Evaluation of front-end higher energy collision-induced dissociation on a benchtop dual-pressure linear ion trap mass spectrometer for shotgun proteomics.

We report the implementation of front-end higher energy collision-induced dissociation (fHCD) on a benchtop dual-pressure linear ion trap. Software and hardware modifications were employed, described in detail vide-infra, to allow isolated ions to undergo collisions with ambient gas molecules in an intermediate multipole (q00) of the instrument. Results comparing the performance of fHCD and resonance excitation collision-induced dissociation (RE-CID) in terms of injection time, total number of scans, efficiency, mass measurement accuracy (MMA), unique peptide identifications, and spectral quality of labile modified peptides are presented. fHCD is approximately 23% as efficient as RE-CID, and depending on the search algorithm, it identifies 6.6% more or 15% less peptides (q < 0.01) from a soluble whole-cell lysate ( Caenorhabditis elegans ) than RE-CID using Mascot or Sequest search algorithms, respectively. fHCD offers a clear advantage for the analysis of phosphorylated and glycosylated (O-GlcNAc) peptides as the average cross-correlation score (XCorr) for spectra using fHCD was statistically greater (p < 0.05) than for spectra collected using RE-CID.

[1]  Michael J MacCoss,et al.  Comparison of database search strategies for high precursor mass accuracy MS/MS data. , 2010, Journal of proteome research.

[2]  M. Mann,et al.  Higher-energy C-trap dissociation for peptide modification analysis , 2007, Nature Methods.

[3]  A. Makarov,et al.  The Orbitrap: a new mass spectrometer. , 2005, Journal of mass spectrometry : JMS.

[4]  Matthias Mann,et al.  A Dual Pressure Linear Ion Trap Orbitrap Instrument with Very High Sequencing Speed* , 2009, Molecular & Cellular Proteomics.

[5]  D. Muddiman,et al.  Mass measurement accuracy comparisons between a double-focusing magnetic sector and a time-of-flight mass analyzer. , 2008, Rapid communications in mass spectrometry : RCM.

[6]  J. Futrell,et al.  Tandem mass spectrometry: dissociation of ions by collisional activation , 2000, Journal of mass spectrometry : JMS.

[7]  Michael S. Bereman,et al.  Comparison between procedures using SDS for shotgun proteomic analyses of complex samples , 2011, Proteomics.

[8]  G. Hart,et al.  O-GlcNAc a sensor of cellular state: the role of nucleocytoplasmic glycosylation in modulating cellular function in response to nutrition and stress. , 2004, Biochimica et biophysica acta.

[9]  Markus Brosch,et al.  Accurate and sensitive peptide identification with Mascot Percolator. , 2009, Journal of proteome research.

[10]  Robert J Chalkley,et al.  Identification of protein O-GlcNAcylation sites using electron transfer dissociation mass spectrometry on native peptides , 2009, Proceedings of the National Academy of Sciences.

[11]  William Stafford Noble,et al.  Semi-supervised learning for peptide identification from shotgun proteomics datasets , 2007, Nature Methods.

[12]  K. Parker,et al.  Multiplexed Protein Quantitation in Saccharomyces cerevisiae Using Amine-reactive Isobaric Tagging Reagents*S , 2004, Molecular & Cellular Proteomics.

[13]  John K. Heath,et al.  Large Scale Localization of Protein Phosphorylation by Use of Electron Capture Dissociation Mass Spectrometry , 2009, Molecular & Cellular Proteomics.

[14]  R. Chalkley,et al.  Identification of GlcNAcylation sites of peptides and α-crystallin using Q-TOF mass spectrometry , 2001 .

[15]  B. Simons,et al.  Performance characteristics of a new hybrid quadrupole time-of-flight tandem mass spectrometer (TripleTOF 5600). , 2011, Analytical chemistry.

[16]  B W Gibson,et al.  Low-mass ions produced from peptides by high-energy collision-induced dissociation in tandem mass spectrometry , 1993, Journal of the American Society for Mass Spectrometry.

[17]  Scott A. McLuckey,et al.  SPECIAL FEATURE:TUTORIAL Slow Heating Methods in Tandem Mass Spectrometry , 1997 .

[18]  A. Makarov,et al.  Interfacing the orbitrap mass analyzer to an electrospray ion source. , 2003, Analytical chemistry.

[19]  Douglas H. Phanstiel,et al.  Higher-energy Collision-activated Dissociation Without a Dedicated Collision Cell* , 2011, Molecular & Cellular Proteomics.

[20]  F W McLafferty,et al.  Localization of labile posttranslational modifications by electron capture dissociation: the case of gamma-carboxyglutamic acid. , 1999, Analytical chemistry.

[21]  H. Steen,et al.  Using optimized collision energies and high resolution, high accuracy fragment ion selection to improve glycopeptide detection by precursor ion scanning , 2003, Journal of the American Society for Mass Spectrometry.

[22]  P. Cohen The role of protein phosphorylation in human health and disease. The Sir Hans Krebs Medal Lecture. , 2001, European journal of biochemistry.

[23]  Scott A. Busby,et al.  Novel linear quadrupole ion trap/FT mass spectrometer: performance characterization and use in the comparative analysis of histone H3 post-translational modifications. , 2004, Journal of proteome research.

[24]  M. Senko,et al.  A two-dimensional quadrupole ion trap mass spectrometer , 2002, Journal of the American Society for Mass Spectrometry.

[25]  Aaron A. Klammer,et al.  Effects of modified digestion schemes on the identification of proteins from complex mixtures. , 2006, Journal of proteome research.

[26]  J. Shabanowitz,et al.  Peptide and protein sequence analysis by electron transfer dissociation mass spectrometry. , 2004, Proceedings of the National Academy of Sciences of the United States of America.

[27]  Craig D Wenger,et al.  Phosphoproteomics for the masses. , 2010, ACS chemical biology.

[28]  Lance Wells,et al.  Combining high-energy C-trap dissociation and electron transfer dissociation for protein O-GlcNAc modification site assignment. , 2011, Journal of proteome research.

[29]  J. Hanover,et al.  The hexosamine signaling pathway: O-GlcNAc cycling in feast or famine. , 2010, Biochimica et biophysica acta.