Alignment of the Measurement Scale Mark during Immersion Hydrometer Calibration Using an Image Processing System

The present work presents an improved method to align the measurement scale mark in an immersion hydrometer calibration system of CENAM, the National Metrology Institute (NMI) of Mexico, The proposed method uses a vision system to align the scale mark of the hydrometer to the surface of the liquid where it is immersed by implementing image processing algorithms. This approach reduces the variability in the apparent mass determination during the hydrostatic weighing in the calibration process, therefore decreasing the relative uncertainty of calibration.

[1]  S. V. Gupta,et al.  Practical Density Measurement and Hydrometry , 2002 .

[2]  N Bignell,et al.  Mass comparison of the 1 kg silicon sphere AVO#3 traceable to the International Prototype K , 2009 .

[3]  Salvatore Lorefice,et al.  Bilateral comparisons of hydrometer calibrations between the IMGC-LNE and the IMGC-MIKES , 2000 .

[4]  Kenichi Fujii,et al.  Revised formula for the density of moist air (CIPM-2007) , 2008 .

[5]  Salvatore Lorefice,et al.  Calibration of hydrometers , 2006 .

[6]  Luis Omar Becerra,et al.  Report of the bilateral comparison of the calibrations of hydrometers for liquid density determination between CENAM–Mexico and INRIM–Italy: SIM.M.D-S1 , 2009 .

[7]  Yunwoo Lee,et al.  Automatic alignment method for calibration of hydrometers , 2004 .

[8]  John D. Wright,et al.  Hydrometer calibration by hydrostatic weighing with automated liquid surface positioning , 2007 .

[9]  F. W. Cuckow A new method of high accuracy for the calibration of reference standard hydrometers , 1949 .

[10]  A Malengo,et al.  SHORT COMMUNICATION: An image processing approach to calibration of hydrometers , 2004 .

[11]  Luis Omar Becerra KEY COMPARISON: Final report of comparison of the calibrations of hydrometers for liquid density determination between SIM laboratories: SIM.M.D-K4 , 2009 .

[12]  Martti Heinonen,et al.  The effect of density gradients on hydrometers , 2003 .

[13]  Tanasko Tasic,et al.  An overview of guidance documents for software in metrological applications , 2006, Comput. Stand. Interfaces.

[14]  E. Iso,et al.  Measurement Uncertainty and Probability: Guide to the Expression of Uncertainty in Measurement , 1995 .