The Effect on Hemolysis of the Raceway Profile of Roller Pumps Used in Cardiopulmonary Bypass

This study aims at the influence on hemolysis of the differences between the maximum and minimum amplitudes of pressure in the outlet of three roller pump models adjusted by dynamic calibration method. Tests were performed with silicone tubes (½ × 3/16 inches) in fluid analogous to blood and fresh bovine blood from slaughterhouse. Tests with analogous solution to blood were performed varying the dynamic calibration pressure between 78 and 500 mm Hg. Tests with fresh bovine blood were performed with the three pumps simultaneously, and pressure differences and free hemoglobin in the plasma were measured during 360 minutes. Tests with both analogous solution to blood and fresh bovine blood showed differences of mean pressures of pump 2 related to pumps 1 and 3 (p < 0.01). The different models of roller pumps analyzed presented differences in pressure amplitudes (p < 0.01) and hemolysis (p < 0.01) adjusted for the same dynamic calibration pressure. Raceway profile of pump 2 resulted in smaller pressure amplitude, implying lower hemolysis rate compared with pumps 1 and 3.

[1]  O. Petrucci,et al.  The Influence of the Residual Stress in Silicone Tubes in the Calibration Methods of Roller Pumps Used in Cardiopulmonary Bypass , 2010, ASAIO journal.

[2]  O. Petrucci,et al.  Analysis of the hydrodynamic profile in different roller pumps models used in cardiopulmonary bypass. , 2009, Revista brasileira de cirurgia cardiovascular : orgao oficial da Sociedade Brasileira de Cirurgia Cardiovascular.

[3]  Hyuk Choi,et al.  Comparison of hemolytic properties of different shapes of occlusion of blood sac in occlusive-type pulsatile blood pump. , 2008, Artificial organs.

[4]  J. Dunning,et al.  Does the use of a centrifugal pump offer any additional benefit for patients having open heart surgery? , 2006, Interactive cardiovascular and thoracic surgery.

[5]  Tohru Takaseya,et al.  Non-occlusive condition with the Better-Header roller pump: impacts of flow dynamics and hemolysis. , 2004, Annals of thoracic and cardiovascular surgery : official journal of the Association of Thoracic and Cardiovascular Surgeons of Asia.

[6]  F. Merkle,et al.  History of extracorporeal circulation: the invention and modification of blood pumps. , 2003, The journal of extra-corporeal technology.

[7]  L. V. von Segesser,et al.  Smart suction device for less blood trauma: a comparison with Cell Saver. , 2001, European Journal of Cardio-Thoracic Surgery.

[8]  D. Sharpe,et al.  Haemolysis during cardiopulmonary bypass: an in vivo comparison of standard roller pumps, nonocclusive roller pumps and centrifugal pumps , 1999, Perfusion.

[9]  Y Sankai,et al.  Species differences in erythrocyte mechanical fragility: comparison of human, bovine, and ovine cells. , 1998, ASAIO journal.

[10]  M. Oz,et al.  Clinical evaluation of setting pump occlusion by the dynamic method: effect on flow , 1998, Perfusion.

[11]  J. B. Riley,et al.  An under-occluded roller pump is less hemolytic than a centrifugal pump. , 1997, The journal of extra-corporeal technology.

[12]  E. Leonard,et al.  A Dynamic Method for Setting Roller Pumps Nonocclusively Reduces Hemolysis and Predicts Retrograde Flow , 1997, ASAIO journal.

[13]  T. Demmy,et al.  In vitro evaluation of five commercially available perfusion systems , 1994 .

[14]  G. Wright,et al.  Cellular aggregation and trauma in cardiotomy suction systems. , 1979, Thorax.

[15]  L. Edmunds,et al.  Relationship between platelet count and cardiotomy suction return. , 1978, The Annals of thoracic surgery.