Evaluation of intraocular lens position during phacoemulsification using intraoperative spectral‐domain optical coherence tomography

Purpose To assess the position of intraocular lenses (IOLs) at the end of standard phacoemulsification with intraoperative spectral‐domain optical coherence tomography (SD‐OCT). Settings Department of Ophthalmology, Rudolf Foundation Hospital, Vienna, Austria. Design Prospective case series. Methods Standard phacoemulsification with IOL implantation was performed. The Rescan 700 SD‐OCT system was used for intraoperative imaging. The anterior segment of the eye was scanned using SD‐OCT at the end of the surgery. The distance from the IOL optic center and the IOL optic edge to the posterior capsule was measured postoperatively using graphic software. Results The study comprised 74 patients (101 eyes). The mean axial length was 23.97 mm (range 21.43 to 28.61 mm). The mean IOL power was 20.39 diopters (D) (range 6.5 to 27.5 D). Contact between the IOL and posterior capsule was absent in 88 cases (87.13%), and partial or full contact was present in 13 cases (12.87%). The mean distance between the IOL central optic and posterior capsule was 0.71 pixel (range 0.06 to 1.38 pixels) in 99 cases (98.02%). In 42 cases (57.53%), partial contact between the IOL edges and the posterior capsule was noticed. The mean distance between the IOL edge and posterior capsule was 0.21 pixel (range 0.04 to 0.92 pixel). Conclusions Intraoperative SD‐OCT facilitated the imaging of IOL position during standard phacoemulsification. Contact between the IOL central optic and posterior capsule at the end of the surgery occurred rarely. Improved IOL design should be considered. Financial Disclosure Drs. Binder and Glittenberg are consultants to Carl Zeiss Meditech AG. None of the other authors has a financial or proprietary interest in any material or method mentioned.

[1]  J. Danias,et al.  Haptic-induced postoperative complications. Evaluation using ultrasound biomicroscopy. , 2013, Canadian journal of ophthalmology. Journal canadien d'ophtalmologie.

[2]  Riccardo Barberi,et al.  Analysis of intraocular lens surface adhesiveness by atomic force microscopy , 2009, Journal of cataract and refractive surgery.

[3]  Chen D. Lu,et al.  Retinal, anterior segment and full eye imaging using ultrahigh speed swept source OCT with vertical-cavity surface emitting lasers , 2012, Biomedical optics express.

[4]  O. Findl,et al.  The capsular tension ring: Designs, applications, and techniques , 2000, Journal of cataract and refractive surgery.

[5]  D. Apple,et al.  Posterior capsule opacification: a review of the aetiopathogenesis, experimental and clinical studies and factors for prevention. , 2004, Indian journal of ophthalmology.

[6]  D. Apple,et al.  Post cataract-intraocular lens (IOL) surgery opacification , 2002, Eye.

[7]  D. Apple,et al.  Effect of in‐the‐bag intraocular lens fixation on the prevention of posterior capsule opacification , 2001, Journal of cataract and refractive surgery.

[8]  Oliver Findl,et al.  Using continuous intraoperative optical coherence tomography measurements of the aphakic eye for intraocular lens power calculation , 2014, British Journal of Ophthalmology.

[9]  O. Nishi,et al.  Inhibition of migrating lens epithelial cells at the capsular bend created by the rectangular optic edge of a posterior chamber intraocular lens. , 1998, Ophthalmic surgery and lasers.

[10]  D. Apple,et al.  Surgical prevention of posterior capsule opacification. Part 3: Intraocular lens optic barrier effect as a second line of defense. , 2000, Journal of cataract and refractive surgery.

[11]  Thomas Kohnen,et al.  Optic edge design as long-term factor for posterior capsular opacification rates. , 2008, Ophthalmology.

[12]  Oliver Findl,et al.  Predicting the postoperative intraocular lens position using continuous intraoperative optical coherence tomography measurements. , 2013, Investigative ophthalmology & visual science.

[13]  M. Campbell,et al.  Biometric, optical and physical changes in the isolated human crystalline lens with age in relation to presbyopia , 1999, Vision Research.

[14]  B. Dick,et al.  Capsule measuring ring to predict capsular bag diameter and follow its course after foldable intraocular lens implantation , 2003, Journal of cataract and refractive surgery.

[15]  Susanne Binder,et al.  FEASIBILITY OF INTRASURGICAL SPECTRAL-DOMAIN OPTICAL COHERENCE TOMOGRAPHY , 2011, Retina.

[16]  R. Chuck,et al.  Vitreous Hyper-Reflective Dots in Optical Coherence Tomography and Cystoid Macular Edema after Uneventful Phacoemulsification Surgery , 2014, PloS one.

[17]  Eva Lankenau,et al.  Future Prospects: Assessment of Intraoperative Optical Coherence Tomography in Ab Interno Glaucoma Surgery , 2015, Current eye research.