Difference in the detection limits of flaws in the depths of multi-layered and continuous aluminum plates using low-frequency eddy current testing

We examined the difference in the detection limits of flaws in the depths of multi-layered and continuous aluminum plates using low-frequency eddy current testing. The detection limits were measured by using a magneto-resistive sensor. Comparing the frequency of an applied magnetic field, the detection limit at 50 Hz is deeper than that at 1 kHz. Comparing the sample structure, the detection limit in the multi-layered samples is deeper than that in the continuous samples. These results are likely due to the differences in the skin depth and conductivity of the sample.