the IRE.

[3]  E. Topsakal,et al.  Design of a Dual-Band Implantable Antenna and Development of Skin Mimicking Gels for Continuous Glucose Monitoring , 2008, IEEE Transactions on Microwave Theory and Techniques.

[4]  Shigekazu Shibuya A Basic Atlas of Radio-Wave Propagation , 1987 .

[5]  Clemens Elster,et al.  Verification of humans using the electrocardiogram , 2007, Pattern Recognit. Lett..

[6]  C.M. Furse,et al.  Miniaturized biocompatible microstrip antenna using genetic algorithm , 2005, IEEE Transactions on Antennas and Propagation.

[7]  C.M. Furse,et al.  Design of implantable microstrip antenna for communication with medical implants , 2004, IEEE Transactions on Microwave Theory and Techniques.

[8]  Philip R. Troyk,et al.  Implantable myoelectric sensors (IMES) for upper-extremity prosthesis control- preliminary work , 2003, Proceedings of the 25th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (IEEE Cat. No.03CH37439).

[9]  A. V. Vorst,et al.  Applications of RF/microwaves in medicine , 2002 .

[10]  Carmen C. Y. Poon,et al.  A novel biometrics method to secure wireless body area sensor networks for telemedicine and m-health , 2006, IEEE Communications Magazine.

[11]  Jun-ichi Takada Extension of Friis' Transmission Formula for Ultra-Wideband Systems , 2003 .

[12]  Patrick E. McSharry,et al.  Advanced Methods And Tools for ECG Data Analysis , 2006 .

[13]  William G. Scanlon,et al.  Radiowave propagation from a tissue-implanted source at 418 MHz and 916.5 MHz , 2000, IEEE Transactions on Biomedical Engineering.

[14]  Richard D. Beach,et al.  Towards a miniature implantable in vivo telemetry monitoring system dynamically configurable as a potentiostat or galvanostat for two- and three-electrode biosensors , 2005, IEEE Transactions on Instrumentation and Measurement.

[15]  C. Balanis Antenna theory , 1982 .