Diaphragm Activity Pre and Post Extubation in Ventilated Critically Ill Infants and Children Measured With Transcutaneous Electromyography*

Supplemental Digital Content is available in the text. OBJECTIVES: Swift extubation is important to prevent detrimental effects of invasive mechanical ventilation but carries the risk of extubation failure. Accurate tools to assess extubation readiness are lacking. This study aimed to describe the effect of extubation on diaphragm activity in ventilated infants and children. Our secondary aim was to compare diaphragm activity between failed and successfully extubated patients. DESIGN: Prospective, observational study. SETTING: Single-center tertiary neonatal ICU and PICU. PATIENTS: Infants and children receiving invasive mechanical ventilation longer than 24 hours. INTERVENTIONS: None. MEASUREMENTS AND MAIN RESULTS: Diaphragm activity was measured with transcutaneous electromyography, from 15 minutes before extubation till 180 minutes thereafter. Peak and tonic activity, inspiratory amplitude, inspiratory area under the curve, and respiratory rate were calculated from the diaphragm activity waveform. One hundred forty-seven infants and children were included (median postnatal age, 1.9; interquartile range, 0.9–6.7 wk). Twenty patients (13.6%) failed extubation within 72 hours. Diaphragm activity increased rapidly after extubation and remained higher throughout the measurement period. Pre extubation, peak (end-inspiratory) diaphragm activity and tonic (end-inspiratory) diaphragm activity were significantly higher in failure, compared with success cases (5.6 vs 7.0 μV; p = 0.04 and 2.8 vs 4.1 μV; p = 0.04, respectively). Receiver operator curve analysis showed the highest area under the curve for tonic (end-inspiratory) diaphragm activity (0.65), with a tonic (end-inspiratory) diaphragm activity greater than 3.4 μV having a combined sensitivity and specificity of 55% and 77%, respectively, to predict extubation outcome. After extubation, diaphragm activity remained higher in patients failing extubation. CONCLUSIONS: Diaphragm activity rapidly increased after extubation. Patients failing extubation had a higher level of diaphragm activity, both pre and post extubation. The predictive value of the diaphragm activity variables alone was limited. Future studies are warranted to assess the additional value of electromyography of the diaphragm in combined extubation readiness assessment.

[1]  W. Shalish,et al.  Diaphragmatic activity and neural breathing variability during a 5-min endotracheal continuous positive airway pressure trial in extremely preterm infants , 2020, Pediatric Research.

[2]  A. Greenough,et al.  Prediction of extubation success using the diaphragmatic electromyograph results in ventilated neonates , 2020, Journal of perinatal medicine.

[3]  Yang Xue,et al.  A prospective observational study on critically ill children with diaphragmatic dysfunction: clinical outcomes and risk factors , 2020, BMC Pediatrics.

[4]  A. V. van Kaam,et al.  Diaphragmatic electromyography in preterm infants: The influence of electrode positioning , 2019, Pediatric pulmonology.

[5]  G J Hutten,et al.  Processing transcutaneous electromyography measurements of respiratory muscles, a review of analysis techniques. , 2019, Journal of electromyography and kinesiology : official journal of the International Society of Electrophysiological Kinesiology.

[6]  M. Aalbers,et al.  Amsterdam, Netherlands , 2018, The Statesman’s Yearbook Companion.

[7]  M. Kneyber Ventilator-induced lung injury: does it occur in children? , 2018, Minerva Anestesiologica.

[8]  R. Berg,et al.  Progressive Diaphragm Atrophy in Pediatric Acute Respiratory Failure* , 2018, Pediatric critical care medicine : a journal of the Society of Critical Care Medicine and the World Federation of Pediatric Intensive and Critical Care Societies.

[9]  Wissam Shalish,et al.  Predictors of extubation readiness in preterm infants: a systematic review and meta-analysis , 2018, Archives of Disease in Childhood: Fetal and Neonatal Edition.

[10]  R. Darnall,et al.  Does Diaphragmatic Electrical Activity in Preterm Infants Predict Extubation Success? , 2017, Respiratory Care.

[11]  K. Sward,et al.  Mechanical Ventilation and Decision Support in Pediatric Intensive Care. , 2017, Pediatric clinics of North America.

[12]  L. Brochard,et al.  Critical illness-associated diaphragm weakness , 2017, Intensive Care Medicine.

[13]  Narayanaswamy P. R. Iyer,et al.  Neural breathing pattern in newborn infants pre‐ and postextubation , 2017, Acta paediatrica.

[14]  G. Rafferty,et al.  Risk Factors for Pediatric Extubation Failure: The Importance of Respiratory Muscle Strength* , 2017, Critical care medicine.

[15]  F. D. de Jongh,et al.  Electrical activity of the diaphragm during nCPAP and high flow nasal cannula , 2017, Archives of Disease in Childhood: Fetal and Neonatal Edition.

[16]  A. V. van Kaam,et al.  Diaphragmatic activity during weaning from respiratory support in preterm infants , 2016, Archives of Disease in Childhood: Fetal and Neonatal Edition.

[17]  G. Rafferty,et al.  Pediatric extubation readiness tests should not use pressure support , 2016, Intensive Care Medicine.

[18]  P. Breheny,et al.  Do Clinical Parameters Predict First Planned Extubation Outcome in the Pediatric Intensive Care Unit? , 2015, Journal of intensive care medicine.

[19]  L. Ducharme-Crevier,et al.  Evolution of inspiratory diaphragm activity in children over the course of the PICU stay , 2014, Intensive Care Medicine.

[20]  P. Davis,et al.  Definitions of extubation success in very premature infants: a systematic review , 2013, Archives of Disease in Childhood: Fetal and Neonatal Edition.

[21]  O. Moerer,et al.  Electrical activity of the diaphragm (EAdi) as a monitoring parameter in difficult weaning from respirator: a pilot study , 2013, Critical Care.

[22]  G. Badger,et al.  Obstetric and Neonatal Care Practices for Infants 501 to 1500 g From 2000 to 2009 , 2013, Pediatrics.

[23]  T. Similowski,et al.  Diaphragm electromyographic activity as a predictor of weaning failure , 2012, Intensive Care Medicine.

[24]  A. Greenough,et al.  Prediction of extubation outcome in infants using the tension time index , 2010, Archives of Disease in Childhood: Fetal and Neonatal Edition.

[25]  B. Markovitz,et al.  Characteristics of children intubated and mechanically ventilated in 16 PICUs. , 2009, Chest.

[26]  Carol E. Nicholson,et al.  Weaning and extubation readiness in pediatric patients* , 2009, Pediatric critical care medicine : a journal of the Society of Critical Care Medicine and the World Federation of Pediatric Intensive and Critical Care Societies.

[27]  A. Zaritsky,et al.  Spontaneous breathing trial predicts successful extubation in infants and children , 2006, Pediatric critical care medicine : a journal of the Society of Critical Care Medicine and the World Federation of Pediatric Intensive and Critical Care Societies.

[28]  Guillaume Emeriaud,et al.  Diaphragm Electrical Activity During Expiration in Mechanically Ventilated Infants , 2006, Pediatric Research.

[29]  C. O. Kamlin,et al.  Predicting successful extubation of very low birthweight infants , 2005, Archives of Disease in Childhood - Fetal and Neonatal Edition.

[30]  S. D. Baisch,et al.  Extubation failure in pediatric intensive care incidence and outcomes , 2005, Pediatric critical care medicine : a journal of the Society of Critical Care Medicine and the World Federation of Pediatric Intensive and Critical Care Societies.

[31]  F. Frutos,et al.  What is the daily practice of mechanical ventilation in pediatric intensive care units? A multicenter study , 2004, Intensive Care Medicine.

[32]  S. Epstein,et al.  Extubation failure: magnitude of the problem, impact on outcomes, and prevention , 2003, Current opinion in critical care.

[33]  A. Greenough,et al.  Assessment of respiratory drive and muscle function in the pediatric intensive care unit and prediction of extubation failure , 2000, Pediatric critical care medicine : a journal of the Society of Critical Care Medicine and the World Federation of Pediatric Intensive and Critical Care Societies.

[34]  S. Venkataraman,et al.  Predictors of extubation success and failure in mechanically ventilated infants and children. , 1996, Critical care medicine.

[35]  A. C. Bryan,et al.  Importance of inspiratory muscle tone in maintenance of FRC in the newborn. , 1981, Journal of applied physiology: respiratory, environmental and exercise physiology.

[36]  G. Elie,et al.  INTERMITTENT HYPOXEMIA IN PRETERM INFANTS , 2018 .

[37]  C. Gaultier Respiratory muscle function in infants. , 1995, The European respiratory journal.