Improving Intra-Fractional Target Position Accuracy Using a 3D Surface Surrogate for Left Breast Irradiation Using the Respiratory-Gated Deep-Inspiration Breath-Hold Technique

Purpose To evaluate the use of 3D optical surface imaging as a surrogate for respiratory gated deep-inspiration breath-hold (DIBH) for left breast irradiation. Material and Methods Patients with left-sided breast cancer treated with lumpectomy or mastectomy were selected as candidates for DIBH treatment for their external beam radiation therapy. Treatment plans were created on both free breathing (FB) and DIBH computed tomography (CT) simulation scans to determine dosimetric benefits from DIBH. The Real-time Position Management (RPM) system was used to acquire patient's breathing trace during DIBH CT acquisition and treatment delivery. The reference 3D surface models from FB and DIBH CT scans were generated and transferred to the “AlignRT” system for patient positioning and real-time treatment monitoring. MV Cine images were acquired during treatment for each beam as quality assurance for intra-fractional position verification. The chest wall excursions measured on these images were used to define the actual target position during treatment, and to investigate the accuracy and reproducibility of RPM and AlignRT. Results Reduction in heart dose can be achieved using DIBH for left breast/chest wall radiation. RPM was shown to have inferior correlation with the actual target position, as determined by the MV Cine imaging. Therefore, RPM alone may not be an adequate surrogate in defining the breath-hold level. Alternatively, the AlignRT surface imaging demonstrated a superior correlation with the actual target positioning during DIBH. Both the vertical and magnitude real-time deltas (RTDs) reported by AlignRT can be used as the gating parameter, with a recommended threshold of ±3 mm and 5 mm, respectively. Conclusion The RPM system alone may not be sufficient for the required level of accuracy in left-sided breast/CW DIBH treatments. The 3D surface imaging can be used to ensure patient setup and monitor inter- and intra- fractional motions. Furthermore, the target position accuracy during DIBH treatment can be improved by AlignRT as a superior surrogate, in addition to the RPM system.

[1]  L. Rutqvist,et al.  Mortality by laterality of the primary tumour among 55,000 breast cancer patients from the Swedish Cancer Registry. , 1990, British Journal of Cancer.

[2]  W. Manning,et al.  Reduction of cardiac volume in left-breast treatment fields by respiratory maneuvers: a CT study. , 2000, International journal of radiation oncology, biology, physics.

[3]  M. Arnfield,et al.  Internal mammary node coverage: an investigation of presently accepted techniques. , 2000, International journal of radiation oncology, biology, physics.

[4]  Y. Ung,et al.  Deep inspiration breath hold to reduce irradiated heart volume in breast cancer patients , 2000, Proceedings of the 22nd Annual International Conference of the IEEE Engineering in Medicine and Biology Society (Cat. No.00CH37143).

[5]  B. E. F. Isher,et al.  Twenty-year follow-up of a randomized trial comparing total mastectomy, lumpectomy, and lumpectomy plus irradiation for the treatment of invasive breast cancer. , 2002 .

[6]  Michael B Sharpe,et al.  Significant reductions in heart and lung doses using deep inspiration breath hold with active breathing control and intensity-modulated radiation therapy for patients treated with locoregional breast irradiation. , 2003, International journal of radiation oncology, biology, physics.

[7]  Michael B Sharpe,et al.  Initial clinical experience with moderate deep-inspiration breath hold using an active breathing control device in the treatment of patients with left-sided breast cancer using external beam radiation therapy. , 2003, International journal of radiation oncology, biology, physics.

[8]  E. Yorke,et al.  Deep inspiration breath hold and respiratory gating strategies for reducing organ motion in radiation treatment. , 2004, Seminars in radiation oncology.

[9]  John M Buatti,et al.  Optically guided patient positioning techniques. , 2005, Seminars in radiation oncology.

[10]  David Djajaputra,et al.  Real-time 3D surface-image-guided beam setup in radiotherapy of breast cancer. , 2004, Medical physics.

[11]  Lena Specht,et al.  Breathing adapted radiotherapy for breast cancer: comparison of free breathing gating with the breath-hold technique. , 2005, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[12]  R. Collins,et al.  Effects of radiotherapy and of differences in the extent of surgery for early breast cancer on local recurrence and 15-year survival: an overview of the randomised trials , 2005, The Lancet.

[13]  Patrick A Kupelian,et al.  Optical tracking technology in stereotactic radiation therapy. , 2007, Medical dosimetry : official journal of the American Association of Medical Dosimetrists.

[14]  W. Schlegel,et al.  Accuracy of a commercial optical 3D surface imaging system for realignment of patients for radiotherapy of the thorax , 2007, Physics in medicine and biology.

[15]  Xiang Li,et al.  The intrafraction motion induced dosimetric impacts in breast 3D radiation treatment: a 4DCT based study. , 2007, Medical physics.

[16]  V. Ferrari,et al.  Cardiac morbidity and mortality after breast conservation treatment in patients with early-stage breast cancer and preexisting cardiac disease. , 2008, Clinical breast cancer.

[17]  B. Zurl,et al.  Wide Tangential Fields Including the Internal Mammary Lymph Nodes in Patients with Left-Sided Breast Cancer , 2009, Strahlentherapie und Onkologie.

[18]  Steve B. Jiang,et al.  Using surface imaging and visual coaching to improve the reproducibility and stability of deep-inspiration breath hold for left-breast-cancer radiotherapy , 2009, Physics in medicine and biology.

[19]  Masahiko Oguchi,et al.  Cardiac-sparing radiotherapy for the left breast cancer with deep breath-holding , 2009, Japanese Journal of Radiology.

[20]  Darren Kahler,et al.  Characterization of a real-time surface image-guided stereotactic positioning system. , 2010, Medical physics.

[21]  X. Li,et al.  Respiratory organ motion and dosimetric impact on breast and nodal irradiation. , 2010, International journal of radiation oncology, biology, physics.

[22]  Hiroki Shirato,et al.  Dynamic gating window for compensation of baseline shift in respiratory-gated radiation therapy. , 2011, Medical physics.

[23]  B. Chauvet,et al.  [Radiation-related heart toxicity]. , 2011, Cancer radiotherapie : journal de la Societe francaise de radiotherapie oncologique.

[24]  B. Chauvet,et al.  [Radiation-related heart toxicity]. , 2011, Cancer radiotherapie : journal de la Societe francaise de radiotherapie oncologique.

[25]  R Peto,et al.  Effect of radiotherapy after breast-conserving surgery on 10-year recurrence and 15-year breast cancer death: meta-analysis of individual patient data for 10 801 women in 17 randomised trials , 2011, The Lancet.

[26]  P. Hall,et al.  Incidence of heart disease in 35,000 women treated with radiotherapy for breast cancer in Denmark and Sweden. , 2011, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[27]  R. Arriagada,et al.  Long-term cardiovascular mortality after radiotherapy for breast cancer. , 2011, Journal of the American College of Cardiology.

[28]  I. Mjaaland,et al.  Cardiac and pulmonary dose reduction for tangentially irradiated breast cancer, utilizing deep inspiration breath-hold with audio-visual guidance, without compromising target coverage , 2011, Acta oncologica.

[29]  G. Sharp,et al.  A voluntary breath-hold treatment technique for the left breast with unfavorable cardiac anatomy using surface imaging. , 2012, International journal of radiation oncology, biology, physics.

[30]  I. Mjaaland,et al.  Radiation during deep inspiration allows loco-regional treatment of left breast and axillary-, supraclavicular- and internal mammary lymph nodes without compromising target coverage or dose restrictions to organs at risk , 2012, Acta oncologica.

[31]  P. Hall,et al.  Risk of ischemic heart disease in women after radiotherapy for breast cancer. , 2013, The New England journal of medicine.

[32]  K. Sasai,et al.  The dosimetric impact of respiratory breast movement and daily setup error on tangential whole breast irradiation using conventional wedge, field-in-field and irregular surface compensator techniques , 2012, Journal of radiation research.

[33]  Jan-Jakob Sonke,et al.  Assessment of set-up variability during deep inspiration breath hold radiotherapy for breast cancer patients by 3D-surface imaging. , 2013, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[34]  Jan-Jakob Sonke,et al.  Accuracy evaluation of a 3-dimensional surface imaging system for guidance in deep-inspiration breath-hold radiation therapy. , 2013, International journal of radiation oncology, biology, physics.

[35]  Sanford L Meeks,et al.  Clinical evaluation of interfractional variations for whole breast radiotherapy using 3-dimensional surface imaging. , 2013, Practical radiation oncology.

[36]  Gregg Tracton,et al.  Clinical experience with 3-dimensional surface matching-based deep inspiration breath hold for left-sided breast cancer radiation therapy. , 2014, Practical radiation oncology.