Dosimetric effect of respiration-gated beam on IMRT delivery.

Intensity modulated radiation therapy (IMRT) with a dynamic multileaf collimator (DMLC) requires synchronization of DMLC leaf motion with dose delivery. A delay in DMLC communication is known to cause leaf lag and lead to dosimetric errors. The errors may be exacerbated by gated operation. The purpose of this study was to investigate the effect of leaf lag on the accuracy of doses delivered in gated IMRT. We first determined the effective leaf delay time by measuring the dose in a stationary phantom delivered by wedge-shaped fields. The wedge fields were generated by a DMLC at various dose rates. The so determined delay varied from 88.3 to 90.5 ms. The dosimetric effect of this delay on gated IMRT was studied by delivering wedge-shaped and clinical IMRT fields to moving and stationary phantoms at dose rates ranging from 100 to 600 MU/min, with and without gating. Respiratory motion was simulated by a linear sinusoidal motion of the phantom. An ionization chamber and films were employed for absolute dose and 2-D dose distribution measurements. Discrepancies between gated and nongated delivery to the stationary phantom were observed in both absolute dose and 2-D dose distribution measurements. These discrepancies increased monotonically with dose rate and frequency of beam interruptions, and could reach 3.7% of the total dose delivered to a 0.6 cm3 ion chamber. Isodose lines could be shifted by as much as 3 mm. The results are consistent with the explanation that beam hold-offs in gated delivery allowed the lagging leaves to catch up with the delivered monitor units each time that the beam was interrupted. Low dose rates, slow leaf speeds and low frequencies of beam interruptions reduce the effect of this delay-and-catch-up cycle. For gated IMRT it is therefore important to find a good balance between the conflicting requirements of rapid dose delivery and delivery accuracy.

[1]  Gary A. Ezzell,et al.  The overshoot phenomenon in step‐and‐shoot IMRT delivery , 2001, Journal of applied clinical medical physics.

[2]  T LoSasso,et al.  Testing of dynamic multileaf collimation. , 1996, Medical physics.

[3]  Ping Xia,et al.  Communication and sampling rate limitations in IMRT delivery with a dynamic multileaf collimator system. , 2002, Medical physics.

[4]  D. Low,et al.  A technique for the quantitative evaluation of dose distributions. , 1998, Medical physics.

[5]  E. B. Butler,et al.  Intensity-modulated radiation therapy (IMRT) for meningioma. , 2002, International journal of radiation oncology, biology, physics.

[6]  C. Ling,et al.  Planning, delivery, and quality assurance of intensity-modulated radiotherapy using dynamic multileaf collimator: a strategy for large-scale implementation for the treatment of carcinoma of the prostate. , 1997, International journal of radiation oncology, biology, physics.

[7]  Timothy D. Solberg,et al.  Patient specific quality assurance for the delivery of intensity modulated radiotherapy , 2003, Journal of applied clinical medical physics.

[8]  Benedick A Fraass,et al.  Incorporation of realistic delivery limitations into dynamic MLC treatment delivery. , 2002, Medical physics.

[9]  D A Jaffray,et al.  The effects of intra-fraction organ motion on the delivery of dynamic intensity modulation. , 1998, Physics in medicine and biology.

[10]  C. Ling,et al.  Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapy. , 1998, Medical physics.

[11]  John Wong,et al.  Intensity modulation to improve dose uniformity with tangential breast radiotherapy: initial clinical experience. , 2000 .

[12]  Michael J. Zelefsky,et al.  High-dose intensity modulated radiation therapy for prostate cancer: early toxicity and biochemical outcome in 772 patients. , 2002, International journal of radiation oncology, biology, physics.

[13]  K. Ohara,et al.  Irradiation synchronized with respiration gate. , 1989, International journal of radiation oncology, biology, physics.

[14]  Jean M. Moran,et al.  Verification of dynamic and segmental IMRT delivery by dynamic log file analysis , 2002, Journal of applied clinical medical physics.

[15]  S Mutic,et al.  Characterization of a commercial multileaf collimator used for intensity modulated radiation therapy. , 2001, Medical physics.

[16]  Nzhde Agazaryan,et al.  An evaluation of gating window size, delivery method, and composite field dosimetry of respiratory-gated IMRT. , 2002, Medical physics.

[17]  Gikas S. Mageras,et al.  Fluoroscopic evaluation of diaphragmatic motion reduction with a respiratory gated radiotherapy system , 2001, Journal of applied clinical medical physics.

[18]  H D Kubo,et al.  Compatibility of Varian 2100C gated operations with enhanced dynamic wedge and IMRT dose delivery. , 2000, Medical physics.

[19]  J. Deasy,et al.  A prospective study of salivary function sparing in patients with head-and-neck cancers receiving intensity-modulated or three-dimensional radiation therapy: initial results. , 2001, International journal of radiation oncology, biology, physics.

[20]  H. Mostafavi,et al.  Breathing-synchronized radiotherapy program at the University of California Davis Cancer Center. , 2000, Medical physics.

[21]  C. Ling,et al.  Effect of respiratory gating on reducing lung motion artifacts in PET imaging of lung cancer. , 2002, Medical physics.

[22]  M. Jiang,et al.  Blind deblurring of spiral CT images-comparative studies on edge-to-noise ratios. , 2002, Medical physics.

[23]  Steve B. Jiang,et al.  Effects of intra-fraction motion on IMRT dose delivery: statistical analysis and simulation. , 2002, Physics in medicine and biology.

[24]  C. Ling,et al.  Evaluation of respiratory movement during gated radiotherapy using film and electronic portal imaging. , 2002, International journal of radiation oncology, biology, physics.

[25]  T LoSasso,et al.  Comprehensive quality assurance for the delivery of intensity modulated radiotherapy with a multileaf collimator used in the dynamic mode. , 2001, Medical physics.