The osteoarthritis initiative (OAI) magnetic resonance imaging quality assurance methods and results.

OBJECTIVE To outline the osteoarthritis initiative's (OAI) magnetic resonance (MR) system quality assurance (QA) processes and present the first 3 years' results. METHOD OAI MR QA included acquisitions evaluated manually at each of the four sites and other acquisitions assessed by an automated computer program. Key image characteristics such as signal-to-noise, contrast-to-noise, signal uniformity, T2 relaxation times, local and global geometric distortion were quantified monthly using the automated program. RESULTS Uniformly high quality, artifact-free subject images were obtained from all four OAI 3 tesla MR facilities. Over a 3-year period, key criteria for quantitative cartilage morphometry were excellent with a 190.0 mm diameter and 148.0 mm length object having reproducible diameter (0.04% RMS CV) and length (0.56% RMS CV). This resulted in spherical volume reproducibility of 0.46% RMS CV. Ghost levels were consistently <0.2%. T2 relaxation time varied longitudinally site-by-site from 2.3 to 18.8% RMS CV. All other measures of MR system stability were met except: 3.0 mm and 5.0 mm slice thicknesses were consistently larger than expected; knee coil signal uniformity and signal level varied significantly over time. CONCLUSIONS OAI MR QA results compared favorably to prior publications and identified similar technical issues for geometric measurements. The longitudinal variations measured in the OAI QA process should have minimal impact on the accuracy and reproducibility of cartilage thickness and volume quantification. This stability should enable direct comparison of baseline and follow-up images. Cross-comparison of the results from all four OAI sites reveals that the MR systems are sufficiently uniform to enable results to be combined.

[1]  F Eckstein,et al.  Double echo steady state magnetic resonance imaging of knee articular cartilage at 3 Tesla: a pilot study for the Osteoarthritis Initiative , 2005, Annals of the rheumatic diseases.

[2]  O Henriksen,et al.  Quality assessment in in vivo NMR spectroscopy: IV. A multicentre trial of test objects and protocols for performance assessment in clinical NMR spectroscopy. , 1995, Magnetic resonance imaging.

[3]  L. Axel,et al.  Quality assurance methods and phantoms for magnetic resonance imaging: report of AAPM nuclear magnetic resonance Task Group No. 1. , 1990, Medical physics.

[4]  F Eckstein,et al.  Two year longitudinal change and test-retest-precision of knee cartilage morphology in a pilot study for the osteoarthritis initiative. , 2007, Osteoarthritis and cartilage.

[5]  S K Mun,et al.  Acceptance testing of magnetic resonance imaging systems: report of AAPM Nuclear Magnetic Resonance Task Group No. 6. , 1992, Medical physics.

[6]  Magnetic Resonance Imaging of the Knee , 1988 .

[7]  David M. Doddrell,et al.  Geometric Distortion in Structural Magnetic Resonance Imaging , 2005 .

[8]  P. M. Walker,et al.  V. Multi-center trial with protocols and prototype test objects for the assessment of MRI equipment , 1988 .

[9]  Deming Wang,et al.  Geometric distortion in clinical MRI systems Part II: correction using a 3D phantom. , 2004, Magnetic resonance imaging.

[10]  P L Carson,et al.  MRI scanner variability studies using a semi-automated analysis system. , 1994, Magnetic resonance imaging.

[11]  Richard I. Kitney,et al.  Standardization of performance evaluation in MRI: 13 Years' experience of intersystem comparison , 2002 .

[12]  J. D. de Certaines,et al.  Performance assessment and quality control in MRI by Eurospin test objects and protocols. , 1993, Magnetic resonance imaging.

[13]  D Matthaei,et al.  1H NMR chemical shift selective (CHESS) imaging. , 1985, Physics in medicine and biology.

[14]  S Napel,et al.  Quantifying MRI geometric distortion in tissue , 1994, Magnetic resonance in medicine.

[15]  David M. Doddrell,et al.  Geometric distortion in clinical MRI systems Part II: correction using a 3D phantom. , 2004, Magnetic resonance imaging.

[16]  Deming Wang,et al.  Geometric distortion in clinical MRI systems Part I: evaluation using a 3D phantom. , 2004, Magnetic resonance imaging.