Comments on MR heating tests of critical implants

MAGNETIC RESONANCE IMAGING (MRI)-related heating for certain implants may be excessive, causing substantial patient injury. The determination of implant heating is particularly challenging because of the numerous variables that must be considered in order to properly identify both MRIand implant-related conditions that can impact the findings (1,2). Therefore, I applaud the recent commentary by Kainz (3), a member of the Food and Drug Administration (FDA), for presenting several of the confounding factors that complicate tests used to characterize implant heating. Kainz’s (3) guest editorial refers to possible issues associated with the in vitro assessment of MRI-related heating, with an emphasis on “MR critical implants.” MR critical implants were defined as: “active implantable medical devices (AIMDs); semiactive implants, i.e., implants powered from outside the body; and elongated metallic structures that are in the range of the critical length (defined later in the guest editorial) (3). Notably, there was no mention of devices that have ”resonant conducting loops,“ which may likewise generate excessive heating in association with MRI (4). Examples include cervical fixation devices and external fixation systems. As a member of the MRI community involved in the evaluation of implants since 1988 (5,6), I welcome the opportunity to respond to the content of this guest editorial insofar as our group and other investigators have extensive experience testing many of the so-called MR critical implants (7–23). In addition, we have been involved in the development of the methodology utilized to evaluate these devices (24). Importantly, findings from many studies have resulted in labeling approved by the U.S. Food and Drug Administration (FDA) to permit MRI procedures to be performed in patients with active, semiactive, and elongated devices, including cochlear implants, bone fusion stimulators, Foley catheters with temperature sensors, implantable infusion pumps, a deep brain stimulation system, spinal cord stimulators, a vagus nerve stimulator, a microstimulator (Bion), relatively long, orthopedic implants, and many others (7–23). As a result of this work, patients with “MR critical implants” throughout the world undergo MRI procedures safely, as MR healthcare workers follow FDA-approved guidelines (or guidelines approved by other agencies), which usually state specific instructions for a given device with respect to MRI parameters and other instructions. These instructions often indicate a particular whole-body averaged specific absorption rate (WB-SAR) that should not be exceeded during the MRI examination. For the aforementioned devices, the evaluation of MRI safety included an assessment of heating caused by MRI procedures. Interestingly, as the guest editorial indicates, “Most MR heating tests refer to [American Society for Testing and Materials International] ASTM Standard 2182-02a which is for passive implants only” (3). One of the reasons for this is that, currently, there is no standard for testing MR critical implants and, therefore, the ASTM standard used for passive implants has been modified to test active, semiactive, and elongated implants. In spite of this issue, the fact remains that the FDA approved MRI-related labeling for implants that utilized findings from heating tests suggested to be problematic. Notably, to date, there is no evidence that patient injuries have occurred in relation to the use of labeling approved by the FDA for these devices relative to the heating information. By comparison, there have been serious patient injuries related to not following specific instructions intended to prevent excessive implant heating (25–27).

[1]  Frank G Shellock,et al.  Simple design changes to wires to substantially reduce MRI-induced heating at 1.5 T: implications for implanted leads. , 2005, Magnetic resonance imaging.

[2]  M. Dewhirst,et al.  Summary, conclusions and recommendations: adverse temperature levels in the human body , 2003, International journal of hyperthermia : the official journal of European Society for Hyperthermic Oncology, North American Hyperthermia Group.

[3]  J. McDougall,et al.  Absence of radiofrequency heating from auditory implants during magnetic resonance imaging. , 1995, Bioelectromagnetics.

[4]  F. Shellock MR imaging of metallic implants and materials: a compilation of the literature. , 1988, AJR. American journal of roentgenology.

[5]  F. Shellock,et al.  MR procedures: biologic effects, safety, and patient care. , 2004, Radiology.

[6]  F G Shellock,et al.  High-field-strength MR imaging and metallic biomedical implants: an ex vivo evaluation of deflection forces. , 1988, AJR. American journal of roentgenology.

[7]  Liong Liem,et al.  Magnetic resonance imaging and spinal cord stimulation systems , 1997, Pain.

[8]  Ashwini Sharan,et al.  Neurostimulation systems for deep brain stimulation: In vitro evaluation of magnetic resonance imaging–related heating at 1.5 tesla , 2002, Journal of magnetic resonance imaging : JMRI.

[9]  Scalp burns from halo pins following magnetic resonance imaging. Case illustration. , 2003, Journal of neurosurgery.

[10]  J. Nyenhuis,et al.  Evaluation of MRI compatibility of the modified nucleus multichannel auditory brainstem and cochlear implants. , 1996, The American journal of otology.

[11]  M. Khoo,et al.  Estimation of chemoreflex loop gain using pseudorandom binary CO/sub 2/ stimulation , 1997, IEEE Transactions on Biomedical Engineering.

[12]  V. Tronnier,et al.  Active deep brain stimulation during MRI: A feasibility study , 2004, Magnetic resonance in medicine.

[13]  Wolfgang Kainz MR heating tests of MR critical implants , 2007, Journal of magnetic resonance imaging : JMRI.

[14]  Frank G. Shellock,et al.  Reference Manual for Magnetic Resonance Safety, Implants, and Devices , 2009 .

[15]  Daniel S Berman,et al.  Cardiac pacemaker: in vitro assessment at 1.5 T. , 2006, American heart journal.

[16]  C H Chouard,et al.  [Nuclear magnetic resonance and cochlear implant]. , 1997, Annales d'oto-laryngologie et de chirurgie cervico faciale : bulletin de la Societe d'oto-laryngologie des hopitaux de Paris.

[17]  Jean A. Tkach,et al.  Evaluation of specific absorption rate as a dosimeter of MRI‐related implant heating , 2004, Journal of magnetic resonance imaging : JMRI.

[18]  F. Shellock,et al.  Implantable Microstimulator: Magnetic Resonance Safety at 1.5 Tesla , 2004, Investigative radiology.

[19]  Jorge Gonzalez-Martinez,et al.  MR imaging-related heating of deep brain stimulation electrodes: in vitro study. , 2002, AJNR. American journal of neuroradiology.

[20]  E S Hochmair,et al.  Cochlear implants: in vitro investigation of electromagnetic interference at MR imaging--compatibility and safety aspects. , 1998, Radiology.

[21]  F. Shellock,et al.  MR safety in patients with implanted deep brain stimulation systems (DBS). , 2003, Acta neurochirurgica. Supplement.

[22]  J. McDougall,et al.  RF heating of implanted spinal fusion stimulator during magnetic resonance imaging , 1996, IEEE Transactions on Biomedical Engineering.

[23]  Shantanu Sinha,et al.  Bilateral neurostimulation systems used for deep brain stimulation: in vitro study of MRI-related heating at 1.5 T and implications for clinical imaging of the brain. , 2005, Magnetic resonance imaging.

[24]  Jean A. Tkach,et al.  Permanent Neurological Deficit Related to Magnetic Resonance Imaging in a Patient with Implanted Deep Brain Stimulation Electrodes for Parkinson’s Disease: Case Report , 2005, Neurosurgery.

[25]  Jean A. Tkach,et al.  Is magnetic resonance imaging safe for patients with neurostimulation systems used for deep brain stimulation? , 2005, Neurosurgery.

[26]  E S Hochmair,et al.  Magnetic resonance imaging and cochlear implants: Compatibility and safety aspects , 1999, Journal of magnetic resonance imaging : JMRI.

[27]  F G Shellock,et al.  MR imaging and electronically activated devices. , 2001, Radiology.

[28]  Jean A. Tkach,et al.  Variability in RF‐induced heating of a deep brain stimulation implant across MR systems , 2006, Journal of magnetic resonance imaging : JMRI.

[29]  Robert J. Witte,et al.  Magnetic Resonance Imaging and Deep Brain Stimulation , 2002, Neurosurgery.

[30]  F. Shellock,et al.  Cardiac pacemakers and implantable cardioverter defibrillators: in vitro magnetic resonance imaging evaluation at 1.5-tesla. , 2007, Journal of cardiovascular magnetic resonance : official journal of the Society for Cardiovascular Magnetic Resonance.

[31]  V. Tronnier,et al.  Magnetic resonance imaging with implanted neurostimulators: an in vitro and in vivo study. , 1999, Neurosurgery.

[32]  B. Condon,et al.  Investigation of the factors responsible for burns during MRI , 2001, Journal of magnetic resonance imaging : JMRI.

[33]  Jean A. Tkach,et al.  Reduction of Magnetic Resonance Imaging-related Heating in Deep Brain Stimulation Leads Using a Lead Management Device , 2005, Neurosurgery.

[34]  S. Esterby American Society for Testing and Materials , 2006 .

[35]  H. Hricak,et al.  The Effect of Magnetic Resonance Imagers on Implanted Neurostimulators , 1992, Pacing and clinical electrophysiology : PACE.

[36]  Peter Boesiger,et al.  In vivo heating of pacemaker leads during magnetic resonance imaging. , 2005, European heart journal.